BeneVision Central Monitoring System

K162607510(k) TraditionalCleared Apr 20, 2017213-day reviewClass II21 CFR 870.2300Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · MSX System, Network And Communication, Physiological Monitors

Cleared on Apr 20, 2017 as substantially equivalent to K150632 (Shenzhen Mindray Bio-Medical Electronics Co., Ltd., 2015) after a 213-day review. Its family tree reaches back 7 generations to 1994. 12 later devices descend from it. It has been recalled once.

Decided
Apr 20, 2017
Review
213 days
Class
Class II
Product code
MSX
Recalls
1
Predicates
1

Its longest line of descent runs back 7 generations to K933143 (1994). 23 ancestors in all, the oldest K781836 (1978). 12 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2001: 1 other ancestor1996: 1 other ancestor1995: 3 other ancestors1994: 2 other ancestors1986: 1 other ancestor1978: 1 other ancestor1997: 1 other ancestor1990: 2 other ancestors1989: 2 other ancestors1991: 1 other ancestor2019: 1 descendant cleared2020: 1 descendant cleared2023: 2 descendants cleared2022: 1 descendant cleared2026: 3 descendants cleared2025: 2 descendants cleared2024: 2 descendants clearedReference device: K152902 · Passport Series Patient Monitors (including Passport 17m, Passport 12m and T1) · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2016 · recalledK933143 · Model Eli 300 Electrocardiograph · Mortara Instrument, Inc. · 1994K962970 · Spacelabs Medical 12-Lead Diagnostic Ecg Analysis Option · Spacelabs Medical, Inc. · 1997K972502 · Spacelabs Medical Integrated Multiparameter Module 90496 · Spacelabs Medical, Inc. · 1998 · recalledK983996 · Spacelabs Medical Ultraview Digital Telemtry System · Spacelabs Medical, Inc. · 1999K080192 · Hypervisor VI Central Monitoring System (Including TMS) · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2008 · recalledK132036 · Hypervisor Central Monitoring System (Including Telemetry Monitoring System, TMS-6016) · Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2014K150632 · Hypervisor IX Monitoring System · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2015 · recalled19801990200020102020
  1. Root · 7 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 6 generations backK962970 Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc. · 1997
  3. 5 generations backK972502 Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc. · 1998 · Recalled
  4. 4 generations backK983996 Spacelabs Medical Ultraview Digital Telemtry SystemSpacelabs Medical, Inc. · 1999
  5. 3 generations backK080192 Hypervisor VI Central Monitoring System (Including TMS)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2008 · Recalled
  6. 2 generations backK132036 Hypervisor Central Monitoring System (Including Telemetry Monitoring…Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2014
  7. 1 generation backK150632 Hypervisor IX Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2015 · Recalled
  8. This deviceK162607 BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2017
  9. Descendant · cites it directly · 11 of its ownK183238 BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2019
  10. Descendant · 2 generations on · 10 of its ownK193391 BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2020 · Recalled
  11. Descendant · 3 generations on · 3 of its ownK220058 BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2022 · Recalled
  12. Descendant · 3 generations on · 2 of its ownK223567 CNS-2101 Central MonitorNihon Kohden Corporation · 2023
  13. Descendant · 4 generations on · 1 of its ownK242750 Central StationNihon Kohden Digital Health Solutions, LLC · 2025

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2019: 12020: 12022: 12023: 22024: 22025: 22026: 3
2019high 3 (2026) · low 0 (2021)2026

Most influential descendants

Showing 5 of 12 · See all
  • K183238BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2019 · cites it directly11descendants
  • K193391BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2020 · 2 generations on10descendants
  • K220058BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2022 · 3 generations on3descendants
  • K223567CNS-2101 Central MonitorNihon Kohden Corporation · 2023 · 3 generations on2descendants
  • K242750Central StationNihon Kohden Digital Health Solutions, LLC · 2025 · 4 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K933143Model Eli 300 ElectrocardiographMortara Instrument, Inc.1994—
−6K962970Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc.1997—
−5K972502Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc.19983
−4K983996Spacelabs Medical Ultraview Digital Telemtry SystemSpacelabs Medical, Inc.1999—
−3K080192Hypervisor VI Central Monitoring System (Including TMS)Shenzhen Mindray Bio-Medical Electronics Co., Ltd.20082
−2K132036Hypervisor Central Monitoring System (Including Telemetry Monitoring System…Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd.2014—
−1K150632Hypervisor IX Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.20152
0 (this device)K162607BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.2017—
+1K183238BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.2019—
+2K193391BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.20202
+3K220058BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.20221
+3K223567CNS-2101 Central MonitorNihon Kohden Corporation2023—
+4K242750Central StationNihon Kohden Digital Health Solutions, LLC2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
198610
198920
199020
199110
199420
199530
199610
199710
200110
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202010
202210
202320
202420
202520
202630
FIG. 1 — Lineage of K162607, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MSX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSX device

  1. #1 K242750 Central Stationscore 21.5 · 31 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K150632Hypervisor IX Monitoring System
Summary page 6
…vasive DQA- Oximeter DSB- Impedance plethysmograph Predicate Device: K150632 –Hypervisor IX Monitoring System Reference Predicates: K152902–Passport Series Patient Monitors (Passport 12m,…
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.20152
†K152902Passport Series Patient Monitors (including Passport 17m, Passport 12m and T1)Shenzhen Mindray Bio-Medical Electronics Co., Ltd.20161

Also cleared against K150632: 1 other device →

What descends from it? (12)

See all 12

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K183238BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.201911

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 30, 2018Class IIZ-0170-2019Software designThe firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 1.8 years on median.
  • 5 of the 23 ancestors have a recall of any class.
  • Of its descendants, K183238 (Shenzhen Mindray Bio-Medical Electronics Co., Ltd., 2019) has gone furthest, with 11 descendants of its own.
  • 1 other device also cites its predicate K150632.

How does its category compare?

All MSX devices

MSX System, Network And Communication, Physiological Monitors: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · MSX

4in 2025

2007: 42008: 12009: 12010: 32011: 72012: 42013: 22014: 32015: 52016: 32017: 22018: 42019: 42020: 52021: 32022: 52023: 32024: 42025: 42026 (partial year): 5
2007high 7 (2011) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSX

295in 2025

2007: 58 days median2010: 43 days median2011: 72 days median2012: 145 days median2014: 165 days median2015: 138 days median2016: 174 days median2018: 111 days median2019: 92 days median2020: 193 days median2021: 227 days median2022: 177 days median2023: 244 days median2024: 212 days median2025: 295 days median2026 (partial year): 145 days median
2007high 295 (2025) · low 43 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSX

1in 2025

2007: 12008: 12009: 22010: 12011: 12012: 22015: 22017: 22018: 12022: 12025: 12026 (partial year): 2
2007high 2 (2009) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MSX

19in 2025

2007: 52008: 62009: 72010: 142011: 32012: 42013: 172014: 322015: 312016: 112017: 162018: 152019: 152020: 252021: 172022: 182023: 102024: 322025: 19
2007high 32 (2014) · low 3 (2011)2025

MAUDE reports naming product code MSX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200745815
20081—16
20091—27
2010343114
201177213
2012414524
20132—017
20143165032
20155138231
20163174011
20172—216
20184111115
2019492015
20205193025
20213227017
20225177118
20233244010
20244212032
20254295119
202651452—

What is it indicated for?

The indications for use of the BeneVision Central Monitoring System include: • Real time viewing of patient clinical data and alarms • Storage and Historical review of patient clinical data and alarms • Printing of real time and historical patient data • Configuration of local settings as well as synchronizing settings across the network to a remote device • Transfer of patient clinical data and settings between several CentralStations The BeneVision…

Read the full statement (392 words)

The indications for use of the BeneVision Central Monitoring System include:

• Real time viewing of patient clinical data and alarms

• Storage and Historical review of patient clinical data and alarms

• Printing of real time and historical patient data

• Configuration of local settings as well as synchronizing settings across the network to a remote device

• Transfer of patient clinical data and settings between several CentralStations

The BeneVision Central Monitoring System is a networked patient monitoring system intended for use in a fixed location, installed in professional healthcare facilities to provide clinicians remote patient monitoring. The network connections between the various devices can be any combination of Ethernet (Wired), Wireless WIFI (WLAN), and Wireless WMTS.

The BeneVision Central Monitoring System supports one or more Mindray compatible physiological monitors and will display, store, print, and transfer information received from the compatible monitors; The BeneVision Central Monitoring System supports bi-directional configuration of the compatible monitors. No data processing is done by the BeneVision Central Monitoring System for data received from compatible monitors.

The telemetry monitoring systems are designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The BeneVision Central Monitoring System supports Telemetry Systems: TMS-6016, Telepack-608, TMS60, and TM80.

• The TMS-6016 transmitter is intended for use on Adult and Pediatric patients to monitor ECG and SpO2 physiological data.

• The Panorama Telepack-608 transmitter is intended for use on Adult patients to monitor ECG and SpO2 physiological data.

• The TMS60 transmitter is intended for use on Adult and Pediatric patients over three years old to monitor ECG, SpO2, NIBP and Resp physiological data. The physiological data can be reviewed locally on the display of the transmitter. The CentralStation will support ECG, Heart Rate, SpO2, NIBP, Resp, Pulse Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TMS60.

• The TM80 telemetry monitor is intended for use on Adult and Pediatric patients over three years old to monitor ECG, SpO2, NIBP and Resp physiological data. The physiological data can be reviewed locally on the display of the monitor. The CentralStation will support ECG, Heart Rate, SpO2, NIBP, Resp, Pulse Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TM80.

The BeneVision Central Monitoring System is intended for use in professional healthcare facilities under the direct supervision of a licensed healthcare practitioner.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162607: BeneVision Central Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162607

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026