510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

PhotonBlade

K162053510(k) TraditionalCleared Sep 15, 201652-day reviewClass II21 CFR 878.4400Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Invuity, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Sep 15, 2016 as substantially equivalent to K093695 (Peak Surgical, Inc., 2010) and 1 other predicate after a 52-day review. Its family tree reaches back 5 generations to 1994. 1 later device descends from it. It has been recalled once.

Decided
Sep 15, 2016
Review
52 days
Class
Class II
Product code
GEI
Recalls
1
Predicates
2

Its longest line of descent runs back 5 generations to K933157 (1994). 22 ancestors in all, the oldest K780977 (1978). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1994: 2 other ancestors2008: 2 other ancestors2005: 1 other ancestor1998: 1 other ancestor1996: 3 other ancestors1979: 1 other ancestor2004: 1 other ancestor2000: 1 other ancestor1995: 2 other ancestors1991: 2 other ancestors1978: 1 other ancestor2020: 1 descendant clearedK933157 · Erbotom Icc 200 · Erbe USA, Inc. · 1994K013255 · Surgitron 120 Iec (Also Known as Surgitron 4.0 Dual RF · Ellman Intl., Inc. · 2001K061174 · Surgi-Max · Ellman Int'L, Inc. · 2006K082786 · Pulsar Generator, Peak Plasmablade; 4.0, Needle, and Ext Tissue Dissection Devices · Peak Surgical, Inc. · 2008 · recalledK093695 · Peak Plasmablade 3.OS Tissue Dissection Device · Peak Surgical, Inc. · 201019801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 7 other ancestors5 generations back: 7 other ancestors 1 generation forward: 1 descendant K093695 · Peak Plasmablade 3.OS Tissue Dissection Device · Peak Surgical, Inc. · 2010K082786 · Pulsar Generator, Peak Plasmablade; 4.0, Needle, and Ext Tissue Dissection Devices · Peak Surgical, Inc. · 2008 · recalledK061174 · Surgi-Max · Ellman Int'L, Inc. · 2006K013255 · Surgitron 120 Iec (Also Known as Surgitron 4.0 Dual RF · Ellman Intl., Inc. · 2001K933157 · Erbotom Icc 200 · Erbe USA, Inc. · 1994−50
  1. Root · 5 generations backK933157 Erbotom Icc 200Erbe USA, Inc. · 1994
  2. 4 generations backK013255 Surgitron 120 Iec (Also Known as Surgitron 4.0 Dual RFEllman Intl., Inc. · 2001
  3. 3 generations backK061174 Surgi-MaxEllman Int'L, Inc. · 2006
  4. 2 generations backK082786 Pulsar Generator, Peak Plasmablade; 4.0, Needle, and Ext Tissue…Peak Surgical, Inc. · 2008 · Recalled
  5. 1 generation backK093695 Peak Plasmablade 3.OS Tissue Dissection DevicePeak Surgical, Inc. · 2010
  6. This deviceK162053 PhotonBladeInvuity, Inc. · 2016
  7. Descendant · cites it directly · 0 of its ownK191583 PhotonBlade with Smoke EvacuationInvuity, Inc. · 2020

Most influential descendants

Showing 1 of 1 · See all
  • K191583PhotonBlade with Smoke EvacuationInvuity, Inc. · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K933157Erbotom Icc 200Erbe USA, Inc.1994—
−4K013255Surgitron 120 Iec (Also Known as Surgitron 4.0 Dual RFEllman Intl., Inc.2001—
−3K061174Surgi-MaxEllman Int'L, Inc.2006—
−2K082786Pulsar Generator, Peak Plasmablade; 4.0, Needle, and Ext Tissue Dissection…Peak Surgical, Inc.20081
−1K093695Peak Plasmablade 3.OS Tissue Dissection DevicePeak Surgical, Inc.2010—
0 (this device)K162053PhotonBladeInvuity, Inc.2016—
+1K191583PhotonBlade with Smoke EvacuationInvuity, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
197910
199120
199420
199520
199630
199810
200010
200410
200510
200820

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
FIG. 1 — Lineage of K162053, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K093695Peak Plasmablade 3.OS Tissue Dissection Device
Summary page 4
…Panel: General and Plastic Surgery Device Classification: Class 11 5.5 Predicate device(s): • PEAK Surgical PlasmaBlade 3.0S (K093695) • PARE NOVA ES Pencil with Surgical Light Source (K944363) 5.6 Device description The PhotonBlade…
Peak Surgical, Inc.2010—
*K944363Model PBD Esu Pencil
Summary page 4
…ication: Class 11 5.5 Predicate device(s): • PEAK Surgical PlasmaBlade 3.0S (K093695) • PARE NOVA ES Pencil with Surgical Light Source (K944363) 5.6 Device description The PhotonBlade is a single use, sterile, electrosurgical device with a light.…
Coral Medical, Inc.1994—

Also cleared against K093695: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191583PhotonBlade with Smoke EvacuationInvuity, Inc.20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 12, 2017Class IIZ-2778-2017Device DesignA small number of customer observations were recently received which related to possible unintended energy discharge with the potential to cause tissue damage to patients.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 5 companies; a new generation was cleared every 4.6 years on median.
  • 5 of the 22 ancestors have a recall of any class.
  • Of its descendants, K191583 (Invuity, Inc., 2020) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K093695.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The PhotonBlade is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of

tissue during general surgical procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162053: PhotonBlade.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162053

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026