510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Orthopantomograph OP300

K163423510(k) TraditionalCleared Aug 31, 2017268-day reviewClass II21 CFR 892.1750Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Palodex Group OY · OAS X-Ray, Tomography, Computed, Dental

Cleared on Aug 31, 2017 as substantially equivalent to K110839 (Soredex Palodex Group OY, 2011) and 1 other predicate after a 268-day review. Its family tree reaches back 8 generations to 1997. 2 later devices descend from it. It has been recalled once.

Decided
Aug 31, 2017
Review
268 days
Class
Class II
Product code
OAS
Recalls
1
Predicates
2

Its longest line of descent runs back 8 generations to K973598 (1997). 18 ancestors in all, the oldest K930338 (1993). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2011: 1 other ancestor2008: 2 other ancestors2006: 1 other ancestor2001: 1 other ancestor2003: 1 other ancestor2000: 2 other ancestors1997: 1 other ancestor1993: 1 other ancestor2023: 1 descendant cleared2024: 1 descendant clearedK973598 · Siremobil CO2 · Siemens Medical Solutions USA, Inc. · 1997 · recalledK030450 · 3D Accu-I-Tomo Xyz Slice View Tomograph · J. Morita USA, Inc. · 2003×2: 2 generations cleared close together, K052587 (2005), K060328 (2006) K073350 · Scanora 3D · Soredex Palodex Group OY · 2007K093683 · Orthopantomograph OP300 · Instrumentarium Dental, Palodex Group OY · 2010K122018 · OP300 · Palodex Group OY · 2012K133544 · OP300 · Palodex Group OY · 20141990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 1 other ancestor4 generations back: 2 other ancestors5 generations back: 2 other ancestors6 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K133544 · OP300 · Palodex Group OY · 2014K122018 · OP300 · Palodex Group OY · 2012K093683 · Orthopantomograph OP300 · Instrumentarium Dental, Palodex Group OY · 2010K073350 · Scanora 3D · Soredex Palodex Group OY · 2007K060328 · Planmeca Promax 3D · Planmeca Oy · 2006K052587 · 3D Accu-I-Tomo Xyz Slice View Tomograph, Model MCT-1 Exf · J. Morita USA, Inc. · 2005K030450 · 3D Accu-I-Tomo Xyz Slice View Tomograph · J. Morita USA, Inc. · 2003K973598 · Siremobil CO2 · Siemens Medical Solutions USA, Inc. · 1997 · recalled−50
  1. Root · 8 generations backK973598 Siremobil CO2Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 7 generations backK030450 3D Accu-I-Tomo Xyz Slice View TomographJ. Morita USA, Inc. · 2003
  3. 6 generations backK052587 3D Accu-I-Tomo Xyz Slice View Tomograph, Model MCT-1 ExfJ. Morita USA, Inc. · 2005
  4. 5 generations backK060328 Planmeca Promax 3DPlanmeca Oy · 2006
  5. 4 generations backK073350 Scanora 3DSoredex Palodex Group OY · 2007
  6. 3 generations backK093683 Orthopantomograph OP300Instrumentarium Dental, Palodex Group OY · 2010
  7. 2 generations backK122018 OP300Palodex Group OY · 2012
  8. 1 generation backK133544 OP300Palodex Group OY · 2014
  9. This deviceK163423 Orthopantomograph OP300Palodex Group OY · 2017
  10. Descendant · cites it directly · 1 of its ownK230505 Orthopantomograph™ OP 3D LXPalodex Group OY · 2023
  11. Descendant · 2 generations on · 0 of its ownK241249 Orthopantomograph™ OP 3D LX (PAN 3D); Orthopantomograph™ OP 3D LX…Palodex Group OY · 2024

Most influential descendants

Showing 2 of 2 · See all
  • K230505Orthopantomograph™ OP 3D LXPalodex Group OY · 2023 · cites it directly1descendant
  • K241249Orthopantomograph™ OP 3D LX (PAN 3D); Orthopantomograph™ OP 3D LX (PAN CEPH 3D)Palodex Group OY · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K973598Siremobil CO2Siemens Medical Solutions USA, Inc.19971
−7K0304503D Accu-I-Tomo Xyz Slice View TomographJ. Morita USA, Inc.2003—
−6K0525873D Accu-I-Tomo Xyz Slice View Tomograph, Model MCT-1 ExfJ. Morita USA, Inc.2005—
−5K060328Planmeca Promax 3DPlanmeca Oy2006—
−4K073350Scanora 3DSoredex Palodex Group OY2007—
−3K093683Orthopantomograph OP300Instrumentarium Dental, Palodex Group OY2010—
−2K122018OP300Palodex Group OY2012—
−1K133544OP300Palodex Group OY2014—
0 (this device)K163423Orthopantomograph OP300Palodex Group OY2017—
+1K230505Orthopantomograph™ OP 3D LXPalodex Group OY2023—
+2K241249Orthopantomograph™ OP 3D LX (PAN 3D); Orthopantomograph™ OP 3D LX (PAN CEPH 3D)Palodex Group OY2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199710
200020
200110
200310
200610
200820
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
FIG. 1 — Lineage of K163423, 1997–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OAS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OAS device

  1. #1 K213226 RCT700score 18.0 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K110839Scanora 3D
Summary page 4
…aphy x-ray system CFR Number: 892.1750 Device Class: II Product Code: OAS Predicate Devices: Proprietary Name: Scanora 3D 510(k) Number K110839 Common Name: X-ray, Tomography, Computed, Dental Classification Name: Computed tomography x-ray system CFR Number:…
Soredex Palodex Group OY2011—
*K133544OP300
Summary page 4
…tomography x-ray system CFR Number: 892.1750 Device Class: II Product Code: OAS Proprietary Name: Orthopantomograph OP300 510(k) Number K133544 Common Name: X-ray, Tomography, Computed, Dental Classification Name: Computed tomography x-ray system CFR Number:…
Palodex Group OY2014—

Also cleared against K110839: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K230505Orthopantomograph™ OP 3D LXPalodex Group OY20231

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 28, 2021Class IIZ-2413-2024Radiation Control for Health and Safety ActIt was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 6 companies; a new generation was cleared every 2.6 years on median.
  • 2 of the 18 ancestors have a recall of any class.
  • Of its descendants, K230505 (Palodex Group OY, 2023) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K110839.

How does its category compare?

All OAS devices

OAS X-Ray, Tomography, Computed, Dental: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · OAS

9in 2025

2007: 22008: 22009: 22010: 12011: 62012: 72013: 102014: 52015: 52016: 122017: 42018: 112019: 72020: 92021: 112022: 62023: 122024: 42025: 92026 (partial year): 9
2007high 12 (2016) · low 1 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OAS

144in 2025

2011: 168 days median2012: 110 days median2013: 155 days median2014: 170 days median2015: 164 days median2016: 154 days median2017: 209 days median2018: 57 days median2019: 55 days median2020: 113 days median2021: 92 days median2022: 105 days median2023: 77 days median2024: 129 days median2025: 144 days median2026 (partial year): 77 days median
2007high 209 (2017) · low 55 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OAS

3in 2025

2015: 12025: 3
2007high 3 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OAS

1in 2025

2008: 12019: 12022: 12023: 12024: 22025: 1
2007high 2 (2024) · low 0 (2007)2025

MAUDE reports naming product code OAS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—01
20092—00
20101—00
2011616800
2012711000
20131015500
2014517000
2015516410
20161215400
2017420900
2018115700
201975501
2020911300
2021119200
2022610501
2023127701
2024412902
2025914431
20269770—

What is it indicated for?

The Orthopantomograph OP300 panoramic, cephalometric and cone beam computed tomography x-ray device is intended to image the head and neck areas for diagnostic support. This includes temporomandibular Joints (TMJs) and dentomaxillofacial areas, and with the 13x15 cm field of view (FOV) this additionally includes the ear, nose and throat (ENT) regions. The x-ray device produces conventional 2D x-ray images and x-ray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K163423: Orthopantomograph OP300.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K163423

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026