Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6100C Series: 6100CA, 6100CP, 6100CI, 6100CN), Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6101C Series: 6101CA, 6101CP, 6101CI, 6101CN), Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6102C Series: 6102CA, 6102CP, 6102CI, 6102CN), Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor, Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Senso

K160865510(k) TraditionalCleared Aug 5, 2016128-day reviewClass II21 CFR 870.2700

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nonin Medical, Inc. · DQA Oximeter

Cleared on Aug 5, 2016 as substantially equivalent to K132402 (Nonin Medical, Inc., 2014) and 1 other predicate after a 128-day review. Its family tree reaches back 13 generations to 1993. 1 of the 85 devices in its family, counting itself, has a Class I recall. 1 later device descends from it.

Decided
Aug 5, 2016
Review
128 days
Class
Class II
Product code
DQA
Recalls
None
Predicates
3

Its longest line of descent runs back 13 generations to K931154 (1993). 83 ancestors in all, the oldest K850978 (1985). 1 later device descends from it, 1 citing it directly. 1 of the 85 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 4 other ancestors2009: 4 other ancestors2007: 2 other ancestors2010: 2 other ancestors2005: 3 other ancestors2004: 6 other ancestors2000: 6 other ancestors2003: 2 other ancestors2002: 4 other ancestors2001: 3 other ancestors1999: 2 other ancestors1997: 5 other ancestors, 1 with a Class I recall11996: 3 other ancestors2006: 2 other ancestors1998: 6 other ancestors1994: 1 other ancestor1991: 2 other ancestors1988: 1 other ancestor1986: 1 other ancestor1995: 4 other ancestors1993: 2 other ancestors1992: 2 other ancestors1990: 2 other ancestors1985: 1 other ancestor2024: 1 descendant clearedK931154 · Evis 200 System · Olympus Corp. · 1993K963033 · BF 240/P240/IT240 Bronchovideoscope & Accessories · Olympus America, Inc. · 1996 · recalledK981543 · Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and Ancillary Equipment · The Olympus Optical Co. · 1998 · recalledK002231 · LF-DP Gastrointestinal and Sigmoid Fiberscope, Accessories and Ancillary Equipment · Olympus Optical Co., Ltd. · 2001K011869 · Xenf-DP Rhino-Laryngofiberscope, Its Accessories and Ancillary Equipment · The Olympus Optical Co. · 2001K040684 · T-Stat Microvascular Tissue Oximeter, Model 303 · Spectros Corporation · 2004K051257 · Adult Cerebral Oximeter, Model 2040 · Cas Medical Systems, Inc. · 2005K061960 · Fore-Sight Cerebral Oximeter Monitor, Model 2040 · Cas Medical Systems, Inc. · 2006×2: 2 generations cleared close together, K083892 (2009), K091452 (2009) K102715 · Regional Oximeter · Nonin Medical, Inc. · 2010K113215 · Model 7600 Regional Oximeter System with Equanox Technology · Nonin Medical, Inc. · 2012K132402 · Sensmart Model X-100 Univeral Oximetry System · Nonin Medical, Inc. · 2014 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 11 other ancestors4 generations back: 16 other ancestors, 1 with a Class I recall15 generations back: 17 other ancestors6 generations back: 14 other ancestors7 generations back: 6 other ancestors8 generations back: 2 other ancestors 1 generation forward: 1 descendant K132402 · Sensmart Model X-100 Univeral Oximetry System · Nonin Medical, Inc. · 2014 · recalledK113215 · Model 7600 Regional Oximeter System with Equanox Technology · Nonin Medical, Inc. · 2012K102715 · Regional Oximeter · Nonin Medical, Inc. · 2010K091452 · Fore-Sight, Model MC-2000 Series (MC 2000, MC2010, MC-2020,MC2030) · Cas Medical Systems, Inc. · 2009K083892 · Fore-Sight, Model MC-2000 Series (MC-2000, MC-2010, MC2020, MC2030) · Cas Medical Systems, Inc. · 2009K061960 · Fore-Sight Cerebral Oximeter Monitor, Model 2040 · Cas Medical Systems, Inc. · 2006K051257 · Adult Cerebral Oximeter, Model 2040 · Cas Medical Systems, Inc. · 2005K040684 · T-Stat Microvascular Tissue Oximeter, Model 303 · Spectros Corporation · 2004K011869 · Xenf-DP Rhino-Laryngofiberscope, Its Accessories and Ancillary Equipment · The Olympus Optical Co. · 2001K002231 · LF-DP Gastrointestinal and Sigmoid Fiberscope, Accessories and Ancillary Equipment · Olympus Optical Co., Ltd. · 2001K981543 · Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and Ancillary Equipment · The Olympus Optical Co. · 1998 · recalledK963033 · BF 240/P240/IT240 Bronchovideoscope & Accessories · Olympus America, Inc. · 1996 · recalledK931154 · Evis 200 System · Olympus Corp. · 1993−10−50
  1. Root · 13 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 9 generations between
  3. 3 generations backK102715 Regional OximeterNonin Medical, Inc. · 2010
  4. 2 generations backK113215 Model 7600 Regional Oximeter System with Equanox TechnologyNonin Medical, Inc. · 2012
  5. 1 generation backK132402 Sensmart Model X-100 Univeral Oximetry SystemNonin Medical, Inc. · 2014 · Recalled
  6. This deviceK160865 Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter…Nonin Medical, Inc. · 2016
  7. Descendant · cites it directly · 0 of its ownK233827 OxiWearOxiwear, Inc. · 2024
Show all 13 generations
  1. Root · 13 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 12 generations backK963033 BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc. · 1996 · Recalled
  3. 11 generations backK981543 Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories…The Olympus Optical Co. · 1998 · Recalled
  4. 10 generations backK002231 LF-DP Gastrointestinal and Sigmoid Fiberscope, Accessories and…Olympus Optical Co., Ltd. · 2001
  5. 9 generations backK011869 Xenf-DP Rhino-Laryngofiberscope, Its Accessories and Ancillary…The Olympus Optical Co. · 2001
  6. 8 generations backK040684 T-Stat Microvascular Tissue Oximeter, Model 303Spectros Corporation · 2004
  7. 7 generations backK051257 Adult Cerebral Oximeter, Model 2040Cas Medical Systems, Inc. · 2005
  8. 6 generations backK061960 Fore-Sight Cerebral Oximeter Monitor, Model 2040Cas Medical Systems, Inc. · 2006
  9. 5 generations backK083892 Fore-Sight, Model MC-2000 Series (MC-2000, MC-2010, MC2020, MC2030)Cas Medical Systems, Inc. · 2009
  10. 4 generations backK091452 Fore-Sight, Model MC-2000 Series (MC 2000, MC2010, MC-2020,MC2030)Cas Medical Systems, Inc. · 2009
  11. 3 generations backK102715 Regional OximeterNonin Medical, Inc. · 2010
  12. 2 generations backK113215 Model 7600 Regional Oximeter System with Equanox TechnologyNonin Medical, Inc. · 2012
  13. 1 generation backK132402 Sensmart Model X-100 Univeral Oximetry SystemNonin Medical, Inc. · 2014 · Recalled
  14. This deviceK160865 Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter…Nonin Medical, Inc. · 2016

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K233827OxiWearOxiwear, Inc. · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K931154Evis 200 SystemOlympus Corp.1993—
−12K963033BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc.19966
−11K981543Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and…The Olympus Optical Co.19985
−10K002231LF-DP Gastrointestinal and Sigmoid Fiberscope, Accessories and Ancillary…Olympus Optical Co., Ltd.2001—
−9K011869Xenf-DP Rhino-Laryngofiberscope, Its Accessories and Ancillary EquipmentThe Olympus Optical Co.2001—
−8K040684T-Stat Microvascular Tissue Oximeter, Model 303Spectros Corporation2004—
−7K051257Adult Cerebral Oximeter, Model 2040Cas Medical Systems, Inc.2005—
−6K061960Fore-Sight Cerebral Oximeter Monitor, Model 2040Cas Medical Systems, Inc.2006—
−5K083892Fore-Sight, Model MC-2000 Series (MC-2000, MC-2010, MC2020, MC2030)Cas Medical Systems, Inc.2009—
−4K091452Fore-Sight, Model MC-2000 Series (MC 2000, MC2010, MC-2020,MC2030)Cas Medical Systems, Inc.2009—
−3K102715Regional OximeterNonin Medical, Inc.2010—
−2K113215Model 7600 Regional Oximeter System with Equanox TechnologyNonin Medical, Inc.2012—
−1K132402Sensmart Model X-100 Univeral Oximetry SystemNonin Medical, Inc.20141
0 (this device)K160865Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors…Nonin Medical, Inc.2016—
+1K233827OxiWearOxiwear, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
198610
198810
199020
199120
199220
199320
199410
199540
199630
199751
199860
199920
200060
200130
200240
200320
200460
200530
200620
200720
200840
200940
201020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
FIG. 1 — Lineage of K160865, 1993–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQA device

  1. #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK090853FDA’s K090853 is Yield Tissue Sampler
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Broncus Technologies, Inc.2009—
*K132402Sensmart Model X-100 Univeral Oximetry System
Summary page 8
…imetry Sensors (Models 8100SS, 8100SM and 8100SL) and Nonin Model 8000Q2 Ear Clip Sensor cleared under K09853 (June 4, 2010), K132402 (February 21, 2014) and K080255 (May 23, 2008), respectively. The proposed and predic…
Nonin Medical, Inc.20141
*K080255Nonin Medical, Inc., Model 7500 Digital Pulse Oximeter
Summary page 8
…100SM and 8100SL) and Nonin Model 8000Q2 Ear Clip Sensor cleared under K09853 (June 4, 2010), K132402 (February 21, 2014) and K080255 (May 23, 2008), respectively. The proposed and predicate sensors utilize Nonin’s PureLight®…
Nonin Medical, Inc.2008—

Also cleared against K132402: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233827OxiWearOxiwear, Inc.20240

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K970209Capnocheck PlusBci Intl., Inc. · 1997 · 4 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 7 companies; a new generation was cleared every 1.7 years on median.
  • 14 of the 83 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K233827 (Oxiwear, Inc., 2024) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K132402.

How does its category compare?

All DQA devices

DQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · DQA

15in 2025

2007: 202008: 222009: 352010: 292011: 152012: 142013: 152014: 112015: 132016: 142017: 92018: 182019: 192020: 152021: 102022: 202023: 132024: 142025: 152026 (partial year): 8
2007high 35 (2009) · low 9 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQA

182in 2025

2007: 122 days median2008: 97 days median2009: 117 days median2010: 127 days median2011: 195 days median2012: 120 days median2013: 182 days median2014: 175 days median2015: 158 days median2016: 246 days median2017: 200 days median2018: 260 days median2019: 158 days median2020: 126 days median2021: 174 days median2022: 265 days median2023: 317 days median2024: 285 days median2025: 182 days median2026 (partial year): 187 days median
2007high 317 (2023) · low 97 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQA

3in 2024

2007: 62008: 52009: 292010: 42011: 12014: 22015: 12018: 322024: 3
2007high 32 (2018) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQA

399in 2025

2007: 3252008: 3122009: 2812010: 1852011: 1642012: 3222013: 4682014: 1,1242015: 2,0822016: 1,7752017: 1,1902018: 6942019: 5482020: 4262021: 5532022: 4782023: 3922024: 5452025: 399
2007high 2,082 (2015) · low 164 (2011)2025

MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007201226325
200822975312
20093511729281
2010291274185
2011151951164
2012141200322
2013151820468
20141117521,124
20151315812,082
20161424601,775
2017920001,190
20181826032694
2019191580548
2020151260426
2021101740553
2022202650478
2023133170392
2024142853545
2025151820399
202681870—

What is it indicated for?

Nonin’s Model 6100C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6100CA: adults > 30 kg / 66 lb; 6100CP: pediatrics > 10 kg / 22 lb.; 6100CI: infants > 2 kg / 4 lb; and 6100CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions. It is intended for use hospitals, medical…

Read the full statement (753 words)

Nonin’s Model 6100C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6100CA: adults > 30 kg / 66 lb; 6100CP: pediatrics > 10 kg / 22 lb.; 6100CI: infants > 2 kg / 4 lb; and 6100CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions.

It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

Nonin’s Model 6101C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6101CA: adults > 30 kg / 66 lb; 6101CP: pediatrics > 10 kg / 22 lb.; 6101CI: infants > 2 kg / 4 lb; and 6101CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions.

It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

Nonin’s Model 6102C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6102CA: adults > 30 kg / 66 lb; 6102CP: pediatrics > 10 kg / 22 lb.; 6102CI: infants > 2 kg / 4 lb; and 6102CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions.

It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

The Model 8100AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.

The Model 8100AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.

It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

The Model 8101AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.

The Model 8101AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.

It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

The Model 8102AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.

The Model 8102AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.

It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

The Model 8100Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

The recommended application site is the earlobe.

The Model 8101Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

The recommended application site is the earlobe.

The Model 8102Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.

The recommended application site is the earlobe.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160865: Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6100C Series: 6100CA, 6100CP, 6100CI, 6100CN), Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6101C Series: 6101CA, 6101CP, 6101CI, 6101CN), Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6102C Series: 6102CA, 6102CP, 6102CI, 6102CN), Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor, Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Senso.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K160865

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.