# K162973: EndoWrist Suction Irrigator

- Pathway: 510(k) (Traditional)
- Applicant: Intuitive Surgical, Inc.
- Decided: Feb 6, 2017 after 104 days
- Product code: NAY (System, Surgical, Computer Controlled Instrument)
- Device class: 2
- Regulation: 21 CFR 876.1500
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 3 generations back to K934094 (1994). 5 ancestors in all.
No later device cites it as a predicate yet.
0 of 6 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K110451: Endowrist One Suction/Irrigator, Intuitive Surgical, Inc. (2011) Summary page 4.

## Recalls of this device
- Jun 4, 2018 · Class II · Z-0249-2020: The firm has identified the potential for a silicone particle to be generated within the Suction Irrigator device. If generated, a particle may be introduced into the patient with irrigation.
- Mar 31, 2017 · Class II · Z-1819-2017: Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.

## Indications for use
The EndoWrist® Suction Irrigator is designed to be used in conjunction with an Intuitive Surgical da Vinci Surgical System and compatible suction and irrigation sources and tubing sets for delivering fluid to the surgical site and for evacuation and aspiration of fluids. The instrument may also be used for retraction and blunt dissection of tissue.