510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Endoform Reconstructive Template

K153633510(k) SpecialCleared Mar 15, 201688-day reviewClass II21 CFR 878.3300Recalls 21

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aroa Biosurgery Limited (Formerly Mesynthes Limited) · FTM Mesh, Surgical

Cleared on Mar 15, 2016 as substantially equivalent to K130547 (Mesynthe, Ltd., 2013) after an 88-day review. Its family tree reaches back 7 generations to 1992. 1 later device descends from it. It has been recalled 21 times.

Decided
Mar 15, 2016
Review
88 days
Class
Class II
Product code
FTM
Recalls
21
Predicates
1

Its longest line of descent runs back 7 generations to K921895 (1992). 25 ancestors in all, the oldest K830889 (1983). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2004: 2 other ancestors1997: 3 other ancestors1983: 1 other ancestor2002: 1 other ancestor2000: 5 other ancestors1992: 1 other ancestor2001: 1 other ancestor1998: 1 other ancestor1996: 1 other ancestor1985: 1 other ancestor2025: 1 descendant clearedK921895 · Tissue Patch, Tissue Graft · Bio-Vascular, Inc. · 1992K961810 · Supple Peri-Guard Pericardium · Bio-Vascular, Inc. · 1996K980431 · Surgisis · Cook Biotech, Inc. · 1998×2: 2 generations cleared close together, K043366 (2004), K050355 (2005) K070560 · LRTM Surgical Mesh · Lifecell Corp. · 2007 · recalledK130547 · Endoform Reconstructive Template · Mesynthe, Ltd. · 201319801990200020102020
  1. Root · 7 generations backK921895 Tissue Patch, Tissue GraftBio-Vascular, Inc. · 1992
  2. 6 generations backK961810 Supple Peri-Guard PericardiumBio-Vascular, Inc. · 1996
  3. 5 generations backK980431 SurgisisCook Biotech, Inc. · 1998
  4. 4 generations backK043366 Permacol Surgical ImplantTissue Science Laboratories, Plc · 2004
  5. 3 generations backK050355 Permacol Surgical Implant T-Piece, Permacol Surgical Implant…Tissue Science Laboratories, Plc · 2005
  6. 2 generations backK070560 LRTM Surgical MeshLifecell Corp. · 2007 · Recalled
  7. 1 generation backK130547 Endoform Reconstructive TemplateMesynthe, Ltd. · 2013
  8. This deviceK153633 Endoform Reconstructive TemplateAroa Biosurgery Limited (Formerly Mesynthes Limited) · 2016
  9. Descendant · cites it directly · 0 of its ownK250598 Endoform Reconstructive Template - PLGAAroa Biosurgery, Ltd. · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K250598Endoform Reconstructive Template - PLGAAroa Biosurgery, Ltd. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K921895Tissue Patch, Tissue GraftBio-Vascular, Inc.1992—
−6K961810Supple Peri-Guard PericardiumBio-Vascular, Inc.1996—
−5K980431SurgisisCook Biotech, Inc.1998—
−4K043366Permacol Surgical ImplantTissue Science Laboratories, Plc2004—
−3K050355Permacol Surgical Implant T-Piece, Permacol Surgical Implant Rectocele-Piece…Tissue Science Laboratories, Plc2005—
−2K070560LRTM Surgical MeshLifecell Corp.20072
−1K130547Endoform Reconstructive TemplateMesynthe, Ltd.2013—
0 (this device)K153633Endoform Reconstructive TemplateAroa Biosurgery Limited (Formerly Mesynthes Limited)2016—
+1K250598Endoform Reconstructive Template - PLGAAroa Biosurgery, Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198510
199210
199610
199730
199810
200050
200110
200210
200420
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K153633, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FTM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FTM device

  1. #1 K262406 EluPro Antibiotic-Eluting BioEnvelopescore 15.1 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K130547Endoform Reconstructive Template
Summary page 5
…Common or Usual Name: Surgical Mesh Classification Name: Mesh, Surgical (21 CFR §878.3300, Product Code FTM) Predicate Device: Endoform Reconstructive Template (K130547) Device Description: The Aroa Endoform Reconstructive Template (“ERT”) is a surgical mesh manufactured by layering sheets of ovi…
Mesynthe, Ltd.2013—

Also cleared against K130547: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K250598Endoform Reconstructive Template - PLGAAroa Biosurgery, Ltd.20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 6, 2018Class IIZ-0420-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0429-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0434-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0435-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0415-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0417-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0418-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0421-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0428-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0430-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0431-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0432-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0433-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0416-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0419-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0422-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0423-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0424-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0425-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0426-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…
Apr 6, 2018Class IIZ-0427-2019OtherDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 6 companies; a new generation was cleared every 2.8 years on median.
  • 2 of the 25 ancestors have a recall of any class.
  • Of its descendants, K250598 (Aroa Biosurgery, Ltd., 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K130547.

How does its category compare?

All FTM devices

FTM Mesh, Surgical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FTM

1in 2025

2007: 92008: 92009: 72010: 22011: 62012: 42013: 42014: 82015: 42016: 72017: 92018: 22019: 62020: 42021: 42023: 22024: 32025: 12026 (partial year): 1
2007high 9 (2007) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FTM

29in 2024

2007: 104 days median2008: 86 days median2009: 86 days median2011: 100 days median2012: 66 days median2013: 123 days median2014: 153 days median2015: 316 days median2016: 88 days median2017: 147 days median2019: 107 days median2020: 59 days median2021: 61 days median2024: 29 days median
2007high 316 (2015) · low 29 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FTM

1in 2023

2007: 72008: 12010: 12011: 82018: 212019: 42020: 12021: 12023: 1
2007high 21 (2018) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FTM

232in 2025

2007: 602008: 1412009: 4762010: 1972011: 2882012: 4672013: 3,3512014: 1,0932015: 9772016: 5092017: 5292018: 3572019: 4182020: 4572021: 3692022: 3752023: 4852024: 3832025: 232
2007high 3,351 (2013) · low 60 (2007)2025

MAUDE reports naming product code FTM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20079104760
20089861141
20097860476
20102—1197
201161008288
20124660467
2013412303,351
2014815301,093
201543160977
20167880509
201791470529
20182—21357
201961074418
20204591457
20214611369
20220—0375
20232—1485
20243290383
20251—0232
20261—0—

What is it indicated for?

Endoform Reconstructive Template is intended for use as a surgical mesh to reinforce and/or repair soft tissue where weakness exists. Indications for use include the repair of hernias and/or abdominal wall defects that require the use of reinforcing or bridging material to obtain the desired surgical outcome.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K153633: Endoform Reconstructive Template.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K153633

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026