510(k) Report

510(k) · Cleared

Cleared

SurgiCube

K163455510(k) TraditionalCleared Sep 28, 2017293-day reviewClass II21 CFR 878.5070

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Surgicube International B.V. · ORC Air Filter Portable Apparatus

Cleared on Sep 28, 2017 as substantially equivalent to K153498 (Toul Meditech AB, 2016) after a 293-day review. Its family tree reaches back 4 generations to 1984.

Decided
Sep 28, 2017
Review
293 days
Class
Class II
Product code
ORC
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K831531 (1984). No later device cites it as a predicate yet.

  1. Root · 4 generations backK831531 Surgical Isolation Barrier SystemLone Star Medical Products, Inc. · 1984
  2. 3 generations backK092801 Abs Blower, Abs Nozzle, Model 1001Nimbic Systems, LLC · 2011
  3. 2 generations backK123006 Air Barrier System (Abs)Nimbic Systems, Inc. · 2013
  4. 1 generation backK153498 OperioToul Meditech AB · 2016
  5. This deviceK163455 SurgiCubeSurgicube International B.V. · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K831531Surgical Isolation Barrier SystemLone Star Medical Products, Inc.1984—
−3K092801Abs Blower, Abs Nozzle, Model 1001Nimbic Systems, LLC2011—
−2K123006Air Barrier System (Abs)Nimbic Systems, Inc.2013—
−1K153498OperioToul Meditech AB2016—
0 (this device)K163455SurgiCubeSurgicube International B.V.2017—
FIG. 1 — Lineage of K163455, 1984–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in ORC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K153498Operio
Summary page 4
…21 CFR 878.5070 Product code ORC Predicate device The legally marketed device to which substantial equivalence is claimed is: Operio K153498 Cleared 07/20/2016 K163455 Page 1 of 5…
Toul Meditech AB2016—

Also cleared against K153498: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 5 companies; a new generation was cleared every 2.6 years on median.
  • None of the 4 ancestors has a recall.
  • 1 other device also cites its predicate K153498.

How does its category compare?

All ORC devices

ORC Air Filter Portable Apparatus: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · ORC

1in 2019

2011: 12013: 12016: 12017: 22019: 1
2007high 2 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ORC

Too few ORC 510(k)s in any year since 2007 to give a median.

Recalls per year · ORC

No recall names ORC since 2007.

Adverse event reports per year · ORC

No MAUDE report names ORC since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20111—00
20120—00
20131—00
20140—00
20150—00
20161—00
20172—00
20180—00
20191—00
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

SurgiCube® is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments. The surgical site of the patient is intended to…

Read the full statement (124 words)

SurgiCube® is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments.

The surgical site of the patient is intended to be placed under the air flow which is directed downwards to the surgical site and/or instruments within 32" (80 cm) in length. Device effectiveness may not be reliably detectable at a distance of 32" (80 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K163455: SurgiCube.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K163455

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026