510(k) Report

510(k) · Cleared

Cleared

Medline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use with Chemotherapy Dr)

K161473510(k) TraditionalCleared Oct 19, 2016145-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medline Industries, Inc. · LZA Polymer Patient Examination Glove

Cleared on Oct 19, 2016 as substantially equivalent to K141623 (Eco Medi Glove Sdn. Bhd., 2014) after a 145-day review. Its family tree reaches back 3 generations to 2002.

Decided
Oct 19, 2016
Review
145 days
Class
Class I
Product code
LZA
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K022765 (2002). No later device cites it as a predicate yet.

  1. Root · 3 generations backK022765 Nitrile Powder-Free Examination GlovesAllegiance Healthcare Corp. · 2002
  2. 2 generations backK110921 Nitrile Powder-Free Examination Gloves Tested for Use with…Cardinal Health, Inc. · 2011
  3. 1 generation backK141623 Emg Blue Nitrile Medical Examination Glove with Chemotherapy Drug…Eco Medi Glove Sdn. Bhd. · 2014
  4. This deviceK161473 Medline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use…Medline Industries, Inc. · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K022765Nitrile Powder-Free Examination GlovesAllegiance Healthcare Corp.2002—
−2K110921Nitrile Powder-Free Examination Gloves Tested for Use with Chemotherapy DrugsCardinal Health, Inc.2011—
−1K141623Emg Blue Nitrile Medical Examination Glove with Chemotherapy Drug Labeling ClaimEco Medi Glove Sdn. Bhd.2014—
0 (this device)K161473Medline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use with…Medline Industries, Inc.2016—
FIG. 1 — Lineage of K161473, 2002–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZA device

  1. #1 K192333 Blue Nitrile Examination Gloves Powder Freescore 71.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141623Emg Blue Nitrile Medical Examination Glove with Chemotherapy Drug Labeling Claim
Summary page 4
…Hospital Regulatory Class: I Regulation #: 21 CFR 880.6250 Predicate Device EMG Blue Nitrile Examination Gloves Powder Free with Tested for Use with Chemotherapy Drugs K141623 A1-001 Medline Industries, Inc.…
Eco Medi Glove Sdn. Bhd.2014—

Also cleared against K141623: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 3.5 years on median.
  • None of the 3 ancestors has a recall.
  • 2 other devices also cite its predicate K141623.

How does its category compare?

All LZA devices

LZA Polymer Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · LZA

24in 2025

2007: 212008: 182009: 282010: 312011: 262012: 332013: 202014: 272015: 202016: 302017: 442018: 222019: 252020: 302021: 1392022: 1462023: 622024: 302025: 242026 (partial year): 15
2007high 146 (2022) · low 18 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZA

110in 2025

2007: 95 days median2008: 97 days median2009: 112 days median2010: 97 days median2011: 122 days median2012: 110 days median2013: 130 days median2014: 197 days median2015: 112 days median2016: 171 days median2017: 124 days median2018: 178 days median2019: 212 days median2020: 101 days median2021: 111 days median2022: 111 days median2023: 92 days median2024: 89 days median2025: 110 days median2026 (partial year): 103 days median
2007high 212 (2019) · low 89 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZA

3so far in 2026, the most since 2007

2015: 12016: 12026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZA

73in 2025

2007: 72008: 22009: 122010: 122011: 92012: 112013: 102014: 82015: 122016: 252017: 42018: 32019: 212020: 42021: 202022: 192023: 242024: 762025: 73
2007high 76 (2024) · low 2 (2008)2025

MAUDE reports naming product code LZA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007219507
2008189702
200928112012
20103197012
20112612209
201233110011
201320130010
20142719708
201520112112
201630171125
20174412404
20182217803
201925212021
20203010104
2021139111020
2022146111019
20236292024
20243089076
202524110073
2026151033—

What is it indicated for?

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The following chemicals have been tested with these gloves. Chemotherapy…

Read the full statement (199 words)

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemicals have been tested with these gloves.

Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes

Carmustine 3.3mg/ml - 40.4 min.

Cisplatin 1.0mg/ml - No breakthrough up to 240 min.

Cyclophosphamide 20.0mg.ml - No breakthrough up to 240 min.

Dacarbazine 2.0mg/ml - No breakthrough up to 240 min.

Doxorubicin Hydrochloride 2/0mg.ml - No breakthrough up to 240 min.

Etoposide 20.0mg/ml - No breakthrough up to 240 min.

Fluoruouracil 50mg/ml - No breakthrough up to 240 min.

Methotrexate 25mg - No breakthrough up to 240 min.

Mitoxantrone 2.0mg/ml - No breakthrough up to 240 min.

Paclitaxel 5.0mg/ml - No breakthrough up to 240 min.

Thiotepa 10mg/ml - 60.6 min.

Please note that the following drugs have extremely low permeation times: Carmustine (3.3 mg/ml) has a minimum breakthrough time of 40.4 minutes; Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 60.6 minutes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161473: Medline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use with Chemotherapy Dr).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161473

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026