Bovie J-Plasma Precise 360 Handpiece

K152570510(k) TraditionalCleared Nov 2, 201554-day reviewClass II21 CFR 878.4400Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bovie Medical Corporation · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Nov 2, 2015 as substantially equivalent to K151325 (Bovie Medical Corporation, 2015) after a 54-day review. Its family tree reaches back 6 generations to 1989. 12 later devices descend from it. It has been recalled 5 times.

Decided
Nov 2, 2015
Review
54 days
Class
Class II
Product code
GEI
Recalls
5
Predicates
1

Its longest line of descent runs back 6 generations to K894762 (1989). 10 ancestors in all, the oldest K871435 (1987). 12 later devices descend from it, 2 citing it directly.

Other ancestors, by generationDescendants, by generation 4 generations back: 1 other ancestor5 generations back: 2 other ancestors6 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 4 descendants3 generations forward: 3 descendants4 generations forward: 2 descendants5 generations forward: 1 descendant K151325 · Bovie J-Plasma Handpiece · Bovie Medical Corporation · 2015 · recalledK112233 · Bovie J-Plasma Handpiece with Retractable Cutting Blade · Bovie Medical Corporation · 2012K090586 · Bovie Icon GS Electrosurgical Generator (GS100), Handpiece (GS-OP-R), Reusable Handpiece Cable (GS807-RC) · Bovie Medical · 2009K030819 · Plasmajet System · Plasma Surgical, Ltd. · 2004K963189 · Erbe Apc 300 Argon Plasma Coagulator and Accessories - Argon Plasma Coagulator and Accessories · Erbe USA, Inc. · 1996K894762 · B100 Disposable Handpiece · Beacon Laboratories, Inc. · 1989−50+5
  1. Root · 6 generations backK894762 B100 Disposable HandpieceBeacon Laboratories, Inc. · 1989
  2. 5 generations backK963189 Erbe Apc 300 Argon Plasma Coagulator and Accessories - Argon Plasma…Erbe USA, Inc. · 1996
  3. 4 generations backK030819 Plasmajet SystemPlasma Surgical, Ltd. · 2004
  4. 3 generations backK090586 Bovie Icon GS Electrosurgical Generator (GS100), Handpiece…Bovie Medical · 2009
  5. 2 generations backK112233 Bovie J-Plasma Handpiece with Retractable Cutting BladeBovie Medical Corporation · 2012
  6. 1 generation backK151325 Bovie J-Plasma HandpieceBovie Medical Corporation · 2015 · Recalled
  7. This deviceK152570 Bovie J-Plasma Precise 360 HandpieceBovie Medical Corporation · 2015
  8. Descendant · cites it directly · 11 of its ownK170188 Bovie J-Plasma Precise Open handpiece & Bovie Ultimate Gen 2 High…Bovie Medical Corporation · 2017 · Recalled
  9. Descendant · 2 generations on · 6 of its ownK183610 Renuvion/J-Plasma Precise Handpiece, Renuvion/J-Plasma Precise Open…Bovie Medical Corporation · 2019 · Recalled
  10. Descendant · 2 generations on · 4 of its ownK191542 Apyx Plasma/RF HandpieceBovie Medical Corporation Dba Apyx Medical Corporation · 2019
  11. Descendant · 3 generations on · 3 of its ownK220970 Renuvion APR HandpieceApyx Medical Corporation (Formerly Bovie Medical Corporation) · 2022
  12. Descendant · 4 generations on · 2 of its ownK223262 Renuvion® APR HandpieceApyx Medical Corporation · 2023

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2017: 22019: 42021: 12022: 22023: 3
2017high 4 (2019) · low 0 (2018)2026

Most influential descendants

Showing 5 of 12 · See all
  • K170188Bovie J-Plasma Precise Open handpiece & Bovie Ultimate Gen 2 High Frequency…Bovie Medical Corporation · 2017 · cites it directly11descendants
  • K183610Renuvion/J-Plasma Precise Handpiece, Renuvion/J-Plasma Precise Open handpieces…Bovie Medical Corporation · 2019 · 2 generations on6descendants
  • K191542Apyx Plasma/RF HandpieceBovie Medical Corporation Dba Apyx Medical Corporation · 2019 · 2 generations on4descendants
  • K220970Renuvion APR HandpieceApyx Medical Corporation (Formerly Bovie Medical Corporation) · 2022 · 3 generations on3descendants
  • K223262Renuvion® APR HandpieceApyx Medical Corporation · 2023 · 4 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K894762B100 Disposable HandpieceBeacon Laboratories, Inc.1989—
−5K963189Erbe Apc 300 Argon Plasma Coagulator and Accessories - Argon Plasma Coagulator…Erbe USA, Inc.1996—
−4K030819Plasmajet SystemPlasma Surgical, Ltd.2004—
−3K090586Bovie Icon GS Electrosurgical Generator (GS100), Handpiece (GS-OP-R), Reusable…Bovie Medical2009—
−2K112233Bovie J-Plasma Handpiece with Retractable Cutting BladeBovie Medical Corporation2012—
−1K151325Bovie J-Plasma HandpieceBovie Medical Corporation20151
0 (this device)K152570Bovie J-Plasma Precise 360 HandpieceBovie Medical Corporation2015—
+1K170188Bovie J-Plasma Precise Open handpiece & Bovie Ultimate Gen 2 High Frequency…Bovie Medical Corporation20176
+2K183610Renuvion/J-Plasma Precise Handpiece, Renuvion/J-Plasma Precise Open…Bovie Medical Corporation20191
+2K191542Apyx Plasma/RF HandpieceBovie Medical Corporation Dba Apyx Medical Corporation2019—
+3K220970Renuvion APR HandpieceApyx Medical Corporation (Formerly Bovie Medical Corporation)2022—
+4K223262Renuvion® APR HandpieceApyx Medical Corporation2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199020
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201720
201940
202110
202220
202330
FIG. 1 — Lineage of K152570, 1989–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K151325Bovie J-Plasma Handpiece
Summary page 5
…e Precise 360 510(k) Summary Page 1 of 7 Legally Marketed Predicate Device(s): 510(k) Number: K151325 Predicate Device Name: Bovie J-Plasma Handpiece…
Bovie Medical Corporation20151

Also cleared against K151325: 5 other devices →

What descends from it? (12)

See all 12

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170188Bovie J-Plasma Precise Open handpiece & Bovie Ultimate Gen 2 High Frequency…Bovie Medical Corporation201711
K170777Bovie J-Plasma Precise FLEX HandpieceBovie Medical Corporation20171

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 18, 2018Class IIZ-0198-2019Nonconforming Material/ComponentThere is a potential risk associated with the PEEK tip of the J-Plasma Precise 360 Handpiece separating from the handpiece and retained in the tissue.
Jul 18, 2016Class IIZ-2646-2016Nonconforming Material/ComponentThe adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.
Jul 18, 2016Class IIZ-2648-2016Nonconforming Material/ComponentThe adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.
Jul 18, 2016Class IIZ-2647-2016Nonconforming Material/ComponentThe adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.
Jul 18, 2016Class IIZ-2649-2016Nonconforming Material/ComponentThe adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 5 companies; a new generation was cleared every 5.0 years on median.
  • 2 of the 10 ancestors have a recall of any class.
  • Of its descendants, K170188 (Bovie Medical Corporation, 2017) has gone furthest, with 11 descendants of its own.
  • 5 other devices also cite its predicate K151325.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and

ablation of soft tissue during open and laparascopic surgical procedures.

The Bovie J-Plasma Precise 360 Handpiece is compatible only with Bovie J-Plasma generators.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152570: Bovie J-Plasma Precise 360 Handpiece.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152570

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026