510(k) Report

510(k) · Cleared

Cleared

TRACOE smart Cuff Manager

K152865510(k) TraditionalCleared Sep 27, 2016363-day reviewClass II21 CFR 868.5750

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tracoe Medical GmbH · BSK Cuff, Tracheal Tube, Inflatable

Cleared on Sep 27, 2016 as substantially equivalent to K833327 (Respironics, Inc., 1983) and 1 other predicate after a 363-day review. Its family tree reaches back 1 generation to 1979.

Decided
Sep 27, 2016
Review
363 days
Class
Class II
Product code
BSK
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K791045 (1979). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1983: 1 other ancestorK791045 · Model GV-10 Lanz Pressure-Valve · Extracorporeal Medical Specialities, Inc. · 19791980199020002010
Other ancestors, by generation 1 generation back: 1 other ancestor K791045 · Model GV-10 Lanz Pressure-Valve · Extracorporeal Medical Specialities, Inc. · 19790
  1. Root · 1 generation backK791045 Model GV-10 Lanz Pressure-ValveExtracorporeal Medical Specialities, Inc. · 1979
  2. This deviceK152865 TRACOE smart Cuff ManagerTracoe Medical GmbH · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K791045Model GV-10 Lanz Pressure-ValveExtracorporeal Medical Specialities, Inc.1979—
0 (this device)K152865TRACOE smart Cuff ManagerTracoe Medical GmbH2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
FIG. 1 — Lineage of K152865, 1979–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BSK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSK device

  1. #1 K261499 TrachFlushscore 13.0 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K833327Respironics PR Valve
Summary page 4
…Device Product Code 510(k) # Smith Medical Inc. PressureEasy Cuff Pressure Controller BSK K833327 Covidien Model GV-10 Lanz Pressure Valve BTR K791045 Device Description: The T…
Respironics, Inc.1983—
*K791045Model GV-10 Lanz Pressure-Valve
Summary page 4
…PressureEasy Cuff Pressure Controller BSK K833327 Covidien Model GV-10 Lanz Pressure Valve BTR K791045 Device Description: The TRACOE® smart Cuff Manager is an accessory device for high volume low pressure (HVLP) cuffed endotrache…
Extracorporeal Medical Specialities, Inc.1979—

Also cleared against K833327: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • 2 other devices also cite its predicate K833327.

How does its category compare?

All BSK devices

BSK Cuff, Tracheal Tube, Inflatable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · BSK

2in 2025

2008: 12010: 12011: 32012: 22013: 12015: 12016: 42018: 22019: 12020: 22021: 12022: 12023: 12025: 22026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSK

376in 2016

2011: 171 days median2016: 376 days median
2007high 376 (2016) · low 171 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSK

1in 2017

2013: 1, 1 Class I2017: 12026 (partial year): 1
2007high 1 (2013) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · BSK

46in 2025

2008: 22009: 32010: 52011: 152012: 622013: 162014: 52015: 12016: 12017: 32018: 42019: 22020: 12021: 92022: 42023: 512024: 1332025: 46
2007high 133 (2024) · low 0 (2007)2025

MAUDE reports naming product code BSK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—02
20090—03
20101—05
20113171015
20122—062
20131—1 (1)16
20140—05
20151—01
2016437601
20170—13
20182—04
20191—02
20202—01
20211—09
20221—04
20231—051
20240—0133
20252—046
20261—1—

What is it indicated for?

The TRACOE® smart Cuff Manager is intended to maintain the HVLP (high volume low pressure) cuff pressure, of an endotracheal tube or tracheostomy tube, within the range of 20 to 30 cm H20 through passive control.

It is for single patient use, and indicated for mechanically ventilated or spontaneously breathing patients. It is used under medical supervision in hospitals, pre-hospital (EMS), extended care facilities or outpatient clinics and is suitable for inter- or intra-facility transport.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152865: TRACOE smart Cuff Manager.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K152865

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026