510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

HIGH V+

K161114510(k) TraditionalCleared Apr 12, 2017357-day reviewClass II21 CFR 888.3027Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Teknimed Sas · NDN Cement, Bone, Vertebroplasty

Cleared on Apr 12, 2017 as substantially equivalent to K090435 (Teknimed, S.A., 2009) and 1 other predicate after a 357-day review. Its family tree reaches back 1 generation to 2004. 42 later devices descend from it. It has been recalled once.

Decided
Apr 12, 2017
Review
357 days
Class
Class II
Product code
NDN
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K032945 (2004). 2 ancestors in all. 42 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2018: 1 descendant cleared2019: 3 descendants cleared2022: 3 descendants cleared2020: 3 descendants cleared2023: 8 descendants cleared2024: 10 descendants cleared2021: 2 descendants cleared2025: 5 descendants cleared2026: 7 descendants clearedK032945 · Simplex X · Stryker Corp. · 2004 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 4 descendants3 generations forward: 11 descendants4 generations forward: 13 descendants5 generations forward: 9 descendants6 generations forward: 2 descendants7 generations forward: 1 descendant K032945 · Simplex X · Stryker Corp. · 2004 · recalled0+5
  1. Root · 1 generation backK032945 Simplex XStryker Corp. · 2004 · Recalled
  2. This deviceK161114 HIGH V+Teknimed Sas · 2017
  3. Descendant · cites it directly · 41 of its ownK180498 NuVasive Reline Fenestrated Screws, High V+ Bone CementNu Vasive, Incorporated · 2018
  4. Descendant · cites it directly · 28 of its ownK190526 High V+ Bone Cement, CarboClear® Fenestrated Pedicle ScrewsCarboFix Orthopedics, Ltd. · 2019
  5. Descendant · 2 generations on · 27 of its ownK200596 G21 Cement, VADER® Pedicle SystemIcotec AG · 2020 · Recalled
  6. Descendant · 2 generations on · 14 of its ownK221926 Invictus® Bone Cement, Invictus Spinal Fixation SystemAlphatec Spine, Inc. · 2022
  7. Descendant · 3 generations on · 11 of its ownK232275 Invictus® Spinal Fixation SystemAlphatec Spine, Inc. · 2023

Showing 5 of 42 descendants. See all descendants

Descendants cleared per year

2018: 12019: 32020: 32021: 22022: 32023: 82024: 102025: 52026: 7
2018high 10 (2024) · low 1 (2018)2026

Most influential descendants

Showing 5 of 42 · See all
  • K180498NuVasive Reline Fenestrated Screws, High V+ Bone CementNu Vasive, Incorporated · 2018 · cites it directly41descendants
  • K190526High V+ Bone Cement, CarboClear® Fenestrated Pedicle ScrewsCarboFix Orthopedics, Ltd. · 2019 · cites it directly28descendants
  • K200596G21 Cement, VADER® Pedicle SystemIcotec AG · 2020 · 2 generations on27descendants
  • K221926Invictus® Bone Cement, Invictus Spinal Fixation SystemAlphatec Spine, Inc. · 2022 · 2 generations on14descendants
  • K232275Invictus® Spinal Fixation SystemAlphatec Spine, Inc. · 2023 · 3 generations on11descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K032945Simplex XStryker Corp.20042
0 (this device)K161114HIGH V+Teknimed Sas2017—
+1K180498NuVasive Reline Fenestrated Screws, High V+ Bone CementNu Vasive, Incorporated2018—
+1K190526High V+ Bone Cement, CarboClear® Fenestrated Pedicle ScrewsCarboFix Orthopedics, Ltd.2019—
+2K200596G21 Cement, VADER® Pedicle SystemIcotec AG20201
+2K221926Invictus® Bone Cement, Invictus Spinal Fixation SystemAlphatec Spine, Inc.2022—
+3K232275Invictus® Spinal Fixation SystemAlphatec Spine, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
201930
202030
202120
202230
202380
2024100
202550
202670
FIG. 1 — Lineage of K161114, 2004–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NDN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NDN device

  1. #1 K241775 XeliteMed VertehighFix High Viscosity Spinal…score 14.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090435Vertecem
Summary page 4
…Product Code: NDN CFR section: 21 CFR section 888.3027 Device panel: Orthopedic PrimaryLegally Vertecem, K090435 Marketed Predicate Device: Secondary Legally Spineplex K032945 Market…
Teknimed, S.A.2009—
*K032945Simplex X
Summary page 4
…panel: Orthopedic PrimaryLegally Vertecem, K090435 Marketed Predicate Device: Secondary Legally Spineplex K032945 Marketed Predicate Device: Submitter: Teknimed, SAS…
Stryker Corp.20042

Also cleared against K090435: 3 other devices →

What descends from it? (42)

See all 42

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180498NuVasive Reline Fenestrated Screws, High V+ Bone CementNu Vasive, Incorporated201841
K190526High V+ Bone Cement, CarboClear® Fenestrated Pedicle ScrewsCarboFix Orthopedics, Ltd.201928

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 15, 2022Class IIZ-1099-2023Employee errorIncorrect expiry date. Product is labeled with expiry date 2027-02-28. The correct expiration date is 2025-02-28.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K180498 (Nu Vasive, Incorporated, 2018) has gone furthest, with 41 descendants of its own.
  • 3 other devices also cite its predicate K090435.

How does its category compare?

All NDN devices

NDN Cement, Bone, Vertebroplasty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · NDN

2in 2025

2007: 92008: 12009: 62010: 52011: 112012: 72013: 72014: 72015: 62016: 42017: 52018: 32019: 22020: 42021: 22022: 22023: 22024: 32025: 22026 (partial year): 1
2007high 11 (2011) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NDN

90in 2024

2007: 128 days median2009: 211 days median2010: 215 days median2011: 164 days median2012: 90 days median2013: 118 days median2014: 106 days median2015: 76 days median2016: 154 days median2017: 125 days median2018: 111 days median2020: 157 days median2024: 90 days median
2007high 215 (2010) · low 76 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NDN

4in 2013

2009: 12013: 4
2007high 4 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NDN

626in 2025

2007: 212008: 392009: 702010: 3962011: 982012: 892013: 2422014: 2312015: 2042016: 1822017: 1912018: 1442019: 2272020: 2122021: 3442022: 2702023: 3112024: 4542025: 626
2007high 626 (2025) · low 21 (2007)2025

MAUDE reports naming product code NDN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20079128021
20081—039
20096211170
201052150396
201111164098
2012790089
201371184242
201471060231
20156760204
201641540182
201751250191
201831110144
20192—0227
202041570212
20212—0344
20222—0270
20232—0311
20243900454
20252—0626
20261—0—

What is it indicated for?

The HIGH V+ is used for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures.

Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K161114: HIGH V+.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K161114

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026