510(k) · Cleared
Recalled · Class I · 2026Arrow Quickflash Arterial Catheterization Device
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Arrow International, Inc. (Subsidiary of Teleflex, Inc.) · DQX Wire, Guide, Catheter
Cleared on Jun 13, 2016 as substantially equivalent to K810675 (Arrow Intl., Inc., 1981) after a 160-day review. Its family tree reaches back 1 generation to 1981. 2 of the 2 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times, including 2 Class I recalls.
- Decided
- Jun 13, 2016
- Review
- 160 days
- Class
- Class II
- Product code
- DQX
- Recalls
- 2 (2 Class I)
- Predicates
- 1
Its longest line of descent runs back 1 generation to K810675 (1981). No later device cites it as a predicate yet. 2 of the 2 devices in this family, counting this one, have a Class I recall.
- Root · 1 generation backK810675 Arrow Radial Artery Catheterization SetArrow Intl., Inc. · 1981 · Recalled (Class I)
- This deviceK160018 Arrow Quickflash Arterial Catheterization DeviceArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2016
Class I recalls in this family
- K810675 Arrow Radial Artery Catheterization Setdirect predicate · 1981
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K810675 | Arrow Radial Artery Catheterization Set | Arrow Intl., Inc. | 1981 | Class I |
| 0 (this device) | K160018 | Arrow Quickflash Arterial Catheterization Device | Arrow International, Inc. (Subsidiary of Teleflex, Inc.) | 2016 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQX device
- #1 K213854 SavvyWirescore 15.7 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K810675 | Arrow Radial Artery Catheterization SetSummary page 4…Name: Catheter guide wire, (Class II, DQX, 21 CFR 870.1330) Percutaneous catheter, (Class II, DQY, 21 CFR 870.1250) 3. Predicate Devices • K810675: Arrow® Radial Artery Catheterization Set 4. Device Description The primary purpose of this Traditional 510(k) submission is… | 1981 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 21, 2026 | Class I | Z-2397-2026 | Nonconforming Material/Component | Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a… |
| Feb 12, 2024 | Class I | Z-1267-2024 | Process control | Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use.… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K810675Arrow Radial Artery Catheterization SetArrow Intl., Inc. · 1981 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
- Its one ancestor has a recall, Class I.
- 8 other devices also cite its predicate K810675.
How does its category compare?
All DQX devicesDQX Wire, Guide, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · DQX
10in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DQX
177in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQX
3in 2024
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · DQX
2,178in 2025
MAUDE reports naming product code DQX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 85 | 12 | 760 |
| 2008 | 19 | 87 | 7 | 715 |
| 2009 | 18 | 87 | 49 | 687 |
| 2010 | 21 | 63 | 34 | 987 |
| 2011 | 18 | 43 | 3 | 1,311 |
| 2012 | 20 | 62 | 5 | 1,337 |
| 2013 | 18 | 119 | 7 (4) | 1,474 |
| 2014 | 18 | 103 | 0 | 1,323 |
| 2015 | 15 | 112 | 1 | 1,323 |
| 2016 | 14 | 179 | 2 (2) | 1,466 |
| 2017 | 19 | 85 | 17 | 1,524 |
| 2018 | 19 | 114 | 9 | 1,737 |
| 2019 | 13 | 173 | 2 | 1,679 |
| 2020 | 14 | 106 | 1 (1) | 1,536 |
| 2021 | 6 | 161 | 0 | 1,713 |
| 2022 | 15 | 212 | 0 | 1,647 |
| 2023 | 10 | 136 | 0 | 2,068 |
| 2024 | 11 | 134 | 3 | 2,210 |
| 2025 | 10 | 177 | 0 | 2,178 |
| 2026 | 12 | 145 | 4 (1) | — |
What is it indicated for?
The Arrow Arterial Catheterization Device permits access to the peripheral arterial circulation or to other small vessels.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K160018 · printed Nov 5, 2026
Cite this page
510(k) Report. “K160018: Arrow Quickflash Arterial Catheterization Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K160018
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026