510(k) Report

510(k) · Cleared

Cleared

Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon Catheters

K152557510(k) TraditionalCleared Feb 23, 2016168-day reviewClass II21 CFR 888.3027

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

G21, S.R.L. · NDN Cement, Bone, Vertebroplasty

Cleared on Feb 23, 2016 as substantially equivalent to K130430 (Stryker Corporation, 2013) and 6 other predicates after a 168-day review. Its family tree reaches back 7 generations to 1998. 2 later devices descend from it.

Decided
Feb 23, 2016
Review
168 days
Class
Class II
Product code
NDN
Recalls
None
Predicates
7

Its longest line of descent runs back 7 generations to K981251 (1998). 11 ancestors in all. 2 later devices descend from it, 1 citing it directly.

  1. Root · 7 generations backK981251 Kyphon Inflatable Bone TampKyphon, Inc. · 1998 · Recalled
  2. 6 generations backK032212 Kyphx Directional Inflatable Bone Tamp, Models K16A and K17AKyphon, Inc. · 2003
  3. 5 generations backK041454 Kyphx Xpander Inflatable Bone Tamps Model#KO8A,KO9A,K13AKyphon, Inc. · 2004 · Recalled
  4. 4 generations backK093419 Ivas 2-10MM (10 Gauge) Balloon Catheter, Ivas 2-15 MM (10…Stryker Instruments, Instruments Div. · 2010
  5. 3 generations backK103807 Ivas 2-10MM (10 Guage) Ballon CatheterStryker Corporation · 2011
  6. 2 generations backK113477 Ivas 20MM (10 Gauge) Balloon CatheterStryker Instruments, Instruments Div. · 2012
  7. 1 generation backK123942 Stryker Ivas Balloon CatheterStryker Corporation · 2013 · Recalled
  8. This deviceK152557 Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon CathetersG21, S.R.L. · 2016
  9. Descendant · cites it directly · 1 of its ownK172214 Modified Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon CathetersG-21 S.R.L. · 2017
  10. Descendant · 2 generations on · 0 of its ownK200763 Osteo-site Vertebral BalloonIzi Medical · 2020

Most influential descendants

Showing 2 of 2 · See all
  • K172214Modified Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon CathetersG-21 S.R.L. · 2017 · cites it directly1descendant
  • K200763Osteo-site Vertebral BalloonIzi Medical · 2020 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K981251Kyphon Inflatable Bone TampKyphon, Inc.19983
−6K032212Kyphx Directional Inflatable Bone Tamp, Models K16A and K17AKyphon, Inc.2003—
−5K041454Kyphx Xpander Inflatable Bone Tamps Model#KO8A,KO9A,K13AKyphon, Inc.20045
−4K093419Ivas 2-10MM (10 Gauge) Balloon Catheter, Ivas 2-15 MM (10 Gauge)0705-115-500…Stryker Instruments, Instruments Div.2010—
−3K103807Ivas 2-10MM (10 Guage) Ballon CatheterStryker Corporation2011—
−2K113477Ivas 20MM (10 Gauge) Balloon CatheterStryker Instruments, Instruments Div.2012—
−1K123942Stryker Ivas Balloon CatheterStryker Corporation20131
0 (this device)K152557Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon CathetersG21, S.R.L.2016—
+1K172214Modified Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon CathetersG-21 S.R.L.2017—
+2K200763Osteo-site Vertebral BalloonIzi Medical2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
201110
201210
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202010
FIG. 1 — Lineage of K152557, 1998–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NDN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NDN device

  1. #1 K241775 XeliteMed VertehighFix High Viscosity Spinal…score 14.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (7)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K130430Ivas 10MM and 15MM Balloon Catheter
Summary page 5
…code STRYKER IVAS BALLOON K130430, K123942, K113154, NDN, STRYKER…
Stryker Corporation20131
*K123942Stryker Ivas Balloon Catheter
Summary page 5
…code STRYKER IVAS BALLOON K130430, K123942, K113154, NDN, STRYKER CATHE…
Stryker Corporation20131
*K113154Ivas Balloon Catheter
Summary page 5
…code STRYKER IVAS BALLOON K130430, K123942, K113154, NDN, STRYKER CATHETER…
Stryker Corporation2012—
*K113477Ivas 20MM (10 Gauge) Balloon Catheter
Summary page 5
…942, K113154, NDN, STRYKER CATHETER K113477, K103807, K093419 HRX…
Stryker Instruments, Instruments Div.2012—
*K103807Ivas 2-10MM (10 Guage) Ballon Catheter
Summary page 5
…154, NDN, STRYKER CATHETER K113477, K103807, K093419 HRX…
Stryker Corporation2011—
*K093419Ivas 2-10MM (10 Gauge) Balloon Catheter, Ivas 2-15 MM (10 Gauge)0705-115-500, Ivas…
Summary page 5
…NDN, STRYKER CATHETER K113477, K103807, K093419 HRX…
Stryker Instruments, Instruments Div.2010—
*K110998Affirm (TM) VCF System
Summary page 5
…NDN, ALGEA THERAPIES AFFIRM 15P K110998…
Algea Therapies2011—
†K131820Avamax Vertebral Balloon System, 8G, Avamax Vertebral Balloon System, 10G, Avamax…Care Fusion2013—
†K103064Avamax Vertebral BalloonCare Fusion2011—
†K093463Avamax Vertebral Balloon Model VBS1010, VBS1015, VBS1020Care Fusion2010—

Also cleared against K130430: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172214Modified Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon CathetersG-21 S.R.L.20171

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 1.3 years on median.
  • 5 of the 11 ancestors have a recall of any class.
  • Of its descendants, K172214 (G-21 S.R.L., 2017) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K130430.

How does its category compare?

All NDN devices

NDN Cement, Bone, Vertebroplasty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · NDN

2in 2025

2007: 92008: 12009: 62010: 52011: 112012: 72013: 72014: 72015: 62016: 42017: 52018: 32019: 22020: 42021: 22022: 22023: 22024: 32025: 22026 (partial year): 1
2007high 11 (2011) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NDN

90in 2024

2007: 128 days median2009: 211 days median2010: 215 days median2011: 164 days median2012: 90 days median2013: 118 days median2014: 106 days median2015: 76 days median2016: 154 days median2017: 125 days median2018: 111 days median2020: 157 days median2024: 90 days median
2007high 215 (2010) · low 76 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NDN

4in 2013

2009: 12013: 4
2007high 4 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NDN

626in 2025

2007: 212008: 392009: 702010: 3962011: 982012: 892013: 2422014: 2312015: 2042016: 1822017: 1912018: 1442019: 2272020: 2122021: 3442022: 2702023: 3112024: 4542025: 626
2007high 626 (2025) · low 21 (2007)2025

MAUDE reports naming product code NDN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20079128021
20081—039
20096211170
201052150396
201111164098
2012790089
201371184242
201471060231
20156760204
201641540182
201751250191
201831110144
20192—0227
202041570212
20212—0344
20222—0270
20232—0311
20243900454
20252—0626
20261—0—

What is it indicated for?

The Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutancous vertebral augmentation. This system is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cements for use during percutaneous vertebral augmentation, such as kyphoplasty.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K152557: Winch Kyphoplasty (15 and 20 mm) 11 Gauge Balloon Catheters.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152557

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026