510(k) Report

510(k) · Cleared

Cleared

Idylla Respiratory (IFV-RSV) Panel

K163628510(k) TraditionalCleared Aug 30, 2017251-day reviewClass II21 CFR 866.3980

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Janssen Pharmaceutica NV · OCC Respiratory Virus Panel Nucleic Acid Assay System

Cleared on Aug 30, 2017 as substantially equivalent to K103209 (Nanosphere, Inc., 2011) after a 251-day review. Its family tree reaches back 6 generations to 2008.

Decided
Aug 30, 2017
Review
251 days
Class
Class II
Product code
OCC
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to DEN070013 (2008). 13 ancestors in all, the oldest DEN060008 (2006). No later device cites it as a predicate yet.

  1. Root · 6 generations backDEN070013 ID-Tag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc. · 2008
  2. 5 generations backK081483 Xtag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc. · 2008 · Recalled
  3. 4 generations backK091667 Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112…Luminex Molecular Diagnostics, Inc. · 2009 · Recalled
  4. 3 generations backK092500 Modification to Proflu+ AssayProdesse, Inc. · 2009
  5. 2 generations backK101855 Profast+ AssayGen-Probe Prodesse, Inc. · 2010
  6. 1 generation backK103209 Verigene Respiratory Virus Nucleic Acid Test +(Rvnat+)Nanosphere, Inc. · 2011 · Recalled
  7. This deviceK163628 Idylla Respiratory (IFV-RSV) PanelJanssen Pharmaceutica NV · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6DEN070013ID-Tag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc.2008—
−5K081483Xtag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc.20081
−4K091667Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112…Luminex Molecular Diagnostics, Inc.20091
−3K092500Modification to Proflu+ AssayProdesse, Inc.2009—
−2K101855Profast+ AssayGen-Probe Prodesse, Inc.2010—
−1K103209Verigene Respiratory Virus Nucleic Acid Test +(Rvnat+)Nanosphere, Inc.20111
0 (this device)K163628Idylla Respiratory (IFV-RSV) PanelJanssen Pharmaceutica NV2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200610
200830
200920
201010
FIG. 1 — Lineage of K163628, 2008–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OCC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OCC device

  1. #1 K262747 ID NOW Influenza A & B 2; ID NOW RSV; ID NOW…score 15.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103209Verigene Respiratory Virus Nucleic Acid Test +(Rvnat+)
Summary page 5
…ments Proposed Predicate Device Nanosphere Verigene® Respiratory Virus Plus Nucleic Acid Test (RV+) on the Nanosphere Verigene® System (K103209) − Device Description The Idylla™ Respiratory (IFV-RSV) Panel is a self-contained molecular diagnostic test designed to work…
Nanosphere, Inc.20111

Also cleared against K103209: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 5 companies; a new generation was cleared every 11 months on median.
  • 3 of the 13 ancestors have a recall of any class.
  • 1 other device also cites its predicate K103209.

How does its category compare?

All OCC devices

OCC Respiratory Virus Panel Nucleic Acid Assay System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · OCC

1in 2024

2008: 42009: 62010: 22011: 92012: 62013: 62014: 42015: 82016: 72017: 102018: 42019: 52020: 22021: 12022: 12023: 22024: 12026 (partial year): 3
2007high 10 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OCC

31in 2019

2008: 28 days median2009: 24 days median2011: 82 days median2012: 125 days median2013: 79 days median2014: 94 days median2015: 85 days median2016: 92 days median2017: 136 days median2018: 134 days median2019: 31 days median2026 (partial year): 130 days median
2007high 136 (2017) · low 24 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OCC

1in 2024

2014: 12018: 12019: 12024: 1
2007high 1 (2014) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OCC

2in 2025

2011: 22012: 12014: 522015: 12016: 362017: 12018: 12019: 22020: 172021: 6862022: 22023: 12024: 32025: 2
2007high 686 (2021) · low 0 (2007)2025

MAUDE reports naming product code OCC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
200842800
200962400
20102—00
201198202
2012612501
201367900
2014494152
201588501
2016792036
20171013601
2018413411
201953112
20202—017
20211—0686
20221—02
20232—01
20241—13
20250—02
202631300—

What is it indicated for?

The Idylla Respiratory (IFV-RSV) Panel is an in vitro assay intended for the qualitative detection of nucleic acids for Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype 2009 H1, H275Y mutation of Influenza A subtype 2009 H1, Influenza B and Respiratory Syncytial Virus (A and B) from nasopharyngeal swabs in viral transport media of adult and pediatric patients. The test uses the Idylla system to aid in the diagnosis of…

Read the full statement (217 words)

The Idylla Respiratory (IFV-RSV) Panel is an in vitro assay intended for the qualitative detection of nucleic acids for Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype 2009 H1, H275Y mutation of Influenza A subtype 2009 H1, Influenza B and Respiratory Syncytial Virus (A and B) from nasopharyngeal swabs in viral transport media of adult and pediatric patients.

The test uses the Idylla system to aid in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings.

Negative results do not preclude respiratory virus infection or co-infection with other viruses and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Performance characteristics for Influenza A were established when influenza A/2009 H1 and H3 were the predominant influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL3+ facility is available to receive and culture specimens.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K163628: Idylla Respiratory (IFV-RSV) Panel.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K163628

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026