510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Touch Ultrasound

K152467510(k) AbbreviatedCleared Dec 9, 2015100-day reviewClass II21 CFR 892.1550Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carestream Health, Inc. · IYN System, Imaging, Pulsed Doppler, Ultrasonic

Cleared on Dec 9, 2015 as substantially equivalent to K043524 (B-K Medical A/S, 2005) and 4 other predicates after a 100-day review. Its family tree reaches back 4 generations to 2000. It has been recalled 2 times.

Decided
Dec 9, 2015
Review
100 days
Class
Class II
Product code
IYN
Recalls
2
Predicates
5

Its longest line of descent runs back 4 generations to K000567 (2000). 10 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2015: 1 other ancestor2014: 1 other ancestor2013: 1 other ancestor2012: 1 other ancestor2001: 2 other ancestorsK000567 · Ultrasound Scanner Type 2102 · B & K Medical A/S · 2000 · recalledK024236 · Ultrasound Scanner, Type 2400 · B-K Medical A/S · 2003 · recalledK043524 · Ultrasound Scanner Pro Focus, Model 2202 · B-K Medical A/S · 2005 · recalledK100919 · Pro Focus 2202, Pro Focus 2202 UV Model: Type 2202 · B-K Medical Aps · 20112000200520102015
  1. Root · 4 generations backK000567 Ultrasound Scanner Type 2102B & K Medical A/S · 2000 · Recalled
  2. 3 generations backK024236 Ultrasound Scanner, Type 2400B-K Medical A/S · 2003 · Recalled
  3. 2 generations backK043524 Ultrasound Scanner Pro Focus, Model 2202B-K Medical A/S · 2005 · Recalled
  4. 1 generation backK100919 Pro Focus 2202, Pro Focus 2202 UV Model: Type 2202B-K Medical Aps · 2011
  5. This deviceK152467 Touch UltrasoundCarestream Health, Inc. · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K000567Ultrasound Scanner Type 2102B & K Medical A/S20001
−3K024236Ultrasound Scanner, Type 2400B-K Medical A/S20032
−2K043524Ultrasound Scanner Pro Focus, Model 2202B-K Medical A/S20051
−1K100919Pro Focus 2202, Pro Focus 2202 UV Model: Type 2202B-K Medical Aps2011—
0 (this device)K152467Touch UltrasoundCarestream Health, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200120
201210
201310
201410
201510
FIG. 1 — Lineage of K152467, 2000–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYN device

  1. #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043524Ultrasound Scanner Pro Focus, Model 2202
Summary page 11
…0) Diagnostic Ultrasonic Transducer (90 ITX, CFR 892.1570) Identification of predicate, legally marketed devices: B-K Medical Ultrasound Scanner Pro Focus 2202, K043524 B-K Medical Ultrasound Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Carestream Health Touch Ultr…
B-K Medical A/S20051
*K100919Pro Focus 2202, Pro Focus 2202 UV Model: Type 2202
Summary page 11
…R 892.1570) Identification of predicate, legally marketed devices: B-K Medical Ultrasound Scanner Pro Focus 2202, K043524 B-K Medical Ultrasound Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Carestream Health Touch Ultrasound, K150342 Siemens Ultrasound Scanner System Acuso…
B-K Medical Aps2011—
*K140428Ultrasound Scanner System BK2300
Summary page 11
…lly marketed devices: B-K Medical Ultrasound Scanner Pro Focus 2202, K043524 B-K Medical Ultrasound Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Carestream Health Touch Ultrasound, K150342 Siemens Ultrasound Scanner System Acuson S1000, K130619 Device description: Ultrasou…
B-K Medical Aps20141
*K150342Touch Ultrasound
Summary page 11
…Scanner Pro Focus 2202, K043524 B-K Medical Ultrasound Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Carestream Health Touch Ultrasound, K150342 Siemens Ultrasound Scanner System Acuson S1000, K130619 Device description: Ultrasound scanner and transducers for B, Tissue and…
Carestream Health, Inc.2015—
*K130619Acuson S1000 Diagnostic Ultrasound System
Summary page 11
…Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Carestream Health Touch Ultrasound, K150342 Siemens Ultrasound Scanner System Acuson S1000, K130619 Device description: Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler,…
Siemens Medical Solutions USA, Inc.2013—

Also cleared against K043524: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 21, 2016Class IIZ-1052-2017Software designSoftware error; Carestream Health Inc, received a complaint stating that when a user accidentally obtains a measurement value of 0 and corrects the value in the report, the resulting measurement unit is not displayed, i.e., centimeters or…
Jul 8, 2016Class IIZ-2217-2016Software designA software issue related to the generic volume measurement functionality could result in the volume of the subject anatomy being overestimated. For example, measurement results of the anatomy may appear enlarged or distended when it…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 4.0 years on median.
  • 4 of the 10 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K043524.

How does its category compare?

All IYN devices

IYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · IYN

57in 2025

2007: 132008: 252009: 182010: 292011: 392012: 432013: 512014: 562015: 542016: 602017: 642018: 592019: 472020: 572021: 462022: 532023: 442024: 502025: 572026 (partial year): 44
2007high 64 (2017) · low 13 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYN

124in 2025

2007: 26 days median2008: 18 days median2009: 29 days median2010: 67 days median2011: 46 days median2012: 37 days median2013: 60 days median2014: 39 days median2015: 55 days median2016: 56 days median2017: 83 days median2018: 85 days median2019: 79 days median2020: 72 days median2021: 105 days median2022: 105 days median2023: 115 days median2024: 119 days median2025: 124 days median2026 (partial year): 107 days median
2007high 124 (2025) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYN

10in 2025

2007: 162008: 122009: 82010: 102011: 82012: 32013: 82014: 22015: 32016: 22018: 72020: 42021: 82022: 102023: 42024: 12025: 102026 (partial year): 2, 1 Class I
2007high 16 (2007) · low 0 (2017)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · IYN

610in 2025

2007: 32008: 22009: 92010: 172011: 182012: 202013: 182014: 482015: 1162016: 1452017: 1462018: 852019: 592020: 672021: 2132022: 1862023: 2292024: 2662025: 610
2007high 610 (2025) · low 2 (2008)2025

MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071326163
20082518122
2009182989
201029671017
20113946818
20124337320
20135160818
20145639248
201554553116
201660562145
201764830146
20185985785
20194779059
20205772467
2021461058213
20225310510186
2023441154229
2024501191266
20255712410610
2026441072 (1)—

What is it indicated for?

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K152467: Touch Ultrasound.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152467

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026