510(k) Report

510(k) · Cleared

Cleared

Immunalysis Benzoylecgonine Urine Enzyme Immunoassay, Immunalysis Benzoylecgonine Urine Calibrators, Immunalysis Benzoylecgonine Urine Control Set

K150925510(k) TraditionalCleared Jun 29, 201584-day reviewClass II21 CFR 862.3250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Immunalysis Corporation · DIO Enzyme Immunoassay, Cocaine And Cocaine Metabolites

Cleared on Jun 29, 2015 as substantially equivalent to K113751 (Randox Laboratories, Ltd., 2012) after an 84-day review. Its family tree reaches back 1 generation to 2012.

Decided
Jun 29, 2015
Review
84 days
Class
Class II
Product code
DIO
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K113751 (2012). No later device cites it as a predicate yet.

K113751 · Randox Cocaine Metabolite · Randox Laboratories, Ltd. · 20122010201220142016
  1. Root · 1 generation backK113751 Randox Cocaine MetaboliteRandox Laboratories, Ltd. · 2012
  2. This deviceK150925 Immunalysis Benzoylecgonine Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K113751Randox Cocaine MetaboliteRandox Laboratories, Ltd.2012—
0 (this device)K150925Immunalysis Benzoylecgonine Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation2015—
FIG. 1 — Lineage of K150925, 2012–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DIO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DIO device

  1. #1 K213211 Carolina Liquid Chemistries Cocaine and…score 3.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K113751Randox Cocaine Metabolite
Summary page 4
…Randox Multidrug Controls, Level 1 and 2 2. Predicate Company: Randox Laboratories, Ltd. 3. Predicate K Number: K113751 Immunalysis Corporation Page 1 of 19 E.…
Randox Laboratories, Ltd.2012—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.

How does its category compare?

All DIO devices

DIO Enzyme Immunoassay, Cocaine And Cocaine Metabolites: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · DIO

1in 2022

2009: 12011: 52012: 52013: 22014: 22015: 22016: 32017: 32018: 12020: 12021: 12022: 1
2007high 5 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DIO

267in 2017

2011: 65 days median2012: 139 days median2016: 88 days median2017: 267 days median
2007high 267 (2017) · low 65 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DIO

1in 2013

2013: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DIO

1in 2023

2011: 42014: 12015: 72016: 92017: 162018: 122019: 222022: 12023: 1
2007high 22 (2019) · low 0 (2007)2025

MAUDE reports naming product code DIO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—00
20100—00
201156504
2012513900
20132—10
20142—01
20152—07
201638809
20173267016
20181—012
20190—022
20201—00
20211—00
20221—01
20230—01
20240—00
20250—00
20260—0—

What is it indicated for?

Immunalysis Benzoylecgonine Urine Enzyme Immunoassay The Immunalysis Benzoylecgonine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a dual cutoff of 150ng/mL and 300ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Benzoylecgonine in human urine with automated clinical chemistry analyzers. This assay is calibrated against Benzoylecgonine. This in-vitro device is for prescription…

Read the full statement (263 words)

Immunalysis Benzoylecgonine Urine Enzyme Immunoassay

The Immunalysis Benzoylecgonine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a dual cutoff of 150ng/mL and 300ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Benzoylecgonine in human urine with automated clinical chemistry analyzers. This assay is calibrated against Benzoylecgonine. This in-vitro device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC-MS or permitting laboratories to establish quality control procedures.

The Immunalysis Benzoylecgonine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Immunalysis Benzoylecgonine Urine Calibrators

The Immunalysis Benzoylecgonine Urine Calibrators are intended for in vitro diagnostic use for the calibration of assays for the analytes currently listed in the package insert: Benzoylecgonine. The calibrators are designed for prescription use with homogenous enzyme immunoassays on automated clinical chemistry analyzers.

Immunalysis Benzoylecgonine Urine Control Set

The Immunalysis Benzoylecgonine Urine Control Set is intended for in vitro diagnostic use to monitor the performance of assays for the analyte currently listed in the package insert: Benzoylecgonine. The controls are designed for prescription use with homogenous enzyme immunoassays on automated clinical chemistry analyzers

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K150925: Immunalysis Benzoylecgonine Urine Enzyme Immunoassay, Immunalysis Benzoylecgonine Urine Calibrators, Immunalysis Benzoylecgonine Urine Control Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150925

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026