# K160753: COOK Vacuum Pump

- Pathway: 510(k) (Traditional)
- Applicant: William A. Cook Australia Pty., Ltd.
- Decided: Jun 13, 2017 after 452 days
- Product code: MQG (Accessory, Assisted Reproduction)
- Device class: 2
- Regulation: 21 CFR 884.6120
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K992070 (1999). 1 ancestor in all.
2 later devices descend from it; 2 cite it directly.
0 of 4 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K992070: Cook Ultra Quiet Vacuum Pump & Regulator, Cook Urological, Inc. (1999) Summary page 4.

## Most influential descendants
- K261083: Wallace® Aspiration Pump (WAPV1); Wallace® Pump Filter Connector  (WFC); Wallace® Vacuum Connection Kit  (WVCK) (2026), 0 descendants of its own
- K260194: Vacuum Pump (Model: AD-VPUM-500) (2026), 0 descendants of its own

## Recalls of this device
- Mar 22, 2018 · Class II · Z-1461-2018: Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.

## Indications for use
The COOK Vacuum Pump is intended for the aspiration of eggs (ova), during assisted reproduction procedures using low flow, intermittent vacuum.

Source page: https://510kreport.com/device/K160753
Methodology: https://510kreport.com/methodology