510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

StageOne Select Cement Spacer Molds for Temporary Hip Replacement

K161166510(k) TraditionalCleared Sep 8, 2016135-day reviewClass II21 CFR 888.3027Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biomet, Inc. · MBB Bone Cement, Antibiotic

Cleared on Sep 8, 2016 as substantially equivalent to K080979 (Biomet Manufacturing Corp, 2008) after a 135-day review. Its family tree reaches back 4 generations to 1998. 2 later devices descend from it. It has been recalled 15 times.

Decided
Sep 8, 2016
Review
135 days
Class
Class II
Product code
MBB
Recalls
15
Predicates
1

Its longest line of descent runs back 4 generations to K974807 (1998). 2 later devices descend from it, 2 citing it directly.

  1. Root · 4 generations backK974807 Osteo Austin Moore Endoprosthesis SystemOsteonics Corp. · 1998
  2. 3 generations backK031841 Spacer G Temporary Hip ProsthesisExactech, Inc. · 2004
  3. 2 generations backK052990 Stageone Disposable Cement Spacer Mold for Temporary Hip Prosthesis…Biomet, Inc. · 2005 · Recalled
  4. 1 generation backK080979 Stageone Select Cement Spacer Molds for Temporary Hip ReplacementBiomet Manufacturing Corp · 2008 · Recalled
  5. This deviceK161166 StageOne Select Cement Spacer Molds for Temporary Hip ReplacementBiomet, Inc. · 2016
  6. Descendant · cites it directly · 0 of its ownK222760 StageOne™ Select Hip Cement Spacer MoldsBiomet, Inc. · 2022
  7. Descendant · cites it directly · 0 of its ownK192041 G21 SpaceFlex HipG21, S.R.L. · 2019

Most influential descendants

Showing 2 of 2 · See all
  • K222760StageOne™ Select Hip Cement Spacer MoldsBiomet, Inc. · 2022 · cites it directly0descendants
  • K192041G21 SpaceFlex HipG21, S.R.L. · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K974807Osteo Austin Moore Endoprosthesis SystemOsteonics Corp.1998—
−3K031841Spacer G Temporary Hip ProsthesisExactech, Inc.2004—
−2K052990Stageone Disposable Cement Spacer Mold for Temporary Hip Prosthesis with…Biomet, Inc.20055
−1K080979Stageone Select Cement Spacer Molds for Temporary Hip ReplacementBiomet Manufacturing Corp200815
0 (this device)K161166StageOne Select Cement Spacer Molds for Temporary Hip ReplacementBiomet, Inc.2016—
+1K222760StageOne™ Select Hip Cement Spacer MoldsBiomet, Inc.2022—
+1K192041G21 SpaceFlex HipG21, S.R.L.2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202210
FIG. 1 — Lineage of K161166, 1998–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBB device

  1. #1 K252443 SpaceFix Shoulder Spacerscore 12.4 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080979Stageone Select Cement Spacer Molds for Temporary Hip Replacement
Summary page 4
…MBB-Polymethylmethacrylate (PMMA) bone cement (21 CFR 888.3027) Legally marketed devices to which substantial equivalence is claimed: x K080979-StageOne™ Select Cement Spacer Molds for Temporary Hip Replacement Device Description The single-use cement spacer molds are ste…
Biomet Manufacturing Corp200815

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K222760StageOne™ Select Hip Cement Spacer MoldsBiomet, Inc.20220
K192041G21 SpaceFlex HipG21, S.R.L.20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 26, 2020Class IIZ-1640-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1638-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1650-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1646-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1642-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1648-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1651-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1641-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1649-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1643-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1645-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1644-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1647-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1652-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
Feb 26, 2020Class IIZ-1639-2020Process controlPotentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 5.9 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K222760 (Biomet, Inc., 2022) has gone furthest, with 0 descendants of its own.

How does its category compare?

All MBB devices

MBB Bone Cement, Antibiotic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · MBB

1in 2025

2007: 12008: 12014: 12015: 12016: 22017: 12019: 22020: 12021: 12022: 32024: 22025: 1
2007high 3 (2022) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBB

106in 2022

2022: 106 days median
2007high 106 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBB

1in 2022

2013: 22022: 12026 (partial year): 2
2007high 2 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MBB

172in 2025

2007: 12008: 32009: 92010: 182011: 312012: 342013: 322014: 412015: 202016: 3272017: 2882018: 3322019: 4972020: 3752021: 6492022: 2752023: 2622024: 2122025: 172
2007high 649 (2021) · low 1 (2007)2025

MAUDE reports naming product code MBB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
20081—03
20090—09
20100—018
20110—031
20120—034
20130—232
20141—041
20151—020
20162—0327
20171—0288
20180—0332
20192—0497
20201—0375
20211—0649
202231061275
20230—0262
20242—0212
20251—0172
20260—2—

What is it indicated for?

Disposable cement spacer molds with stainless steel reinforcement stems, adapters and inserts are indicated for use to mold a temporary hemi-hip replacement for skeletally mature patients undergoing a two-stage revision procedure due to a septic process. The temporary prosthesis is molded using PALACOS® R+G Bone Cement or Refobacin® Bone Cement R, assembled and inserted into the femoral medullary canal and acetabular cavity following removal of the…

Read the full statement (177 words)

Disposable cement spacer molds with stainless steel reinforcement stems, adapters and inserts are indicated for use to

mold a temporary hemi-hip replacement for skeletally mature patients undergoing a two-stage revision procedure due to a

septic process. The temporary prosthesis is molded using PALACOS® R+G Bone Cement or Refobacin® Bone

Cement R, assembled and inserted into the femoral medullary canal and acetabular cavity following removal of the

existing femoral and acetabular implants and debridement. The device is intended for use in conjunction with systemic

antimicrobial antibiotic therapy (standard treatment approach to an infection).

The hemi-hip prosthesis made from the StageOne™ Select disposable cement molds is not intended for use more than 180

days, at which time it must be explanted and permanent devices implanted or another appropriate treatment performed

(e.g. resection arthroplasty, fusion, etc.)

Due to the inherent mechanical limitations of the hemi-hip prosthesis material (PALACOS® R+G Bone Cement and

Refobacin® Bone Cement R), the temporary hemi-hip prosthesis is only indicated for patients who will consistently use

traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K161166: StageOne Select Cement Spacer Molds for Temporary Hip Replacement.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161166

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026