510(k) Report

510(k) · Cleared

Cleared

LIAISON® CMV IgG and LIAISON® CMV IgG Serum Control Set

K162969510(k) SpecialCleared Jan 6, 201774-day reviewClass II21 CFR 866.3175

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

DiaSorin, Inc. · LFZ Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus

Cleared on Jan 6, 2017 as substantially equivalent to K040290 (Diasorin S.P.A, 2005) after a 74-day review. Its family tree reaches back 3 generations to 1992.

Decided
Jan 6, 2017
Review
74 days
Class
Class II
Product code
LFZ
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K915892 (1992). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK915892 CMV Igg Elisa TestGull Laboratories, Inc. · 1992
  2. 2 generations backK955361 Incstar Cytomegalovirus Igg Elisa AssayIncstar Corp. · 1996
  3. 1 generation backK040290 Diasorin Liaison CMV Igm/IggDiasorin S.P.A · 2005
  4. This deviceK162969 LIAISON® CMV IgG and LIAISON® CMV IgG Serum Control SetDiaSorin, Inc. · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K915892CMV Igg Elisa TestGull Laboratories, Inc.1992—
−2K955361Incstar Cytomegalovirus Igg Elisa AssayIncstar Corp.1996—
−1K040290Diasorin Liaison CMV Igm/IggDiasorin S.P.A2005—
0 (this device)K162969LIAISON® CMV IgG and LIAISON® CMV IgG Serum Control SetDiaSorin, Inc.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
FIG. 1 — Lineage of K162969, 1992–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LFZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LFZ device

  1. #1 K220911 Elecsys CMV IgGscore 4.8 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040290Diasorin Liaison CMV Igm/Igg
Summary page 5
…Micorobiology (83) Product Code: LFZ, JJX PREDICATE DEVICE: DiaSorin LIAISON® CMV IgG (K040290) DEVICE DESCRIPTION: The LIAISON® CMV IgG assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer fam…
Diasorin S.P.A2005—

Also cleared against K040290: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 8.9 years on median.
  • None of the 4 ancestors has a recall.
  • 2 other devices also cite its predicate K040290.

How does its category compare?

All LFZ devices

LFZ Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LFZ

2in 2022

2014: 22017: 22018: 12022: 2
2007high 2 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LFZ

Too few LFZ 510(k)s in any year since 2007 to give a median.

Recalls per year · LFZ

1in 2025

2007: 12009: 12022: 12025: 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LFZ

53in 2025

2007: 72008: 72009: 142010: 32011: 72013: 12014: 32015: 72016: 142017: 132018: 92019: 102020: 72021: 232022: 112023: 362024: 342025: 53
2007high 53 (2025) · low 0 (2012)2025

MAUDE reports naming product code LFZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—17
20080—07
20090—114
20100—03
20110—07
20120—00
20130—01
20142—03
20150—07
20160—014
20172—013
20181—09
20190—010
20200—07
20210—023
20222—111
20230—036
20240—034
20250—153
20260—0—

What is it indicated for?

The LIAISON® CMV IgG assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer family* for the qualitative determination of IgG antibodies to human cytomegalovirus (hCMV) in human serum. It is intended to be used as an aid in the determination of serological status to CMV. The DiaSorin LIAISON® CMV IgG Serum Control Set is intended for use as assayed quality control samples to monitor the performance of the LIAISON® CMYV IgG…

Read the full statement (109 words)

The LIAISON® CMV IgG assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer family* for the qualitative determination of IgG antibodies to human cytomegalovirus (hCMV) in human serum. It is intended to be used as an aid in the determination of serological status to CMV.

The DiaSorin LIAISON® CMV IgG Serum Control Set is intended for use as assayed quality control samples to monitor the performance of the LIAISON® CMYV IgG assay on the LIAISON® Analyzer family*. The performance characteristics of the LIAISON® CMYV IgG controls have not been established for any other assay or instrument platforms different from the LIAISON® and LIAISON® XL.

*(LIAISON® and LIAISON® XL).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K162969: LIAISON® CMV IgG and LIAISON® CMV IgG Serum Control Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K162969

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026