PLEX Elite 9000 SS-OCT

K161194510(k) TraditionalCleared Oct 26, 2016182-day reviewClass II21 CFR 886.1570Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carl Zeiss Meditec, Inc. · OBO Tomography, Optical Coherence

Cleared on Oct 26, 2016 as substantially equivalent to K150977 (Carl Zeiss Meditec, Inc., 2015) after a 182-day review. Its family tree reaches back 8 generations to 2002. 7 later devices descend from it. It has been recalled once.

Decided
Oct 26, 2016
Review
182 days
Class
Class II
Product code
OBO
Recalls
1
Predicates
1

Its longest line of descent runs back 8 generations to K012727 (2002). 24 ancestors in all, the oldest K946058 (1995). 7 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2010: 1 other ancestor2009: 1 other ancestor2007: 1 other ancestor2006: 3 other ancestors2004: 2 other ancestors2002: 2 other ancestors2003: 1 other ancestor2001: 1 other ancestor1998: 1 other ancestor1995: 2 other ancestors2019: 1 descendant cleared2024: 3 descendants cleared2023: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant clearedReference device: K143275 · IOLMaster700 · Carl Zeiss Meditec, Inc. · 2015K012727 · Humphrey OCT3 · Carl Zeiss Ophthalmic Systems, Inc. · 2002K030433 · Stratusoct with RNFL Normative Database · Carl Zeiss Meditec, Inc. · 2003 · recalledK033123 · Stratusoct with RNFL & Macula Normative Database · Carl Zeiss Meditec, Inc. · 2004 · recalledK063343 · Bioptigen Spectral Domain Ophthalmic Imaging System · Bioptigen, Inc. · 2006K071250 · CA, Model: Cam · Optovue, Inc. · 2007K101505 · Rtvue Model Rtvue 100 · Optovue, Inc. · 2010 · recalledK111157 · CIRRUS HD-OCT WITH RETINAL NERVE FIBER LAYER, MACULAR, OPTIC NERVE HEAD, and GANGLION CELL NORMATIVE DATABASES · Carl Zeiss Meditec, Inc. · 2012K150977 · Cirrus HD-Oct · Carl Zeiss Meditec, Inc. · 20151995200020052010201520202025
  1. Root · 8 generations backK012727 Humphrey OCT3Carl Zeiss Ophthalmic Systems, Inc. · 2002
  2. 7 generations backK030433 Stratusoct with RNFL Normative DatabaseCarl Zeiss Meditec, Inc. · 2003 · Recalled
  3. 6 generations backK033123 Stratusoct with RNFL & Macula Normative DatabaseCarl Zeiss Meditec, Inc. · 2004 · Recalled
  4. 5 generations backK063343 Bioptigen Spectral Domain Ophthalmic Imaging SystemBioptigen, Inc. · 2006
  5. 4 generations backK071250 CA, Model: CamOptovue, Inc. · 2007
  6. 3 generations backK101505 Rtvue Model Rtvue 100Optovue, Inc. · 2010 · Recalled
  7. 2 generations backK111157 CIRRUS HD-OCT WITH RETINAL NERVE FIBER LAYER, MACULAR, OPTIC NERVE…Carl Zeiss Meditec, Inc. · 2012
  8. 1 generation backK150977 Cirrus HD-OctCarl Zeiss Meditec, Inc. · 2015
  9. This deviceK161194 PLEX Elite 9000 SS-OCTCarl Zeiss Meditec, Inc. · 2016
  10. Descendant · cites it directly · 6 of its ownK181534 Cirrus HD-OctCarl Zeiss Meditec, Inc. · 2019 · Recalled
  11. Descendant · 2 generations on · 1 of its ownK241081 3D OPTICAL COHERENCE TOMOGRAPHY 3D OCT-1 (type: Maestro2); IMAGEnet6…Topcon Corporation · 2024
  12. Descendant · 2 generations on · 1 of its ownK231760 HP-Oct (HP-Oct)Cylite Pty., Ltd. · 2024
  13. Descendant · 2 generations on · 1 of its ownK222200 Cirrus HD-OctCarl Zeiss Meditec, Inc. · 2023
  14. Descendant · 3 generations on · 0 of its ownK253893 Harmony Referral System with Harmony OCT Viewer (Octave)Topcon Healthcare Solutions EMEA Oy · 2026

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2019: 12023: 12024: 32025: 12026: 1
2019high 3 (2024) · low 0 (2020)2026

Most influential descendants

Showing 5 of 7 · See all
  • K181534Cirrus HD-OctCarl Zeiss Meditec, Inc. · 2019 · cites it directly6descendants
  • K2410813D OPTICAL COHERENCE TOMOGRAPHY 3D OCT-1 (type: Maestro2); IMAGEnet6 Ophthalmic Data…Topcon Corporation · 2024 · 2 generations on1descendant
  • K231760HP-Oct (HP-Oct)Cylite Pty., Ltd. · 2024 · 2 generations on1descendant
  • K222200Cirrus HD-OctCarl Zeiss Meditec, Inc. · 2023 · 2 generations on1descendant
  • K253893Harmony Referral System with Harmony OCT Viewer (Octave)Topcon Healthcare Solutions EMEA Oy · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K012727Humphrey OCT3Carl Zeiss Ophthalmic Systems, Inc.2002—
−7K030433Stratusoct with RNFL Normative DatabaseCarl Zeiss Meditec, Inc.20031
−6K033123Stratusoct with RNFL & Macula Normative DatabaseCarl Zeiss Meditec, Inc.20041
−5K063343Bioptigen Spectral Domain Ophthalmic Imaging SystemBioptigen, Inc.2006—
−4K071250CA, Model: CamOptovue, Inc.2007—
−3K101505Rtvue Model Rtvue 100Optovue, Inc.20101
−2K111157CIRRUS HD-OCT WITH RETINAL NERVE FIBER LAYER, MACULAR, OPTIC NERVE HEAD, and…Carl Zeiss Meditec, Inc.2012—
−1K150977Cirrus HD-OctCarl Zeiss Meditec, Inc.2015—
0 (this device)K161194PLEX Elite 9000 SS-OCTCarl Zeiss Meditec, Inc.2016—
+1K181534Cirrus HD-OctCarl Zeiss Meditec, Inc.20192
+2K2410813D OPTICAL COHERENCE TOMOGRAPHY 3D OCT-1 (type: Maestro2); IMAGEnet6…Topcon Corporation2024—
+2K231760HP-Oct (HP-Oct)Cylite Pty., Ltd.2024—
+2K222200Cirrus HD-OctCarl Zeiss Meditec, Inc.2023—
+3K253893Harmony Referral System with Harmony OCT Viewer (Octave)Topcon Healthcare Solutions EMEA Oy2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199520
199810
200110
200220
200310
200420
200510
200630
200710
200910
201010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202310
202430
202510
202610
FIG. 1 — Lineage of K161194, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OBO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OBO device

  1. #1 K250868 SPECTRALIS HRA+OCT and variantsscore 16.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K150977Cirrus HD-Oct
Summary page 4
…mography System Substantial Equivalence Claimed To (21 CFR §807.92(a)(3)) Company: Carl Zeiss Meditec, Inc. Device: CIRRUS™ HD-OCT (K150977) Reference Predicate: Company: Carl Zeiss Meditec AG Device: IOLMaster 700 (K143275)…
Carl Zeiss Meditec, Inc.2015—
†K143275IOLMaster700Carl Zeiss Meditec, Inc.2015—

Also cleared against K150977: 3 other devices →

What descends from it? (7)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181534Cirrus HD-OctCarl Zeiss Meditec, Inc.20196

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 16, 2018Class IIZ-0770-2020Device DesignUnder certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 1.3 years on median.
  • 4 of the 24 ancestors have a recall of any class.
  • Of its descendants, K181534 (Carl Zeiss Meditec, Inc., 2019) has gone furthest, with 6 descendants of its own.
  • 3 other devices also cite its predicate K150977.

How does its category compare?

All OBO devices

OBO Tomography, Optical Coherence: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · OBO

3in 2025

2007: 22008: 22009: 42010: 22011: 22012: 32013: 22014: 32015: 12016: 42017: 42018: 62019: 52020: 22021: 12022: 22023: 72024: 82025: 32026 (partial year): 2
2007high 8 (2024) · low 1 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OBO

125in 2025

2009: 151 days median2012: 269 days median2014: 152 days median2016: 125 days median2017: 118 days median2018: 87 days median2019: 249 days median2023: 262 days median2024: 157 days median2025: 125 days median
2007high 269 (2012) · low 87 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OBO

1in 2022

2010: 12011: 22012: 22018: 12021: 12022: 12026 (partial year): 1
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OBO

4in 2025

2013: 22016: 12018: 12021: 12023: 32024: 32025: 4
2007high 4 (2025) · low 0 (2007)2025

MAUDE reports naming product code OBO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
2009415100
20102—10
20112—20
2012326920
20132—02
2014315200
20151—00
2016412501
2017411800
201868711
2019524900
20202—00
20211—11
20222—10
2023726203
2024815703
2025312504
20262—1—

What is it indicated for?

The PLEX™ Elite 9000 Swept-Source OCT [SS-OCT] is a non-contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head, vitreous and choroid.

The PLEX™ Elite SS-OCT angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161194: PLEX Elite 9000 SS-OCT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K161194

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026