C.R. Bard, Inc. · EZD Catheter, Straight

Cleared on May 11, 2015 as substantially equivalent to K122785 (Rochester Medical Corp., 2013) after an 89-day review. Its family tree reaches back 5 generations to 1991. 1 later device descends from it. It has been recalled 2 times.

Decided
May 11, 2015
Review
89 days
Class
Class II
Product code
EZD
Recalls
2
Predicates
1

Its longest line of descent runs back 5 generations to K910022 (1991). 8 ancestors in all, the oldest K896750 (1990). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2001: 1 other ancestor1990: 1 other ancestor2017: 1 descendant clearedK910022 · MMG/O'neil Urinary Catheterization System Kit · Medical Marketing Group · 1991K973070 · Conveen Easicath Set · Coloplast Corp. · 1997K023254 · Speedicath · Coloplast Corp. · 2003K072808 · Speedicath Compact · Coloplast Corp. · 2007K122785 · Hydrosil; Magic; Personal Catheter · Rochester Medical Corp. · 2013 · recalled1990200020102020
  1. Root · 5 generations backK910022 MMG/O'neil Urinary Catheterization System KitMedical Marketing Group · 1991
  2. 4 generations backK973070 Conveen Easicath SetColoplast Corp. · 1997
  3. 3 generations backK023254 SpeedicathColoplast Corp. · 2003
  4. 2 generations backK072808 Speedicath CompactColoplast Corp. · 2007
  5. 1 generation backK122785 Hydrosil; Magic; Personal CatheterRochester Medical Corp. · 2013 · Recalled
  6. This deviceK150345 Magic3 Go Intermittent Urinary CatheterC.R. Bard, Inc. · 2015
  7. Descendant · cites it directly · 0 of its ownK172247 Magic3 Go Intermittent Urinary CatheterC.R. Bard, Inc. · 2017 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K172247Magic3 Go Intermittent Urinary CatheterC.R. Bard, Inc. · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K910022MMG/O'neil Urinary Catheterization System KitMedical Marketing Group1991—
−4K973070Conveen Easicath SetColoplast Corp.1997—
−3K023254SpeedicathColoplast Corp.2003—
−2K072808Speedicath CompactColoplast Corp.2007—
−1K122785Hydrosil; Magic; Personal CatheterRochester Medical Corp.20133
0 (this device)K150345Magic3 Go Intermittent Urinary CatheterC.R. Bard, Inc.2015—
+1K172247Magic3 Go Intermittent Urinary CatheterC.R. Bard, Inc.20172

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
200010
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
FIG. 1 — Lineage of K150345, 1991–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in EZD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EZD device

  1. #1 K241210 Luja Coudescore 16.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122785Hydrosil; Magic; Personal Catheter
Summary page 4
…Code: EZD Predicate Device: Legally marketed device to which substantial equivalence is claimed  HydroSil, Magic, Personal Catheter – K122785 Device Description Magic3 Go™ Intermittent Urinary Catheter is a ready-to-use silicone intermittent urinary cathete…
Rochester Medical Corp.20133

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172247Magic3 Go Intermittent Urinary CatheterC.R. Bard, Inc.20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 20, 2023Class IIZ-1701-2023Under Investigation by firmSome catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.
Apr 20, 2023Class IIZ-1698-2023Under Investigation by firmSome catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 4 companies; a new generation was cleared every 5.2 years on median.
  • 2 of the 8 ancestors have a recall of any class.
  • Of its descendants, K172247 (C.R. Bard, Inc., 2017) has gone furthest, with 0 descendants of its own.

How does its category compare?

All EZD devices

EZD Catheter, Straight: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · EZD

11so far in 2026, the most since 2007

2007: 42008: 12009: 22010: 22011: 42012: 22013: 42014: 52015: 12016: 32017: 22018: 32020: 12022: 102023: 62024: 52025: 62026 (partial year): 11
2007high 11 (2026) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EZD

130in 2025

2007: 121 days median2011: 116 days median2013: 110 days median2014: 91 days median2016: 83 days median2018: 122 days median2022: 288 days median2023: 214 days median2024: 147 days median2025: 130 days median2026 (partial year): 146 days median
2007high 288 (2022) · low 83 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · EZD

1in 2014

2011: 12012: 12014: 1
2007high 1 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · EZD

288in 2025

2009: 42010: 62011: 42012: 22013: 182014: 172015: 212016: 212017: 1202018: 3902019: 4662020: 5632021: 9942022: 4872023: 5532024: 1162025: 288
2007high 994 (2021) · low 0 (2007)2025

MAUDE reports naming product code EZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007412100
20081—00
20092—04
20102—06
2011411614
20122—12
20134110018
2014591117
20151—021
2016383021
20172—0120
201831220390
20190—0466
20201—0563
20210—0994
2022102880487
202362140553
202451470116
202561300288
2026111460—

What is it indicated for?

The Magic3 Go™ Intermittent Urinary Catheter is intended for urological use only. It is intended for use by adult female patients for bladder management including urine drainage, collection, and measurement. The device is passed to the urinary bladder via the urethra.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K150345: Magic3 Go Intermittent Urinary Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150345

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026