# K151982: Stat Profile Prime ABG Analyzer System

- Pathway: 510(k) (Traditional)
- Applicant: Nova Biomedical Corporation
- Decided: Sep 10, 2015 after 55 days
- Product code: CHL (Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph)
- Device class: 2
- Regulation: 21 CFR 862.1120
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K061830 (2006). 4 ancestors in all.
No later device cites it as a predicate yet.
0 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K142220: Stat Profile Prime ABG Analyzer System (incl. Controls, Calibrators), Nova Biomedical Corporation (2014) Summary page 4.

## Recalls of this device
- Jun 19, 2017 · Class II · Z-2732-2017: Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.

## Indications for use
The Stat Profile Prime ABG Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for the quantitative determination of pH, PCO2, and PO2 in heparinized whole blood. PCO2, PO2, pH: Whole blood measurement of blood gases is used in the diagnosis and treatment of life threatening acid-base disturbances.