# K151929: Persyst 13 (P13) Eeg Review and Analysis Software

- Pathway: 510(k) (Special)
- Applicant: Persyst Development Corporation
- Decided: Aug 12, 2015 after 29 days
- Product code: OMB (Automatic Event Detection Software For Full-Montage Electroencephalograph)
- Device class: 2
- Regulation: 21 CFR 882.1400
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
15 later devices descend from it; 5 cite it directly.
0 of 16 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA’s file for it is a 510(k) statement, which names no predicate.

## Most influential descendants
- K191301: Ceribell Pocket EEG Device (2019), 8 descendants of its own
- K171720: encevis (2018), 6 descendants of its own
- K223504: Ceribell Status Epilepticus Monitor (2023), 3 descendants of its own
- K211452: Encevis (2021), 3 descendants of its own
- K240993: encevis (2.1) (2024), 2 descendants of its own

## Indications for use
1. Persyst 13 EEG Review and Analysis Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information. 2. The Seizure Detection component of Persyst 13 is intended to mark previously acquired sections of adult (greater than or equal to 18 years) EEG recordings that may correspond to electrographic seizures, in order to assist qualified clinical practitioners in the assessment of EEG traces. EEG recordings should be obtained with a full scalp montage according to the standard 10/20 system. 3. The Spike Detection component of Persyst 13 is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to spikes, in order to assist qualified clinical practitioners in the assessment of EEG traces. The Spike Detection component is intended to be used in patients at least one month old. Persyst 13 Spike Detection performance has not been assessed for intracranial recordings. 4. Persyst 13 includes the calculation and display of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include FFT, Rhythmicity, Peak Envelope, Artifact Intensity, Amplitude, Relative Symmetry and Suppression Ratio. Automatic event marking is not applicable to the quantitative measures. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms. 5. The aEEG functionality included in Persyst 13 is intended to monitor the state of the brain. The automated event marking function of Persyst 13 is not applicable to aEEG. 6. Persyst 13 provides notifications for seizure detection, quantitative EEG and aEEG that can be used when processing a record during acquisition. These include an on screen display and the optional sending of an email message. Delays of up to several minutes can occur between the beginning of a seizure and when the Persyst 13 notifications will be shown to a user. Persyst 13 notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert. 7. Persyst AR (Artifact Reduction) is intended to reduce EMG, eye movement, and electrode artifacts in a standard 10-20 EEG recording. AR does not remove the entire artifact signal, and is not effective for other types of artifacts. AR may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms. 8. This device does not provide any diagnostic conclusion about the patient's condition to the user.

Source page: https://510kreport.com/device/K151929