510(k) Report

510(k) · Cleared

Cleared

Immunalysis Barbiturates Urine Enzyme Immunoassay, Immunalysis Multi-Drug Calibrators

K161714510(k) TraditionalCleared Oct 14, 2016115-day reviewClass II21 CFR 862.3150

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Immunalysis Corporation · DIS Enzyme Immunoassay, Barbiturate

Cleared on Oct 14, 2016 as substantially equivalent to K955928 (Diagnostic Reagents, Inc., 1996) after a 115-day review. Its family tree reaches back 1 generation to 1996. 2 later devices descend from it.

Decided
Oct 14, 2016
Review
115 days
Class
Class II
Product code
DIS
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K955928 (1996). 2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2019: 1 descendant cleared2020: 1 descendant clearedK955928 · Barbiturate Eia Essay (Modification) · Diagnostic Reagents, Inc. · 19961995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant K955928 · Barbiturate Eia Essay (Modification) · Diagnostic Reagents, Inc. · 19960
  1. Root · 1 generation backK955928 Barbiturate Eia Essay (Modification)Diagnostic Reagents, Inc. · 1996
  2. This deviceK161714 Immunalysis Barbiturates Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation · 2016
  3. Descendant · cites it directly · 1 of its ownK182719 Quidel Triage TOX Drug Screen, 94600; Quidel Triage® MeterProQuidel Cardiovascular, Inc. · 2019
  4. Descendant · 2 generations on · 0 of its ownK200363 Quidel Triage® TOX Drug Screen, 94600Quidel Cardiovascular, Inc. · 2020

Most influential descendants

Showing 2 of 2 · See all
  • K182719Quidel Triage TOX Drug Screen, 94600; Quidel Triage® MeterProQuidel Cardiovascular, Inc. · 2019 · cites it directly1descendant
  • K200363Quidel Triage® TOX Drug Screen, 94600Quidel Cardiovascular, Inc. · 2020 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K955928Barbiturate Eia Essay (Modification)Diagnostic Reagents, Inc.1996—
0 (this device)K161714Immunalysis Barbiturates Urine Enzyme Immunoassay, Immunalysis Multi-Drug…Immunalysis Corporation2016—
+1K182719Quidel Triage TOX Drug Screen, 94600; Quidel Triage® MeterProQuidel Cardiovascular, Inc.2019—
+2K200363Quidel Triage® TOX Drug Screen, 94600Quidel Cardiovascular, Inc.2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202010
FIG. 1 — Lineage of K161714, 1996–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DIS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K955928Barbiturate Eia Essay (Modification)
Summary page 4
…DKB 5. Predicate Device: DRI ® Barbiturates EIA Assay 6. Predicate Company: Diagnostic Reagents, Inc. 7. Predicate K Number: K955928 Page 1 of 12 D. Device Description The Immunalysis Barbiturates U…
Diagnostic Reagents, Inc.1996—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K182719Quidel Triage TOX Drug Screen, 94600; Quidel Triage® MeterProQuidel Cardiovascular, Inc.20191

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K182719 (Quidel Cardiovascular, Inc., 2019) has gone furthest, with 1 descendant of its own.

How does its category compare?

All DIS devices

DIS Enzyme Immunoassay, Barbiturate: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · DIS

1in 2016

2007: 32011: 22014: 22015: 32016: 1
2007high 3 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DIS

29in 2015

2007: 331 days median2015: 29 days median
2007high 331 (2007) · low 29 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DIS

1in 2016

2011: 12012: 12016: 1
2007high 1 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DIS

1in 2018

2007: 12008: 12018: 1
2007high 1 (2007) · low 0 (2009)2025

MAUDE reports naming product code DIS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007333101
20080—01
20090—00
20100—00
20112—10
20120—10
20130—00
20142—00
201532900
20161—10
20170—00
20180—01
20190—00
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

Immunalysis Barbiturates Urine Enzyme Immunoassay The Immunalysis Barbiturates Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 200 ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Barbiturates in human urine with automated clinical chemistry analyzers. This assay is calibrated against Secobarbital. This in vitro diagnostic device is for prescription use only. The…

Read the full statement (206 words)

Immunalysis Barbiturates Urine Enzyme Immunoassay

The Immunalysis Barbiturates Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 200 ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Barbiturates in human urine with automated clinical chemistry analyzers. This assay is calibrated against Secobarbital. This in vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography/ Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis Barbiturates Urine Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS or LC-MS/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Immunalysis Multi-Drug Calibrators:

The Immunalysis Multi-Drug Calibrators are intended for in vitro diagnostic use for the calibration of assays for the following analytes: Benzoylecgonine, Methamphetamine, Morphine, PCP, Secobarbital and Oxazepam. The calibrators are designed for prescription use with immunoassays.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161714: Immunalysis Barbiturates Urine Enzyme Immunoassay, Immunalysis Multi-Drug Calibrators.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161714

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026