510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Percuvance Percutaneous Surgical System

K153063510(k) TraditionalCleared Apr 8, 2016169-day reviewClass II21 CFR 878.4400Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Teleflexmedical, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Apr 8, 2016 as substantially equivalent to K143299 (Teleflexmedical, Inc., 2015) and 1 other predicate after a 169-day review. Its family tree reaches back 5 generations to 1991. 1 later device descends from it. It has been recalled 9 times.

Decided
Apr 8, 2016
Review
169 days
Class
Class II
Product code
GEI
Recalls
9
Predicates
2

Its longest line of descent runs back 5 generations to K912544 (1991). 14 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 5 generations backK912544 Bipolar ForcepsEverest Medical Corp. · 1991
  2. 4 generations backK972415 Minisite*Bipolar Forceps** DeviceUnited States Surgical, A Division of Tyco Healthc · 1997
  3. 3 generations backK982623 Aesculap Needlescopic Instrument SystemAesculap, Inc. · 1998
  4. 2 generations backK123102 Sovereign MiniAesculap, Inc. · 2013
  5. 1 generation backK143299 Percuvance (R) Percutaneous Surgical SystemTeleflexmedical, Inc. · 2015 · Recalled
  6. This deviceK153063 Percuvance Percutaneous Surgical SystemTeleflexmedical, Inc. · 2016
  7. Descendant · cites it directly · 0 of its ownK181852 Percuvance Percutaneous Surgical SystemTeleflexmedical, Inc. · 2018

Most influential descendants

Showing 1 of 1 · See all
  • K181852Percuvance Percutaneous Surgical SystemTeleflexmedical, Inc. · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K912544Bipolar ForcepsEverest Medical Corp.1991—
−4K972415Minisite*Bipolar Forceps** DeviceUnited States Surgical, A Division of Tyco Healthc1997—
−3K982623Aesculap Needlescopic Instrument SystemAesculap, Inc.1998—
−2K123102Sovereign MiniAesculap, Inc.2013—
−1K143299Percuvance (R) Percutaneous Surgical SystemTeleflexmedical, Inc.20152
0 (this device)K153063Percuvance Percutaneous Surgical SystemTeleflexmedical, Inc.2016—
+1K181852Percuvance Percutaneous Surgical SystemTeleflexmedical, Inc.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199430
199530
199610
199710
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
FIG. 1 — Lineage of K153063, 1991–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K143299Percuvance (R) Percutaneous Surgical System
Summary page 5
…ngeable with components from other percutaneous systems, including those of the existing Percuvance™ Percutaneous Surgical System, which was cleared under 510(k) K143299. Initial access to the surgical site is achieved with the Introducer Tool Tip attached to the Shaft. Once inside the pat…
Teleflexmedical, Inc.20152
*K083754Mini Lap Electrocautery Instruments
Summary page 7
…reference Hem-o-lok® Ligating Clip System, respectively. The following are indications for use of the predicate MiniLap® MiniPolar Electrocautery Instruments (K083754): The MiniPolar Instruments are used to cut and cauterize soft tissue. The following are indications for use…
Mini Lap Technologies, Inc.2009—
†K133202Hem-O-Lok Ligating ClipTeleflex Medical2013—
†K040855Instrumed Laparoscope and AccessoriesInstrumed International, Inc.2004—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181852Percuvance Percutaneous Surgical SystemTeleflexmedical, Inc.20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 3, 2018Class IIZ-0753-2019Process controlThere is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur…
Jan 3, 2018Class IIZ-0754-2019Process controlThere is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur…
Oct 18, 2017Class IIZ-0059-2018Under Investigation by firmTeleflex is recalling these products because the jaws of the tool tip may break, causing a fragment of the tool tip to fall off during use.
Jul 19, 2017Class IIZ-3222-2017Device DesignFor some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
Jul 19, 2017Class IIZ-3220-2017Device DesignFor some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
Jul 19, 2017Class IIZ-3223-2017Device DesignFor some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
Jul 19, 2017Class IIZ-3224-2017Device DesignFor some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
Jul 19, 2017Class IIZ-3221-2017Device DesignFor some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
Jul 19, 2017Class IIZ-3219-2017Device DesignFor some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 2.0 years on median.
  • 1 of the 14 ancestors has a recall of any class.
  • Of its descendants, K181852 (Teleflexmedical, Inc., 2018) has gone furthest, with 0 descendants of its own.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K153063: Percuvance Percutaneous Surgical System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K153063

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026