# K151767: Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Bridge, Trinidad CH Albumin BCP reagent (Alb_P), Trinidad CH Albumin BCP Calibrator

- Pathway: 510(k) (Traditional)
- Applicant: Siemens Healthcare Diagnostics, Inc.
- Decided: Dec 11, 2015 after 164 days
- Product code: JGS (Electrode, Ion Specific, Sodium)
- Device class: 2
- Regulation: 21 CFR 862.1665
- Recalls of this device: 12 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K990346 (1999). 2 ancestors in all.
3 later devices descend from it; 3 cite it directly.
0 of 6 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K990346: Bayer Advia 1650 Chemistry System, Bayer Corp. (1999) Summary page 6.
- K132664: Advia Chemistry Albumin BCP Reagent (Albp), Advia Chemistry Albumin BCP Calibrator, Siemens Healthcare Diagnostics, Inc. (2013) Summary page 6.

## Most influential descendants
- K222438: A-LYTE® Integrated Multisensor (IMT Na K Cl) (2023), 0 descendants of its own
- K222116: Atellica® CI Analyzer,  Atellica® IMThyroid Stimulating Hormone 3-Ultra (TSH3-UL), Atellica® CH Albumin BCP (AlbP) (2023), 0 descendants of its own
- K161954: Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin BCP Reagent (Alb_p), Atellica IM Thyroid Stimulating Hormone TSH, Atellica CH Vancomycin (Vanc) (2017), 0 descendants of its own

## Recalls of this device
- May 22, 2025 · Class II · Z-2027-2025: Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
- Dec 15, 2023 · Class II · Z-1003-2024: Potential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent.
- Nov 18, 2021 · Class II · Z-0449-2022: (1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemen
- May 13, 2021 · Class II · Z-2026-2021: Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. T
- May 4, 2021 · Class II · Z-1870-2021: Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results
- Dec 2, 2020 · Class II · Z-0861-2021: Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results
- Feb 27, 2020 · Class II · Z-1797-2020: Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test de
- Oct 11, 2019 · Class II · Z-0862-2020: On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Pot
- Sep 23, 2019 · Class II · Z-0581-2020: Potential for calibration error resulting in QC failures after the calibration run with QC and patient results being 5 times higher.
- Jul 30, 2019 · Class II · Z-0148-2020: Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below
- Sep 26, 2018 · Class II · Z-0782-2019: Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime
- Jun 29, 2018 · Class II · Z-0194-2019: Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs 

## Indications for use
The Trinidad CH System is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. The system’s chemical and immunochemical assay applications utilize photometric and ion selective electrode technology for clinical use. The TD-LYTE Integrated Multisensor (Na, K, Cl) is intended for in vitro diagnostic use in the quantitative determination of sodium, potassium and chloride (Na, K, Cl) in human serum, plasma and urine using the Trinidad CH System. Measurements of sodium obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. The TD-LYTE IMT Standard A is intended for the calibration of Na, K, and Cl on the Trinidad CH System. The TD-LYTE IMT Standard B + Salt Bridge is intended for the calibration of Na, K, and Cl on the Trinidad CH System. The Albumin BCP Reagent (Alb_P) is intended for in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on the Trinidad CH System. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys. The Albumin BCP calibrator is for in vitro diagnostic use in the calibration of the Trinidad CH Albumin BCP Assay (AOE_P) on the Trinidad CH System.

Source page: https://510kreport.com/device/K151767