# K161531: IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement Module X2; IntelliVue Patient Monitors MP20, MP30, MP40, MP50, MP60; IntelliVue Patient Monitors MP70, MP80, MP90;IntelliVue Patient Monitors MX400, MX430, MX450, MX500, MX550, MX600;IntelliVue Patient Monitors MX700, MX800

- Pathway: 510(k) (Special)
- Applicant: Philips Medizin Systeme Boeblingen GmbH
- Decided: Jul 1, 2016 after 29 days
- Product code: MHX (Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms))
- Device class: 2
- Regulation: 21 CFR 870.1025
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 36 generations back to K941811 (1994). 261 ancestors in all.
31 later devices descend from it; 11 cite it directly.
1 of 293 devices in its family (counting itself) has a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K150975: Philips IntelliVue MP2/MP5/MPSC Patient Monitors, Philips IntelliVue X1 and X2 multi-msrmt-modules for use with IntelliVue patient monitors, Philips Medizin Systeme Boeblingen GmbH (2015) Summary page 10.
- K150310: IntelliVue Patient Monitor, Philips Medizin Systeme Boeblingen GmbH (2015) Summary page 10.
- K151681: IntelliVue Patient Monitor, Philips Medizin Systeme Boeblingen GmbH (2015) Summary page 10.

## Most influential descendants
- K171801: IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3; intelliVue Hemodynamic Extension, Capnography Extension, Microstream Extension;Transpac IV Dual IBP Cable (2017), 8 descendants of its own
- K221704: HemoSphere Advanced Monitor, Right Ventricular Pressure (RVP), Global Hypoperfusion Index (GHI) (2022), 6 descendants of its own
- K210906: IntelliVue Patient Monitor MX750, IntelliVue Patient Monitor MX850, IntelliVue 4-Slot module rack FMX-4 (2021), 6 descendants of its own
- K181314: IntelliVue Patient Monitor MX100, IntelliVue Multi-Measurement Module X3, IntelliVue Microstream Extension (2018), 6 descendants of its own
- K232294: HemoSphere Alta Advanced Monitoring Platform; HemoSphere Alta™ Advanced Monitor (Smart Recovery); HemoSphere Alta™ Advanced Monitor (Cardiac); HemoSphere Alta™ Advanced Monitor (All-on-One) (2023), 5 descendants of its own

## Recalls of this device
- Jul 26, 2016 · Class II · Z-2328-2016: If an affected Patient Monitor has been powered on continuously for several months, any displayed waveforms will contain outdated data and therefore fail to reflect the patients current condition.

## Indications for use
The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients. The monitor is intended to be used for monitoring and recording of, and to generate alarms, for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment. The monitor is also intended for use during patient transport inside and outside of the hospital environment. The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment. The MP5, MP5SC and MP5T monitors are also intended for use during patient transport inside the hospital environment; only the MP5 monitor is for use during patient transport outside of the hospital environment. The MP5, MP5SC and MP5T when used with the TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment. The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment. The SSC Sepsis Protocol, in the Protocol Watch clinical decision support tool, is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients. The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment. The MX400/MX430/MX450/MX500/MX550 monitors are additionally intended for use in transport situations within hospital environments. The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only. BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation. The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients. The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients. The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment. The MP20/MP30/MP40/MP50 monitors are additionally intended for use in transport situations within hospital environments. The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients. The transcutaneous gas measurement (tcGas) is restricted to neonatal patients only. BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation. The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients. The device is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients. The device is intended to be used for monitoring and recording of, and to generate alarms, for, multiple physiological parameters of adults, pediatrics, and neonates. The multi measurement module is intended for use by trained healthcare professionals in a hospital environment. The device is also intended for use during patient transport inside and outside of the hospital environment. The device is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The device is for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

Source page: https://510kreport.com/device/K161531
Methodology: https://510kreport.com/methodology