Vital Sync Informatics Manager & Virtual Patient Monitoring Platform

K160718510(k) TraditionalCleared Nov 14, 2016244-day reviewClass II21 CFR 870.2300Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Covidien · MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Cleared on Nov 14, 2016 as substantially equivalent to K140339 (Covidien, 2014) and 1 other predicate after a 244-day review. Its family tree reaches back 9 generations to 1993. It has been recalled once.

Decided
Nov 14, 2016
Review
244 days
Class
Class II
Product code
MWI
Recalls
1
Predicates
2

Its longest line of descent runs back 9 generations to K921669 (1993). 32 ancestors in all, the oldest K820885 (1982). No later device cites it as a predicate yet.

  1. Root · 9 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 5 generations between
  3. 3 generations backK060833 12SL Ecg Analysis ProgramGe Medical Systems Information Technologies · 2006
  4. 2 generations backK092369 12SL Ecg Analysis ProgramGe Medical Systems Information Technologies · 2010
  5. 1 generation backK133882 Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · Recalled
  6. This deviceK160718 Vital Sync Informatics Manager & Virtual Patient Monitoring PlatformCovidien · 2016
Show all 9 generations
  1. Root · 9 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 8 generations backK964750 Eagle 4000 Patient MonitorMarquette Electronics, Inc. · 1997
  3. 7 generations backK991735 Mac Series Electrocardiographs Including Mac 5000, Mac PC, Mac 6…General Electric Medical Systems Information Techn · 1999
  4. 6 generations backK014108 Mac 5000 Ecg Analysis SystemGeneral Electric Medical Systems Information Techn · 2002
  5. 5 generations backK033492 Modification to Mac 5000 Ecg Analysis SystemGe Medical Systems Information Technologies · 2003
  6. 4 generations backK042177 Modification to Mac 5000 Ecg Analysis SystemGe Medical Systems Information Technologies · 2004
  7. 3 generations backK060833 12SL Ecg Analysis ProgramGe Medical Systems Information Technologies · 2006
  8. 2 generations backK092369 12SL Ecg Analysis ProgramGe Medical Systems Information Technologies · 2010
  9. 1 generation backK133882 Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · Recalled
  10. This deviceK160718 Vital Sync Informatics Manager & Virtual Patient Monitoring PlatformCovidien · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K921669Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc.1993—
−8K964750Eagle 4000 Patient MonitorMarquette Electronics, Inc.1997—
−7K991735Mac Series Electrocardiographs Including Mac 5000, Mac PC, Mac 6, Mac VU, Mac…General Electric Medical Systems Information Techn1999—
−6K014108Mac 5000 Ecg Analysis SystemGeneral Electric Medical Systems Information Techn2002—
−5K033492Modification to Mac 5000 Ecg Analysis SystemGe Medical Systems Information Technologies2003—
−4K042177Modification to Mac 5000 Ecg Analysis SystemGe Medical Systems Information Technologies2004—
−3K06083312SL Ecg Analysis ProgramGe Medical Systems Information Technologies2006—
−2K09236912SL Ecg Analysis ProgramGe Medical Systems Information Technologies2010—
−1K133882Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc.20141
0 (this device)K160718Vital Sync Informatics Manager & Virtual Patient Monitoring PlatformCovidien2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198420
198910
199110
199410
199820
199910
200020
200310
200510
200610
200720
200920
201110
201210
201320
201410
FIG. 1 — Lineage of K160718, 1993–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MWI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MWI device

  1. #1 K230457 Empatica Health Monitoring Platform…score 28.0 · 39 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140339Vital Sync Informatics Manager & Virtual Patient Monitoring Platform
Summary page 4
…and DSI. Primary Predicate Device: Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform (K140339) Secondary Predicate Device: GE Carescape Central Station (K133882)…
Covidien20142
*K133882Carescape Central Station (Formally Known as Cic Pro)
Summary page 4
…ager & Virtual Patient Monitoring Platform (K140339) Secondary Predicate Device: GE Carescape Central Station (K133882) 1 K160718 Page 2 of 5 Device Description The subject Vita…
Ge Medical Systems Information Technologies, Inc.20141

Also cleared against K140339: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 30, 2026Class IIZ-1557-2026Under Investigation by firmDue to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 2.5 years on median.
  • 11 of the 32 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K140339.

How does its category compare?

All MWI devices

MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · MWI

4in 2025

2007: 92008: 92009: 62010: 172011: 112012: 132013: 202014: 162015: 192016: 122017: 132018: 112019: 142020: 82021: 92022: 52023: 82024: 102025: 42026 (partial year): 6
2007high 20 (2013) · low 4 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MWI

247in 2025

2007: 42 days median2008: 49 days median2009: 112 days median2010: 58 days median2011: 78 days median2012: 104 days median2013: 85 days median2014: 148 days median2015: 127 days median2016: 165 days median2017: 205 days median2018: 217 days median2019: 255 days median2020: 138 days median2021: 161 days median2022: 231 days median2023: 250 days median2024: 202 days median2025: 247 days median2026 (partial year): 239 days median
2007high 255 (2019) · low 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MWI

9so far in 2026, the most since 2007

2007: 12008: 12009: 22010: 22011: 32012: 52013: 32014: 62015: 12017: 22019: 42022: 22024: 52026 (partial year): 9
2007high 9 (2026) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MWI

182in 2025

2007: 32008: 32009: 202010: 172011: 162012: 312013: 422014: 302015: 312016: 512017: 852018: 1582019: 1692020: 1102021: 1882022: 1682023: 1792024: 1202025: 182
2007high 188 (2021) · low 3 (2007)2025

MAUDE reports naming product code MWI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200794213
200894913
20096112220
20101758217
20111178316
201213104531
20132085342
201416148630
201519127131
201612165051
201713205285
2018112170158
2019142554169
202081380110
202191610188
202252312168
202382500179
2024102025120
202542470182
202662399—

What is it indicated for?

The Vital Sync™ Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (eMR) and Clinical Information System (CIS). The Vital Sync™ Virtual Patient Monitoring Platform (VPMP) is a display system that provides visual and audible renderings of physiologic data…

Read the full statement (225 words)

The Vital Sync™ Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (eMR) and Clinical Information System (CIS).

The Vital Sync™ Virtual Patient Monitoring Platform (VPMP) is a display system that provides visual and audible renderings of physiologic data, waveforms, alarms and alerts routed through the Vital Sync™ Informatics Manager from supported devices. The Vital Sync™ Virtual Patient Monitoring Platform is intended to be used by healthcare professionals in a hospital or hospital-type facility for the following purposes:

- To remotely view and review patient data, waveforms, alerts and alarm information from supported devices and clinical information systems to facilitate clinical management.

- To facilitate remote collaboration with other healthcare professionals regarding patient data from supported devices.

- To access additional processed parameters to facilitate patient monitoring, assessment and clinical management.

- To set and adjust alert thresholds on supported devices where this capability is not available on the device itself.

- To access data, waveforms and alerts from supported devices where these capabilities are not enabled or available on the device itself.

WARNING: The Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform are notification systems and are not replacements for direct patient observation, patient assessment or clinical judgment

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K160718: Vital Sync Informatics Manager & Virtual Patient Monitoring Platform.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K160718

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026