510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Medline ENFit Connectors

K151628510(k) TraditionalCleared Jul 24, 201537-day reviewClass II21 CFR 876.5980Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medline Industries, Inc. · PIO Enteral Specific Transition Connectors

Cleared on Jul 24, 2015 as substantially equivalent to K140581 (Cedic S.R.L., 2014) after a 37-day review. Its family tree reaches back 3 generations to 1994. It has been recalled once.

Decided
Jul 24, 2015
Review
37 days
Class
Class II
Product code
PIO
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K940556 (1994). 7 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK940556 Sandoz Nutrition Compat (R) Enteral Feeding PumpSandoz Nutrition · 1994 · Recalled
  2. 2 generations backK080340 Compat Dualflo Enteral Delivery Pump Set with Spikeright Piercing…Novartis Nutrition Corp. · 2008
  3. 1 generation backK140581 Cedic Enteral Distal End Enfit Transition Connector, Cedic Enteral…Cedic S.R.L. · 2014
  4. This deviceK151628 Medline ENFit ConnectorsMedline Industries, Inc. · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K940556Sandoz Nutrition Compat (R) Enteral Feeding PumpSandoz Nutrition19942
−2K080340Compat Dualflo Enteral Delivery Pump Set with Spikeright Piercing Spike and…Novartis Nutrition Corp.2008—
−1K140581Cedic Enteral Distal End Enfit Transition Connector, Cedic Enteral Enfit…Cedic S.R.L.2014—
0 (this device)K151628Medline ENFit ConnectorsMedline Industries, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
200820
200910
FIG. 1 — Lineage of K151628, 1994–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PIO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PIO device

  1. #1 K231299 Nutricair enteral ENFit adapter: ENFit male –…score 7.3 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140581Cedic Enteral Distal End Enfit Transition Connector, Cedic Enteral Enfit Transition…
Summary page 6
…L 60060 Medline ENFit G-Tube Connector PREDICATE DEVICE Cedic ENFit Transition Connectors for Enteral Applications (K140581) are identified herein as the predicate device of the Medline ENFit Transition Connector and the Medline ENFit G-Tube Connector.…
Cedic S.R.L.2014—

Also cleared against K140581: 6 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 13, 2026Class IIZ-1690-2026Process change controlConnectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 4 companies; a new generation was cleared every 6.6 years on median.
  • 3 of the 7 ancestors have a recall of any class.
  • 6 other devices also cite its predicate K140581.

How does its category compare?

All PIO devices

PIO Enteral Specific Transition Connectors: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · PIO

2in 2024

2014: 12015: 32016: 12017: 12022: 12024: 2
2007high 3 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PIO

90in 2015

2015: 90 days median
2007high 90 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PIO

1in 2020

2020: 1
2007high 1 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PIO

9in 2025

2015: 1192016: 2152017: 32018: 52019: 72020: 62021: 92022: 52023: 102024: 62025: 9
2007high 215 (2016) · low 0 (2007)2025

MAUDE reports naming product code PIO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20141—00
20153900119
20161—0215
20171—03
20180—05
20190—07
20200—16
20210—09
20221—05
20230—010
20242—06
20250—09
20260—0—

What is it indicated for?

The Medline ENFit Transition Connector is intended for connecting an enteral feeding set with an ENFit connector to a gastronomy tube with funnel.

The Medline ENFit G-Tube Connector is intended for connecting an enteral feeding set equipped with a stepped enteral distal end to a gastronomy tube equipped with an ENFit connector.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K151628: Medline ENFit Connectors.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151628

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026