iTotal® Hip Replacement System

K162719510(k) TraditionalCleared Jun 14, 2017258-day reviewClass II21 CFR 888.3358Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Conformis, Inc. · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Cleared on Jun 14, 2017 as substantially equivalent to K112802 (Pipeline Orthopedics, 2012) and 2 other predicates after a 258-day review. Its family tree reaches back 8 generations to 1994. 17 later devices descend from it. It has been recalled 4 times.

Decided
Jun 14, 2017
Review
258 days
Class
Class II
Product code
LPH
Recalls
4
Predicates
3

Its longest line of descent runs back 8 generations to K931641 (1994). 106 ancestors in all, the oldest K810258 (1981). 17 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor2005: 3 other ancestors2011: 2 other ancestors2008: 2 other ancestors2007: 1 other ancestor2006: 1 other ancestor2004: 2 other ancestors2001: 3 other ancestors2000: 8 other ancestors1996: 12 other ancestors2010: 1 other ancestor2003: 3 other ancestors2002: 2 other ancestors1999: 7 other ancestors1998: 5 other ancestors1997: 6 other ancestors1995: 6 other ancestors1994: 15 other ancestors1993: 7 other ancestors1986: 3 other ancestors1992: 3 other ancestors1985: 2 other ancestors1991: 2 other ancestors1981: 1 other ancestor2019: 2 descendants cleared2020: 1 descendant cleared2025: 2 descendants cleared2022: 4 descendants cleared2021: 1 descendant cleared2023: 7 descendants clearedReference device: K160025 · iTotal Posterior Stabilized (PS) Knee Replacement System, iTotal Cruciate Retaining (CR) Knee Replacement System · Conformis, Inc. · 2016 · recalledReference device: K161668 · ConforMIS iTotal (R) Posterior Stabilized (PS) Knee Replacement System (iTotal PS) · Conformis, Inc. · 2016 · recalledK931641 · Porocoat Porous Coating · Depuy, Inc. · 1994K003800 · Aml Hip Prosthesis · DePuy Orthopaedics, Inc. · 2001K012364 · Aml Hip Prosthesis · DePuy Orthopaedics, Inc. · 2001K030236 · Acumatch P-Series Porous Press-Fit Femoral Component · Exactech, Inc. · 2003×2: 2 generations cleared close together, K041906 (2004), K042842 (2004) K102487 · Novation Femoral Stem, Tapered 12/14 , Standard/Extended Offset, Plasma Coated, Press-Fit, Size 19/20 · Exactech, Inc. · 2010K112802 · Pipeline Total Hip System · Pipeline Orthopedics · 201219801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 10 other ancestors3 generations back: 21 other ancestors4 generations back: 28 other ancestors5 generations back: 17 other ancestors6 generations back: 19 other ancestors7 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 6 descendants3 generations forward: 2 descendants4 generations forward: 4 descendants5 generations forward: 2 descendants K112802 · Pipeline Total Hip System · Pipeline Orthopedics · 2012K102487 · Novation Femoral Stem, Tapered 12/14 , Standard/Extended Offset, Plasma Coated, Press-Fit, Size 19/20 · Exactech, Inc. · 2010K042842 · Novation 12/14 Press-Fit Femoral Stem · Exactech, Inc. · 2004K041906 · Exactech 12/14 Total Hip System · Exactech, Inc. · 2004 · recalledK030236 · Acumatch P-Series Porous Press-Fit Femoral Component · Exactech, Inc. · 2003K012364 · Aml Hip Prosthesis · DePuy Orthopaedics, Inc. · 2001K003800 · Aml Hip Prosthesis · DePuy Orthopaedics, Inc. · 2001K931641 · Porocoat Porous Coating · Depuy, Inc. · 1994−50+5
  1. Root · 8 generations backK931641 Porocoat Porous CoatingDepuy, Inc. · 1994
  2. 7 generations backK003800 Aml Hip ProsthesisDePuy Orthopaedics, Inc. · 2001
  3. 6 generations backK012364 Aml Hip ProsthesisDePuy Orthopaedics, Inc. · 2001
  4. 5 generations backK030236 Acumatch P-Series Porous Press-Fit Femoral ComponentExactech, Inc. · 2003
  5. 4 generations backK041906 Exactech 12/14 Total Hip SystemExactech, Inc. · 2004 · Recalled
  6. 3 generations backK042842 Novation 12/14 Press-Fit Femoral StemExactech, Inc. · 2004
  7. 2 generations backK102487 Novation Femoral Stem, Tapered 12/14 , Standard/Extended Offset…Exactech, Inc. · 2010
  8. 1 generation backK112802 Pipeline Total Hip SystemPipeline Orthopedics · 2012
  9. This deviceK162719 iTotal® Hip Replacement SystemConformis, Inc. · 2017
  10. Descendant · cites it directly · 12 of its ownK192198 Conformis Hip SystemConformis, Inc. · 2019 · Recalled
  11. Descendant · 2 generations on · 8 of its ownK213389 Identity Imprint Cruciate Retaining (CR) Knee Replacement System and…Conformis, Inc. · 2021
  12. Descendant · 2 generations on · 7 of its ownK221059 Identity Imprint Knee Replacement System (including Identity Imprint…Conformis, Inc. · 2022
  13. Descendant · 3 generations on · 6 of its ownK221404 Identity Imprint Knee Replacement System (including Identity Imprint…Conformis, Inc. · 2022
  14. Descendant · cites it directly · 6 of its ownK190904 Conformis BeneFIT Hip SystemConformis, Inc. · 2019 · Recalled

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2019: 22020: 12021: 12022: 42023: 72025: 2
2019high 7 (2023) · low 0 (2024)2026

Most influential descendants

Showing 5 of 17 · See all
  • K192198Conformis Hip SystemConformis, Inc. · 2019 · cites it directly12descendants
  • K213389Identity Imprint Cruciate Retaining (CR) Knee Replacement System and Identity Imprint…Conformis, Inc. · 2021 · 2 generations on8descendants
  • K221059Identity Imprint Knee Replacement System (including Identity Imprint Posterior…Conformis, Inc. · 2022 · 2 generations on7descendants
  • K221404Identity Imprint Knee Replacement System (including Identity Imprint Posterior…Conformis, Inc. · 2022 · 3 generations on6descendants
  • K190904Conformis BeneFIT Hip SystemConformis, Inc. · 2019 · cites it directly6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K931641Porocoat Porous CoatingDepuy, Inc.1994—
−7K003800Aml Hip ProsthesisDePuy Orthopaedics, Inc.2001—
−6K012364Aml Hip ProsthesisDePuy Orthopaedics, Inc.2001—
−5K030236Acumatch P-Series Porous Press-Fit Femoral ComponentExactech, Inc.2003—
−4K041906Exactech 12/14 Total Hip SystemExactech, Inc.20041
−3K042842Novation 12/14 Press-Fit Femoral StemExactech, Inc.2004—
−2K102487Novation Femoral Stem, Tapered 12/14 , Standard/Extended Offset, Plasma…Exactech, Inc.2010—
−1K112802Pipeline Total Hip SystemPipeline Orthopedics2012—
0 (this device)K162719iTotal® Hip Replacement SystemConformis, Inc.2017—
+1K192198Conformis Hip SystemConformis, Inc.20191
+2K213389Identity Imprint Cruciate Retaining (CR) Knee Replacement System and Identity…Conformis, Inc.2021—
+2K221059Identity Imprint Knee Replacement System (including Identity Imprint Posterior…Conformis, Inc.2022—
+3K221404Identity Imprint Knee Replacement System (including Identity Imprint Posterior…Conformis, Inc.2022—
+1K190904Conformis BeneFIT Hip SystemConformis, Inc.20191

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198520
198630
199120
199230
199370
1994150
199560
1996120
199760
199850
199970
200080
200130
200220
200330
200420
200530
200610
200710
200820
201010
201120
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201920
202010
202110
202240
202370
202520
FIG. 1 — Lineage of K162719, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LPH device

  1. #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112802Pipeline Total Hip System
Summary page 5
…Product Code: LPH, JDI, OQG, OQH Devices): Regulation Number: 21 CFR 888.3358 510(k) Number: K112802 Corail AMT Hip Prosthesis Device Class: II…
Pipeline Orthopedics2012—
*K042992Corail Amt Hip Prosthesis
Summary page 5
…Product Code: LZO Regulation Number: 21 CFR 888.3353 510(k) Number: K042992 Signature Planner, Signature Guides Device Class: II…
DePuy Orthopaedics, Inc.20051
*K111863Signature Planner,Signature Guides
Summary page 5
…Code: LPH, LZO, KWZ, JDI, KWY, MAY MEH Regulation Number: 21 CFR 888.3358 510(k) Number: K111863 Identification of the iTotal® CR & PS Knee Replacement Systems Legally Marketed Device Class:…
Materialise NV2012—
†K160025iTotal Posterior Stabilized (PS) Knee Replacement System, iTotal Cruciate Retaining (CR)…Conformis, Inc.20161
†K161668ConforMIS iTotal (R) Posterior Stabilized (PS) Knee Replacement System (iTotal PS)Conformis, Inc.20164

Also cleared against K112802: 5 other devices →

What descends from it? (17)

See all 17

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192198Conformis Hip SystemConformis, Inc.201912
K190904Conformis BeneFIT Hip SystemConformis, Inc.20196
K202484Cordera Hip SystemConformis, Inc.20202

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 22, 2019Class IIZ-0064-2020Under Investigation by firmTwo femoral stems, sizes 13 and 15, were erroneously switched and packed in each other's packaging.
Aug 22, 2019Class IIZ-0063-2020Under Investigation by firmTwo femoral stems, sizes 13 and 15, were erroneously switched and packed in each other's packaging.
Mar 29, 2019Class IIZ-1775-2019Device DesignDiscrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.
Oct 16, 2018Class IIZ-0731-2019Under Investigation by firmThe stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 1.6 years on median.
  • 16 of the 106 ancestors have a recall of any class.
  • Of its descendants, K192198 (Conformis, Inc., 2019) has gone furthest, with 12 descendants of its own.
  • 5 other devices also cite its predicate K112802.

How does its category compare?

All LPH devices

LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LPH

12in 2025

2007: 102008: 102009: 32010: 152011: 162012: 252013: 112014: 202015: 182016: 102017: 122018: 92019: 222020: 72021: 82022: 142023: 102024: 62025: 122026 (partial year): 8
2007high 25 (2012) · low 3 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LPH

153in 2025

2007: 116 days median2008: 86 days median2009: 112 days median2010: 100 days median2011: 110 days median2012: 76 days median2013: 135 days median2014: 129 days median2015: 111 days median2016: 150 days median2017: 121 days median2018: 145 days median2019: 164 days median2020: 164 days median2021: 410 days median2022: 247 days median2023: 111 days median2024: 92 days median2025: 153 days median2026 (partial year): 256 days median
2007high 410 (2021) · low 76 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LPH

4in 2025

2007: 162008: 12009: 22010: 222011: 262012: 302013: 92014: 32015: 12016: 82017: 142018: 22019: 82020: 62021: 42022: 22023: 22024: 12025: 42026 (partial year): 2
2007high 30 (2012) · low 1 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LPH

6,163in 2025

2007: 5092008: 7062009: 1,1562010: 2,0072011: 3,3202012: 3,7852013: 5,0022014: 5,1632015: 5,8302016: 4,9002017: 5,3122018: 5,9542019: 7,1302020: 5,4292021: 5,9782022: 5,2232023: 5,3452024: 5,6262025: 6,163
2007high 7,130 (2019) · low 509 (2007)2025

MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071011616509
200810861706
2009311221,156
201015100222,007
201116110263,320
20122576303,785
20131113595,002
20142012935,163
20151811115,830
20161015084,900
201712121145,312
2018914525,954
20192216487,130
2020716465,429
2021841045,978
20221424725,223
20231011125,345
202469215,626
20251215346,163
202682562—

What is it indicated for?

The iTotal® Hip Replacement System is designed from a patient’s pre-operative CT scan which must include certain necessary anatomic landmarks that are clearly identifiable. Total hip replacement using the iTotal® Hip Replacement System is indicated for use in skeletally mature individuals undergoing total hip replacement due to: • A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis…

Read the full statement (147 words)

The iTotal® Hip Replacement System is designed from a patient’s pre-operative CT scan which must include certain necessary anatomic landmarks that are clearly identifiable. Total hip replacement using the iTotal® Hip Replacement System is indicated for use in skeletally mature individuals undergoing total hip replacement due to:

• A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia.

• Treatment of non-displaced non-unions of the hip, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.

• Revision procedures for failed previous hip surgery (excluding situations where hardware is present).

The iTotal® Hip Replacement System includes standard hip replacement components as well as the following patient specific components: femoral neck, acetabular cup, single use instrumentation.

The iTotal® Hip Replacement implants are intended for cementless fixation using an anterior or posterior surgical approach.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162719: iTotal® Hip Replacement System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162719

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026