510(k) · Cleared
Clearedillumigene Pertussis DNA Amplification Assay
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Meridian Bioscience, Inc. · OZZ Bordetella Pertussis Dna Assay System
Cleared on Nov 10, 2015 after a 90-day review. FDA’s file for it is a 510(k) statement, which names no predicate.
- Decided
- Nov 10, 2015
- Review
- 90 days
- Class
- Class II
- Product code
- OZZ
- Recalls
- None
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OZZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OZZ device
- #1 K243753 cobas liat Bordetella panel nucleic acid testscore 12.5 · 8 generations · no Class I recall
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All OZZ devicesOZZ Bordetella Pertussis Dna Assay System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · OZZ
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OZZ
88in 2018
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OZZ
No recall names OZZ since 2007.
Adverse event reports per year · OZZ
No MAUDE report names OZZ since 2007.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 2 | — | 0 | 0 |
| 2015 | 1 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 2 | — | 0 | 0 |
| 2018 | 3 | 88 | 0 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 0 | 0 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 1 | — | 0 | 0 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The illumigene Pertussis DNA Amplification Assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection of Bordetella pertussis in human nasopharyngeal swab samples taken from patients suspected of having respiratory tract infection attributable to Bordetella pertussis. The illumigene Pertussis DNA Amplification Assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to pertussis by…
Read the full statement (271 words)
The illumigene Pertussis DNA Amplification Assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection of Bordetella pertussis in human nasopharyngeal swab samples taken from patients suspected of having respiratory tract infection attributable to Bordetella pertussis.
The illumigene Pertussis DNA Amplification Assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to pertussis by targeting the IS481 insertional element of the B. pertussis genome. The IS481 insertional element can also be found in B. holmesii and some B. bronchiseptica strains. Respiratory infections with B. pertussis, B. holmesii or B. bronchiseptica may yield positive test results in IS481 assays. B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.
Negative results for the illumigene Pertussis DNA Amplification Assay do not preclude Bordetella pertussis infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the illumigene Pertussis assay should be used in conjunction with information obtained during the patient’s clinical evaluation as an aid in diagnosis of B. pertussis infection and should not be used as the sole basis for treatment or other patient management decisions.
illumigene Pertussis DNA Amplification Assay is intended for use in hospital, reference or state laboratory settings. The device is point-of-care use.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K152285 · printed Nov 5, 2026
Cite this page
510(k) Report. “K152285: illumigene Pertussis DNA Amplification Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K152285
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026