510(k) · Cleared
ClearedEVEREST Spinal System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
K2m, Incorporated · NKB Thoracolumbosacral Pedicle Screw System
Cleared on Jul 21, 2016 as substantially equivalent to K151727 (K2m, Inc., 2015) and 1 other predicate after a 65-day review. Its family tree reaches back 21 generations to 1985. 9 later devices descend from it.
- Decided
- Jul 21, 2016
- Review
- 65 days
- Class
- Class II
- Product code
- NKB
- Recalls
- None
- Predicates
- 2
Its longest line of descent runs back 21 generations to K853033 (1985). 349 ancestors in all, the oldest K781443 (1978). 9 later devices descend from it, 3 citing it directly.
- Root · 21 generations backK853033 Hartshill Rect & Wires, Spinal Fixation SystemA.W. Showell (Surgicraft), Ltd. · 1985
- 17 generations between
- 3 generations backK143334 Range/Denali/Mesa Spinal SystemK2m, Inc. · 2015
- 2 generations backK150325 Range/Denali/Mesa Spinal SystemK2m, Inc. · 2015
- 1 generation backK151727 Everest Spinal SystemK2m, Inc. · 2015
- This deviceK161369 EVEREST Spinal SystemK2m, Incorporated · 2016
- Descendant · cites it directly · 6 of its ownK181188 EVEREST Spinal SystemK2m, Inc. · 2018
- Descendant · 2 generations on · 5 of its ownK181603 EVEREST Spinal System, RANGE (MESA and DENALI) Spinal System…K2m, Inc. · 2018
- Descendant · 3 generations on · 4 of its ownK222684 XIA® 4.5 Spinal System, XIA® 4.5 Cortical Trajectory, XIA® 3 Spinal…Stryker Spine · 2023
- Descendant · cites it directly · 1 of its ownK242361 OZARK Cervical Plate System; PYRENEES® Cervical Plate System…Stryker Spine · 2024
- Descendant · 4 generations on · 1 of its ownK240872 Erisma® Deformity Spinal System; Erisma® Lp Spinal Fixation SystemClariance · 2024
Show all 21 generations
- Root · 21 generations backK853033 Hartshill Rect & Wires, Spinal Fixation SystemA.W. Showell (Surgicraft), Ltd. · 1985
- 20 generations backK965221 Ransford Cervical Fixation SystemSurgicraft, Ltd. · 1997
- 19 generations backK982322 Synthes Spine Occipital-Cervical Plate/Rod and Hook SystemSynthes (Usa) · 1998
- 18 generations backK002733 Summit Occipito-Cervico-Thoracic (Oct) Spinal SystemDepuyacromed · 2000
- 17 generations backK013222 Summit Occipito-Cervico- Thoracic (Oct) Spinal SystemDepuyacromed · 2001
- 16 generations backK022190 Modification to Summit Occipito-Cervico-Thoracic (Oct) Spinal SystemDepuyacromed · 2002
- 15 generations backK030103 Summit Oct Spinal System Minipolyaxial ScrewsDepuyacromed · 2003
- 14 generations backK041203 Summit Oct Spinal SystemDepuy Spine, Inc. · 2004
- 13 generations backK062174 Modification to Expedium Spine SystemDepuy Spine, Inc. · 2006
- 12 generations backK063156 Modification to Expedium Spine SystemDepuy Spine, Inc. · 2006
- 11 generations backK063741 Expedium Spine SystemDepuy Spine, Inc. · 2007
- 10 generations backK080313 Modification to Expedium Spine SystemDepuy Spine, Inc. · 2008
- 9 generations backK111571 Viper SystemsJohnson & Johnson · 2011
- 8 generations backK121016 Range Spinal System (Mesa and Denali)K2m, Inc. · 2012
- 7 generations backK130330 Range Spinal SystemK2m, Inc. · 2013
- 6 generations backK131784 Range Spinal SystemK2m, Inc. · 2013
- 5 generations backK133944 Range, Spinal System, Everest Spinal SystemK2m, Inc. · 2014
- 4 generations backK141873 Range/Mesa/Denali Spinal SystemK2m, Inc. · 2014
- 3 generations backK143334 Range/Denali/Mesa Spinal SystemK2m, Inc. · 2015
- 2 generations backK150325 Range/Denali/Mesa Spinal SystemK2m, Inc. · 2015
- 1 generation backK151727 Everest Spinal SystemK2m, Inc. · 2015
- This deviceK161369 EVEREST Spinal SystemK2m, Incorporated · 2016
Showing 5 of 9 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 9 · See all- K181188EVEREST Spinal SystemK2m, Inc. · 2018 · cites it directly6descendants
- K181603EVEREST Spinal System, RANGE (MESA and DENALI) Spinal System, CASPIAN OCT (MESA Mini and…K2m, Inc. · 2018 · 2 generations on5descendants
- K222684XIA® 4.5 Spinal System, XIA® 4.5 Cortical Trajectory, XIA® 3 Spinal System, Serrato®…Stryker Spine · 2023 · 3 generations on4descendants
- K242361OZARK Cervical Plate System; PYRENEES® Cervical Plate System; EVEREST® Spinal System…Stryker Spine · 2024 · cites it directly1descendant
- K240872Erisma® Deformity Spinal System; Erisma® Lp Spinal Fixation SystemClariance · 2024 · 4 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −21 | K853033 | Hartshill Rect & Wires, Spinal Fixation System | A.W. Showell (Surgicraft), Ltd. | 1985 | — |
| −20 | K965221 | Ransford Cervical Fixation System | Surgicraft, Ltd. | 1997 | — |
| −19 | K982322 | Synthes Spine Occipital-Cervical Plate/Rod and Hook System | Synthes (Usa) | 1998 | — |
| −18 | K002733 | Summit Occipito-Cervico-Thoracic (Oct) Spinal System | Depuyacromed | 2000 | — |
| −17 | K013222 | Summit Occipito-Cervico- Thoracic (Oct) Spinal System | Depuyacromed | 2001 | — |
| −16 | K022190 | Modification to Summit Occipito-Cervico-Thoracic (Oct) Spinal System | Depuyacromed | 2002 | — |
| −15 | K030103 | Summit Oct Spinal System Minipolyaxial Screws | Depuyacromed | 2003 | — |
| −14 | K041203 | Summit Oct Spinal System | Depuy Spine, Inc. | 2004 | — |
| −13 | K062174 | Modification to Expedium Spine System | Depuy Spine, Inc. | 2006 | — |
| −12 | K063156 | Modification to Expedium Spine System | Depuy Spine, Inc. | 2006 | — |
| −11 | K063741 | Expedium Spine System | Depuy Spine, Inc. | 2007 | — |
| −10 | K080313 | Modification to Expedium Spine System | Depuy Spine, Inc. | 2008 | — |
| −9 | K111571 | Viper Systems | Johnson & Johnson | 2011 | — |
| −8 | K121016 | Range Spinal System (Mesa and Denali) | K2m, Inc. | 2012 | — |
| −7 | K130330 | Range Spinal System | K2m, Inc. | 2013 | — |
| −6 | K131784 | Range Spinal System | K2m, Inc. | 2013 | — |
| −5 | K133944 | Range, Spinal System, Everest Spinal System | K2m, Inc. | 2014 | — |
| −4 | K141873 | Range/Mesa/Denali Spinal System | K2m, Inc. | 2014 | — |
| −3 | K143334 | Range/Denali/Mesa Spinal System | K2m, Inc. | 2015 | — |
| −2 | K150325 | Range/Denali/Mesa Spinal System | K2m, Inc. | 2015 | — |
| −1 | K151727 | Everest Spinal System | K2m, Inc. | 2015 | — |
| 0 (this device) | K161369 | EVEREST Spinal System | K2m, Incorporated | 2016 | — |
| +1 | K181188 | EVEREST Spinal System | K2m, Inc. | 2018 | — |
| +2 | K181603 | EVEREST Spinal System, RANGE (MESA and DENALI) Spinal System, CASPIAN OCT… | K2m, Inc. | 2018 | — |
| +3 | K222684 | XIA® 4.5 Spinal System, XIA® 4.5 Cortical Trajectory, XIA® 3 Spinal System… | Stryker Spine | 2023 | — |
| +1 | K242361 | OZARK Cervical Plate System; PYRENEES® Cervical Plate System; EVEREST® Spinal… | Stryker Spine | 2024 | — |
| +4 | K240872 | Erisma® Deformity Spinal System; Erisma® Lp Spinal Fixation System | Clariance | 2024 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1978 | 1 | 0 |
| 1984 | 1 | 0 |
| 1990 | 3 | 0 |
| 1992 | 5 | 0 |
| 1993 | 3 | 0 |
| 1994 | 2 | 0 |
| 1995 | 7 | 0 |
| 1996 | 8 | 0 |
| 1997 | 4 | 0 |
| 1998 | 13 | 0 |
| 1999 | 12 | 0 |
| 2000 | 18 | 0 |
| 2001 | 28 | 0 |
| 2002 | 18 | 0 |
| 2003 | 20 | 0 |
| 2004 | 42 | 0 |
| 2005 | 30 | 0 |
| 2006 | 18 | 0 |
| 2007 | 15 | 0 |
| 2008 | 18 | 0 |
| 2009 | 14 | 0 |
| 2010 | 12 | 0 |
| 2011 | 16 | 0 |
| 2012 | 9 | 0 |
| 2013 | 8 | 0 |
| 2014 | 1 | 0 |
| 2015 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2017 | 1 | 0 |
| 2018 | 2 | 0 |
| 2020 | 1 | 0 |
| 2023 | 1 | 0 |
| 2024 | 3 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NKB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NKB device
- #1 K223494 CD Horizon Spinal Systemscore 40.4 · 30 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K151727 | Everest Spinal SystemSummary page 4…CFR 888.3060, Product Code KWQ) Spinal Interlaminal fixation Orthosis (21 CFR 888.3050, Product Code KWP) Predicate Device(s) Primary Predicate EVEREST Spinal System (K151727) Secondary Predicate RANGE /DENALI/MESA Spinal Systems (K141873) Device Description The EVEREST Spinal System is a top-loading… | 2015 |
| * | K141873 | Range/Mesa/Denali Spinal SystemSummary page 4…hosis (21 CFR 888.3050, Product Code KWP) Predicate Device(s) Primary Predicate EVEREST Spinal System (K151727) Secondary Predicate RANGE /DENALI/MESA Spinal Systems (K141873) Device Description The EVEREST Spinal System is a top-loading, multiple component, posterior (thoracic-lumbar) spinal fixation… | 2014 |
What descends from it? (9)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 31 years and passes through 8 companies; a new generation was cleared every 10 months on median.
- 33 of the 349 ancestors have a recall of any class.
- Of its descendants, K181188 (K2m, Inc., 2018) has gone furthest, with 6 descendants of its own.
- 2 other devices also cite its predicate K151727.
How does its category compare?
All NKB devicesNKB Thoracolumbosacral Pedicle Screw System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · NKB
24in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NKB
61in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NKB
1in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · NKB
1,171in 2025
MAUDE reports naming product code NKB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 38 | 87 | 4 | 34 |
| 2008 | 36 | 59 | 5 | 82 |
| 2009 | 44 | 71 | 33 | 171 |
| 2010 | 37 | 111 | 4 | 462 |
| 2011 | 48 | 84 | 3 | 577 |
| 2012 | 60 | 79 | 13 | 1,747 |
| 2013 | 35 | 73 | 0 | 1,962 |
| 2014 | 46 | 90 | 3 (1) | 1,671 |
| 2015 | 55 | 90 | 3 | 1,446 |
| 2016 | 50 | 84 | 1 | 1,585 |
| 2017 | 63 | 69 | 1 | 1,618 |
| 2018 | 56 | 61 | 3 | 1,684 |
| 2019 | 39 | 84 | 1 | 1,915 |
| 2020 | 48 | 68 | 1 | 2,076 |
| 2021 | 32 | 61 | 0 | 1,745 |
| 2022 | 34 | 88 | 5 | 1,339 |
| 2023 | 44 | 72 | 2 | 919 |
| 2024 | 39 | 57 | 3 | 1,614 |
| 2025 | 24 | 61 | 1 | 1,171 |
| 2026 | 30 | 91 | 3 | — |
What is it indicated for?
The EVEREST Spinal System may be used in conjunction with the RANGE® (MESA® and DENALI®) Spinal Systems, all of which are cleared for the following indications: Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation)…
Read the full statement (154 words)
The EVEREST Spinal System may be used in conjunction with the RANGE® (MESA® and DENALI®) Spinal Systems, all of which are cleared for the following indications:
Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system the EVEREST Spinal System may also be used for the same indications as an adjunct to fusion.
When used for posterior non-cervical pedicle screw fixation in pediatric patients the EVEREST Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K161369 · printed Nov 5, 2026
Cite this page
510(k) Report. “K161369: EVEREST Spinal System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K161369
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026