510(k) Report

510(k) · Cleared

Cleared

Contour Next Link Wireless Blood Glucose Monitoring System, Contour Next Blood Glucose Monitoring System

K160430510(k) SpecialCleared Jun 24, 2016129-day reviewClass II21 CFR 862.1345

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ascensia Diabetes Care U.S., Inc. · NBW System, Test, Blood Glucose, Over The Counter

Cleared on Jun 24, 2016 as substantially equivalent to K122370 (Bayer Healthcare, 2013) and 1 other predicate after a 129-day review. Its family tree reaches back 5 generations to 2003. 2 later devices descend from it.

Decided
Jun 24, 2016
Review
129 days
Class
Class II
Product code
NBW
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K030531 (2003). 9 ancestors in all. 2 later devices descend from it, 1 citing it directly.

  1. Root · 5 generations backK030531 Medtronic Minimed Paradigm Model 512 Insulin Pump and the BD…Minimed, Inc. · 2003 · Recalled
  2. 4 generations backK040603 Paradigm Link Blood Glucose MeterBecton, Dickinson & CO · 2004
  3. 3 generations backK073231 Onetouch Ultralink Blood Glucose Monitoring SystemLifescan, Inc. · 2008
  4. 2 generations backK110894 Contour Next Link Wireless Blood Glucose MeterBayer Healthcare · 2012
  5. 1 generation backK122370 Contour Next Link Wireless Blood Glucose MonitorBayer Healthcare · 2013
  6. This deviceK160430 Contour Next Link Wireless Blood Glucose Monitoring System, Contour…Ascensia Diabetes Care U.S., Inc. · 2016
  7. Descendant · cites it directly · 1 of its ownK191286 CONTOUR® NEXT Blood Glucose Monitoring System, CONTOUR® NEXT USB…Ascensia Diabetes Care · 2019
  8. Descendant · 2 generations on · 0 of its ownK210687 CONTOUR® NEXT ONE Blood Glucose Monitoring SystemAscensia Diabetes Care · 2021

Most influential descendants

Showing 2 of 2 · See all
  • K191286CONTOUR® NEXT Blood Glucose Monitoring System, CONTOUR® NEXT USB Blood Glucose…Ascensia Diabetes Care · 2019 · cites it directly1descendant
  • K210687CONTOUR® NEXT ONE Blood Glucose Monitoring SystemAscensia Diabetes Care · 2021 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K030531Medtronic Minimed Paradigm Model 512 Insulin Pump and the BD Paradigm Link…Minimed, Inc.20033
−4K040603Paradigm Link Blood Glucose MeterBecton, Dickinson & CO2004—
−3K073231Onetouch Ultralink Blood Glucose Monitoring SystemLifescan, Inc.2008—
−2K110894Contour Next Link Wireless Blood Glucose MeterBayer Healthcare2012—
−1K122370Contour Next Link Wireless Blood Glucose MonitorBayer Healthcare2013—
0 (this device)K160430Contour Next Link Wireless Blood Glucose Monitoring System, Contour Next Blood…Ascensia Diabetes Care U.S., Inc.2016—
+1K191286CONTOUR® NEXT Blood Glucose Monitoring System, CONTOUR® NEXT USB Blood Glucose…Ascensia Diabetes Care2019—
+2K210687CONTOUR® NEXT ONE Blood Glucose Monitoring SystemAscensia Diabetes Care2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
200910
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202110
FIG. 1 — Lineage of K160430, 2003–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NBW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NBW device

  1. #1 K260545 CONTOUR® PLUS BLUE Blood Glucose Monitoring…score 14.6 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122370Contour Next Link Wireless Blood Glucose Monitor
Summary page 5
…ext Link Wireless Blood Glucose Monitoring devices: System: Contour Next Link Wireless Blood Glucose Monitoring System K122370 Contour Next Blood Glucose Monitoring System: Contour Next Blood Glucose Monitoring Syst…
Bayer Healthcare2013—
*K121190Contour Next Blood Glucose Meter
Summary page 5
…ng System K122370 Contour Next Blood Glucose Monitoring System: Contour Next Blood Glucose Monitoring System K121190 Page 1 of 11 4) Device Conto…
Bayer Healthcare, LLC2012—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191286CONTOUR® NEXT Blood Glucose Monitoring System, CONTOUR® NEXT USB Blood Glucose…Ascensia Diabetes Care20191

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 5 companies; a new generation was cleared every 2.7 years on median.
  • 1 of the 9 ancestors has a recall of any class.
  • Of its descendants, K191286 (Ascensia Diabetes Care, 2019) has gone furthest, with 1 descendant of its own.

How does its category compare?

All NBW devices

NBW System, Test, Blood Glucose, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · NBW

2in 2025

2007: 292008: 372009: 382010: 312011: 302012: 302013: 472014: 242015: 322016: 202017: 252018: 192019: 102020: 72021: 72022: 122023: 52024: 72025: 22026 (partial year): 5
2007high 47 (2013) · low 2 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NBW

243in 2024

2007: 85 days median2008: 128 days median2009: 140 days median2010: 132 days median2011: 328 days median2012: 230 days median2013: 167 days median2014: 202 days median2015: 237 days median2016: 233 days median2017: 244 days median2018: 168 days median2019: 127 days median2020: 121 days median2021: 465 days median2022: 410 days median2023: 193 days median2024: 243 days median2026 (partial year): 190 days median
2007high 465 (2021) · low 85 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NBW

1in 2025

2007: 92008: 22009: 52010: 112011: 12012: 12013: 5, 2 Class I2014: 12016: 12019: 142025: 1
2007high 14 (2019) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · NBW

4,745in 2025

2007: 7,9912008: 7,7042009: 10,9632010: 11,6312011: 22,9092012: 23,4492013: 35,4082014: 38,5242015: 62,1112016: 76,6212017: 37,5382018: 16,9112019: 14,4722020: 10,7032021: 9,1982022: 8,1032023: 7,2562024: 6,2202025: 4,745
2007high 76,621 (2016) · low 4,745 (2025)2025

MAUDE reports naming product code NBW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007298597,991
20083712827,704
200938140510,963
2010311321111,631
201130328122,909
201230230123,449
2013471675 (2)35,408
201424202138,524
201532237062,111
201620233176,621
201725244037,538
201819168016,911
2019101271414,472
20207121010,703
2021746509,198
20221241008,103
2023519307,256
2024724306,220
20252—14,745
202651900—

What is it indicated for?

The Contour Next blood glucose monitoring system is an over-the-counter (OTC) device utilized by persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for single-patient use only, and should not be shared. The system is intended for self-testing outside the body (in vitro diagnostic use) and is to be used as an aid to monitor the effectiveness of diabetes control. The Contour Next blood glucose…

Read the full statement (399 words)

The Contour Next blood glucose monitoring system is an over-the-counter (OTC) device utilized by persons with diabetes

in home settings for the quantitative measurement of glucose in whole blood, and is for single-patient use only, and should

not be shared. The system is intended for self-testing outside the body (in vitro diagnostic use) and is to be used as an aid

to monitor the effectiveness of diabetes control.

The Contour Next blood glucose monitoring system is indicated for use with fresh capillary whole blood samples drawn

from the fingertip and palm only. Alternative site testing should be done only during steady state times (when glucose is

not changing rapidly).

The system consists of a Contour Next blood glucose meter, Contour Next test strips and Contour Next control solutions.

Contour Next test strips are intended for self-testing by persons with diabetes for the quantitative measurement of glucose

in whole blood samples.

The Contour Next blood glucose monitoring system is not intended for the diagnosis or screening for diabetes mellitus

and is not intended for use on neonates.

The Contour Next Link Wireless Blood Glucose Monitoring System is an over the counter (OTC) device utilized by

persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for single-

patient use only and should not be shared. The System is intended for self-testing outside the body (in vitro diagnostic

use) and is to be used as an aid to monitor the effectiveness of diabetes control.

The Contour Next Link Wireless Blood Glucose Monitoring System is indicated for use with fresh capillary whole blood

samples drawn from the fingertip and palm only. Alternative site testing should be done only during steady state times

(when glucose is not changing rapidly).

Contour Next Test Strips are intended for self-testing by persons with diabetes for the quantitative measurement of

glucose in whole blood samples.

The Contour Next Link Wireless Blood Glucose Monitoring System is intended to be used to transmit glucose values to

Medtronic MiniMed Paradigm Insulin Pumps or Medtronic MiniMed Paradigm REAL-Time Revel Insulin Pumps or

Medtronic MiniMed 530G Insulin Pumps and facilitate transfer of information to Medtronic MiniMed Carelink Therapy

Management Software through use of radio frequency communication.

The Contour Next Link Wireless Blood Glucose Monitoring System is not intended for the diagnosis of or screening for

diabetes mellitus and is not intended for use on neonates.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160430: Contour Next Link Wireless Blood Glucose Monitoring System, Contour Next Blood Glucose Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160430

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026