# K150650: neon3

- Pathway: 510(k) (Traditional)
- Applicant: Ulrich GmbH & Co. KG
- Decided: Sep 24, 2015 after 196 days
- Product code: NKG (Posterior Cervical Screw System)
- Device class: 2
- Regulation: 21 CFR 888.3075
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 3 generations back to K953076 (1997). 4 ancestors in all.
10 later devices descend from it; 3 cite it directly.
0 of 15 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K142838: Synapse Occipital-Cervical-Thoracic (OCT) System, Synthes USA Products, LLC (2015) Summary page 4.

## Most influential descendants
- K161032: neon3™  universal OCT spinal stabilization (2016), 7 descendants of its own
- K191932: Momentum™ Posterior Spinal Fixation System (2019), 2 descendants of its own
- K231809: Momentum® Posterior Spinal Fixation System with G21 V-STEADY Bone Cement; Momentum® MIS Posterior Spinal Fixation System with G21 V-STEADY Bone Cement (2023), 1 descendant of its own
- K171250: BRIDALVEIL Occipital Cervical Thoracic System (2017), 1 descendant of its own
- K251719: Momentum® Posterior Spinal Fixation System (2025), 0 descendants of its own

## Recalls of this device
- Dec 31, 2019 · Class II · Z-1350-2020: Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

## Indications for use
neon3™ is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. neon3™ is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.