510(k) Report

510(k) · Cleared

Cleared

CorSens

K160656510(k) TraditionalCleared Nov 21, 2016258-day reviewClass II21 CFR 870.2320

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Corsens Medical, Ltd. · DXR Ballistocardiograph

Cleared on Nov 21, 2016 as substantially equivalent to K081603 (Heartforce Medical, Inc., 2009) after a 258-day review. Its family tree reaches back 2 generations to 1992.

Decided
Nov 21, 2016
Review
258 days
Class
Class II
Product code
DXR
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K910994 (1992). No later device cites it as a predicate yet.

  1. Root · 2 generations backK910994 Modified SeismocardiographSeismed Instruments, Inc. · 1992
  2. 1 generation backK081603 Digital Ballistocardiograph, Model DGB 300Heartforce Medical, Inc. · 2009
  3. This deviceK160656 CorSensCorsens Medical, Ltd. · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K910994Modified SeismocardiographSeismed Instruments, Inc.1992—
−1K081603Digital Ballistocardiograph, Model DGB 300Heartforce Medical, Inc.2009—
0 (this device)K160656CorSensCorsens Medical, Ltd.2016—
FIG. 1 — Lineage of K160656, 1992–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXR device

  1. #1 K243566 CardioTag™score 10.0 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081603Digital Ballistocardiograph, Model DGB 300
Summary page 5
…The CorSens device is substantially equivalent to the following market- cleared device: • dBG 300 Digital Ballistocardiograph (K081603) in respect to intended use, clinical indication and technologica…
Heartforce Medical, Inc.2009—
†K142827Visi Mobile Monitoring System, Visi Mobile Chest SensorSotera Wireless, Inc.2015—
†K073630Modification to: Bpcard, Model Z-7000-0700Qrs Diagnostic, LLC2008—

Also cleared against K081603: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 17.5 years on median.
  • None of the 2 ancestors has a recall.
  • 2 other devices also cite its predicate K081603.

How does its category compare?

All DXR devices

DXR Ballistocardiograph: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · DXR

1in 2025

2009: 12016: 12025: 1
2007high 1 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXR

Too few DXR 510(k)s in any year since 2007 to give a median.

Recalls per year · DXR

No recall names DXR since 2007.

Adverse event reports per year · DXR

No MAUDE report names DXR since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20161—00
20170—00
20180—00
20190—00
20200—00
20210—00
20220—00
20230—00
20240—00
20251—00
20260—0—

What is it indicated for?

CorSens Device records vibrational waveforms produced by the heart contractions and transmitted to the chest wall.

CorSens Device may be used as a tool to measure the timing of part of the events in the cardiac cycle for adult population.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160656: CorSens.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160656

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026