# K162227: STA® - Liatest® D-Di

- Pathway: 510(k) (Traditional)
- Applicant: Diagnostica Stago
- Decided: Dec 10, 2016 after 124 days
- Product code: DAP (Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control)
- Device class: 2
- Regulation: 21 CFR 864.7320
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K862156 (1986). 4 ancestors in all.
No later device cites it as a predicate yet.
0 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K040882: Vidas D-Dimer Exclusion Assay, Model 30 442, bioMerieux, Inc. (2004) Summary page 4.

## Recalls of this device
- Feb 5, 2026 · Class II · Z-1585-2026: After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
- Sep 12, 2025 · Class II · Z-0205-2026: A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.

## Indications for use
The STA® - Liatest® D-Di kit is an immuno-turbidimetric assay for the quantitative determination of D-dimer in venous plasma (3.2% sodium citrate) for use on STA-R®, STA Compact® and STA Satellite® analyzers by professional laboratory personnel. The STA® - Liatest® D-Di is intended for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT) in outpatients suspected of PE or DVT.