# K161818: GEM Premier 3000

- Pathway: 510(k) (Special)
- Applicant: Instrumentation Laboratory CO
- Decided: Sep 14, 2016 after 75 days
- Product code: CGA (Glucose Oxidase, Glucose)
- Device class: 2
- Regulation: 21 CFR 862.1345
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 5 generations back to K801848 (1980). 10 ancestors in all.
No later device cites it as a predicate yet.
0 of 11 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K052121: Gem Premier 3000, Instrumentation Laboratory CO (2005) Summary page 4.

## Recalls of this device
- Sep 29, 2023 · Class II · Z-0357-2024: Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside
- Sep 29, 2023 · Class II · Z-0358-2024: Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside
- Sep 29, 2023 · Class II · Z-0359-2024: Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside

## Indications for use
The GEM Premier 3000 is a portable system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting. The instrument provides quantitative measurements of whole blood pH, pCO2, pO2, Na+, K+, Ca++, Glucose, Lactate and Hct. These parameters along with derived parameters aid in the diagnosis of a patient's acid/base status, oxygen delivery capacity, and electrolyte and metabolite balance.

Source page: https://510kreport.com/device/K161818
Methodology: https://510kreport.com/methodology