510(k) · Cleared
Recalled · Class II · 2015ECHELON Oval V 5.1 MRI System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Hitachi Medical Systems America, Inc. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Mar 31, 2016 as substantially equivalent to K151015 (Hitachi Medical Systems America, Inc., 2015) after a 111-day review. Its family tree reaches back 9 generations to 1995. 5 later devices descend from it. It has been recalled 2 times.
- Decided
- Mar 31, 2016
- Review
- 111 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 9 generations to K955249 (1995). 10 ancestors in all. 5 later devices descend from it, 1 citing it directly.
- Root · 9 generations backK955249 Sensor PadInventive Products, Inc. · 1995
- 5 generations between
- 3 generations backK083533 Echelon C Mri SystemHitachi Medical Systems America, Inc. · 2009 · Recalled
- 2 generations backK113145 Echelon Oval Mri SystemHitachi Medical Systems America, Inc. · 2012 · Recalled
- 1 generation backK151015 ECHELON Oval V 5.0 MRI SystemHitachi Medical Systems America, Inc. · 2015 · Recalled
- This deviceK153547 ECHELON Oval V 5.1 MRI SystemHitachi Medical Systems America, Inc. · 2016
- Descendant · cites it directly · 4 of its ownK172110 ECHELON OVAL V6.0A MRI SystemHitachi Healthcare Americas Corporation · 2018 · Recalled
- Descendant · 2 generations on · 3 of its ownK223426 ECHELON Synergy MRI systemFujifilm Healthcare Corporation · 2023
- Descendant · 3 generations on · 2 of its ownK233687 ECHELON Synergy V10.0FUJIFILM Healthcare Americas Corporation · 2024
- Descendant · 4 generations on · 1 of its ownK241429 ECHELON Synergy MRI SystemFujifilm Healthcare Corporation · 2024
- Descendant · 5 generations on · 0 of its ownK251386 ECHELON SynergyFujifilm Corporation · 2025
Show all 9 generations
- Root · 9 generations backK955249 Sensor PadInventive Products, Inc. · 1995
- 8 generations backK973450 Sensor PadInventive Products, Inc. · 1997
- 7 generations backK991469 Aware PadAac Consulting Group, Inc. · 1999
- 6 generations backK023390 MY Breast Friend, Breast Self-Exam PadMbf Sales, LLC · 2002
- 5 generations backK050602 AltaireHitachi Medical Systems America, Inc. · 2005
- 4 generations backK052172 HHF1 Magnetic Resonance Imaging SystemHitachi Medical Systems America, Inc. · 2005 · Recalled
- 3 generations backK083533 Echelon C Mri SystemHitachi Medical Systems America, Inc. · 2009 · Recalled
- 2 generations backK113145 Echelon Oval Mri SystemHitachi Medical Systems America, Inc. · 2012 · Recalled
- 1 generation backK151015 ECHELON Oval V 5.0 MRI SystemHitachi Medical Systems America, Inc. · 2015 · Recalled
- This deviceK153547 ECHELON Oval V 5.1 MRI SystemHitachi Medical Systems America, Inc. · 2016
Most influential descendants
Showing 5 of 5 · See all- K172110ECHELON OVAL V6.0A MRI SystemHitachi Healthcare Americas Corporation · 2018 · cites it directly4descendants
- K223426ECHELON Synergy MRI systemFujifilm Healthcare Corporation · 2023 · 2 generations on3descendants
- K233687ECHELON Synergy V10.0FUJIFILM Healthcare Americas Corporation · 2024 · 3 generations on2descendants
- K241429ECHELON Synergy MRI SystemFujifilm Healthcare Corporation · 2024 · 4 generations on1descendant
- K251386ECHELON SynergyFujifilm Corporation · 2025 · 5 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K955249 | Sensor Pad | Inventive Products, Inc. | 1995 | — |
| −8 | K973450 | Sensor Pad | Inventive Products, Inc. | 1997 | — |
| −7 | K991469 | Aware Pad | Aac Consulting Group, Inc. | 1999 | — |
| −6 | K023390 | MY Breast Friend, Breast Self-Exam Pad | Mbf Sales, LLC | 2002 | — |
| −5 | K050602 | Altaire | Hitachi Medical Systems America, Inc. | 2005 | — |
| −4 | K052172 | HHF1 Magnetic Resonance Imaging System | Hitachi Medical Systems America, Inc. | 2005 | 3 |
| −3 | K083533 | Echelon C Mri System | Hitachi Medical Systems America, Inc. | 2009 | 2 |
| −2 | K113145 | Echelon Oval Mri System | Hitachi Medical Systems America, Inc. | 2012 | 3 |
| −1 | K151015 | ECHELON Oval V 5.0 MRI System | Hitachi Medical Systems America, Inc. | 2015 | 1 |
| 0 (this device) | K153547 | ECHELON Oval V 5.1 MRI System | Hitachi Medical Systems America, Inc. | 2016 | — |
| +1 | K172110 | ECHELON OVAL V6.0A MRI System | Hitachi Healthcare Americas Corporation | 2018 | 1 |
| +2 | K223426 | ECHELON Synergy MRI system | Fujifilm Healthcare Corporation | 2023 | — |
| +3 | K233687 | ECHELON Synergy V10.0 | FUJIFILM Healthcare Americas Corporation | 2024 | — |
| +4 | K241429 | ECHELON Synergy MRI System | Fujifilm Healthcare Corporation | 2024 | — |
| +5 | K251386 | ECHELON Synergy | Fujifilm Corporation | 2025 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2018 | 1 | 0 |
| 2023 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K151015 | ECHELON Oval V 5.0 MRI SystemSummary page 5…Radiology Trade/Proprietary Name: ECHELON Oval V5.1 MRI System Predicate Device(s): ECHELON Oval V5.0 MRI System (K151015) Device Intended Use The ECHELON Oval V5.1 System is an imaging device and is intended to provide the physician with physiologic… | 2015 |
What descends from it? (5)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K172110 | ECHELON OVAL V6.0A MRI System | 2018 | 4 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 16, 2015 | Class II | Z-1540-2017 | Software in the Use Environment | Image data transferred from the MRI system to a workstation showed errors on the slice position reference image. |
| Apr 16, 2015 | Class II | Z-1541-2017 | Software in the Use Environment | Image data transferred from the MRI system to a workstation showed errors on the slice position reference image. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 21 years and passes through 4 companies; a new generation was cleared every 2.4 years on median.
- 5 of the 10 ancestors have a recall of any class.
- Of its descendants, K172110 (Hitachi Healthcare Americas Corporation, 2018) has gone furthest, with 4 descendants of its own.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The ECHELON Oval MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons…
Read the full statement (136 words)
The ECHELON Oval MRI System is an imaging device, and is intended to provide the physician with physiological and
clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces
transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head,
body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei)
exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice
relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images
provide information that can be useful in diagnosis determination.
Anatomical Region: Head, Body, Spine, Extremities
Nucleus excited: Proton
Diagnostic uses: T1, T2, proton density weighted imaging
Diffusion weighted imaging
MR Angiography
Image processing
Spectroscopy
Whole Body
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K153547 · printed Nov 6, 2026
Cite this page
510(k) Report. “K153547: ECHELON Oval V 5.1 MRI System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K153547