510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Octavius I, Octavius II, Octavius Iii, Octavius 4D

K160405510(k) TraditionalCleared Oct 20, 2017612-day reviewClass II21 CFR 892.5050Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ptw-Freiburg Physikalisch-Technische-Werkstaetten · IYE Accelerator, Linear, Medical

Cleared on Oct 20, 2017 as substantially equivalent to K131466 (Sun Nuclear Corporation, 2013) after a 612-day review. Its family tree reaches back 2 generations to 1994. It has been recalled 2 times.

Decided
Oct 20, 2017
Review
612 days
Class
Class II
Product code
IYE
Recalls
2
Predicates
1

Its longest line of descent runs back 2 generations to K935928 (1994). No later device cites it as a predicate yet.

K935928 · Rit 113 Film Analysis System · Radiological Imaging Technology, Inc. · 1994K131466 · MAPCHECK2, Arccheck · Sun Nuclear Corporation · 2013199019952000200520102015
  1. Root · 2 generations backK935928 Rit 113 Film Analysis SystemRadiological Imaging Technology, Inc. · 1994
  2. 1 generation backK131466 MAPCHECK2, ArccheckSun Nuclear Corporation · 2013
  3. This deviceK160405 Octavius I, Octavius II, Octavius Iii, Octavius 4DPtw-Freiburg Physikalisch-Technische-Werkstaetten · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K935928Rit 113 Film Analysis SystemRadiological Imaging Technology, Inc.1994—
−1K131466MAPCHECK2, ArccheckSun Nuclear Corporation2013—
0 (this device)K160405Octavius I, Octavius II, Octavius Iii, Octavius 4DPtw-Freiburg Physikalisch-Technische-Werkstaetten2017—
FIG. 1 — Lineage of K160405, 1994–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K131466MAPCHECK2, Arccheck
Summary page 4
…harged-particle radiation therapy system, Product Code: IYE 9. Predicate Device: Sun Nuclear’s “MapCHECK2” and “ArcCHECK” (K131466) 10. Device Description: The OCTAVIUS System is comprised of a two-dimensional ion c…
Sun Nuclear Corporation2013—

Also cleared against K131466: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 7, 2023Class IIZ-2577-2023Under Investigation by firmGenerated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result…
Aug 27, 2021Class IIZ-0479-2022Software designWhen a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 3 companies; a new generation was cleared every 19.1 years on median.
  • None of the 2 ancestors has a recall.
  • 2 other devices also cite its predicate K131466.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The OCTAVIUS System is intended to collect beam data for patient plan verification of a treatment planning system (TPS) and under the aspect of machine QA for the following purposes:

- IMRT patient plan verification

- periodic QA procedures, e.g. constancy checks

- beam data analysis according to international therapy dosimetry protocols

- measurements after repair or replacement of major treatment unit components

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160405: Octavius I, Octavius II, Octavius Iii, Octavius 4D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160405

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026