510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

RA-308 Excimer Laser System, Excimer Laser Catheter

K170349510(k) TraditionalCleared May 24, 2017110-day reviewClass II21 CFR 870.1250Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ra Medical Systems,Inc · PDU Catheter For Crossing Total Occlusions

Cleared on May 24, 2017 as substantially equivalent to K040067 (Spectranetics Corp., 2004) after a 110-day review. Its family tree reaches back 7 generations to 1989. 2 later devices descend from it. It has been recalled 5 times.

Decided
May 24, 2017
Review
110 days
Class
Class II
Product code
PDU
Recalls
5
Predicates
1

Its longest line of descent runs back 7 generations to K890505 (1989). 26 ancestors in all, the oldest K871011 (1987). 2 later devices descend from it, 1 citing it directly.

  1. Root · 7 generations backK890505 Usci Grey/Black Guide WireC.R. Bard, Inc. · 1989
  2. 6 generations backK961271 Usci Commander Series Guide WiresC.R. Bard, Inc. · 1996
  3. 5 generations backK990639 Acs HI-Torque Cross IT 100XT, 200XT and 300XT Guide Wires with…Guidant Corp. · 1999
  4. 4 generations backK011986 Safe-Steer Guide Wire SystemIntra Luminal Therapeutics, Inc. · 2002
  5. 3 generations backK021323 Modification to Safe-Steer Guide Wire SystemIntra Luminal Therapeutics, Inc. · 2002
  6. 2 generations backK031842 Safe-Cross Radio Frequency Total Occlusion Crossing SystemIntra Luminal Therapeutics, Inc. · 2003
  7. 1 generation backK040067 Clirpath Excimer Laser CatheterSpectranetics Corp. · 2004
  8. This deviceK170349 RA-308 Excimer Laser System, Excimer Laser CatheterRa Medical Systems,Inc · 2017
  9. Descendant · cites it directly · 1 of its ownK210664 DABRA Laser System (DABRA Laser model RA-308 and DABRA Catheter…Ra Medical Systems,Inc · 2021
  10. Descendant · 2 generations on · 0 of its ownK220544 DABRA Excimer Laser System (DABRA Laser model RA-308 and DABRA…Ra Medical Systems,Inc · 2022

Most influential descendants

Showing 2 of 2 · See all
  • K210664DABRA Laser System (DABRA Laser model RA-308 and DABRA Catheter model 101)Ra Medical Systems,Inc · 2021 · cites it directly1descendant
  • K220544DABRA Excimer Laser System (DABRA Laser model RA-308 and DABRA Catheter model 2.0)Ra Medical Systems,Inc · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K890505Usci Grey/Black Guide WireC.R. Bard, Inc.1989—
−6K961271Usci Commander Series Guide WiresC.R. Bard, Inc.1996—
−5K990639Acs HI-Torque Cross IT 100XT, 200XT and 300XT Guide Wires with Hydrocoat…Guidant Corp.1999—
−4K011986Safe-Steer Guide Wire SystemIntra Luminal Therapeutics, Inc.2002—
−3K021323Modification to Safe-Steer Guide Wire SystemIntra Luminal Therapeutics, Inc.2002—
−2K031842Safe-Cross Radio Frequency Total Occlusion Crossing SystemIntra Luminal Therapeutics, Inc.2003—
−1K040067Clirpath Excimer Laser CatheterSpectranetics Corp.2004—
0 (this device)K170349RA-308 Excimer Laser System, Excimer Laser CatheterRa Medical Systems,Inc2017—
+1K210664DABRA Laser System (DABRA Laser model RA-308 and DABRA Catheter model 101)Ra Medical Systems,Inc2021—
+2K220544DABRA Excimer Laser System (DABRA Laser model RA-308 and DABRA Catheter model…Ra Medical Systems,Inc2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
198850
198910
199010
199110
199310
199410
199710
199820
199910
200010
200210
200320

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202210
FIG. 1 — Lineage of K170349, 1989–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PDU? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PDU device

  1. #1 K262013 Kiba CTO Crossing Catheterscore 15.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040067Clirpath Excimer Laser Catheter
Summary page 4
…us catheter D. Device Class: Class II (per 870.1250 percutaneous catheter) E. Device Code: PDU Identification of Predicate Device Spectranetics CLiRpath Laser Catheters; K040067. Identification of Reference Devices Ra Medical EX-308; K062963 Spectranetics CVX-300; K052514 Device Description Ra Medical Syst…
Spectranetics Corp.2004—
†K062963Pharos Excimer Laser System, Model EX-308Ra Medical Systems,Inc20072
†K052514Clirpath Turbo Plus Excimer Laser Catheters, Models 314-151, 314-159, 317-152, 317-156…Spectranetics Corp.2005—

Also cleared against K040067: 3 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K210664DABRA Laser System (DABRA Laser model RA-308 and DABRA Catheter model 101)Ra Medical Systems,Inc20211

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 16, 2020Class IIZ-0569-2021Device DesignDue to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.
Jul 24, 2020Class IIZ-0050-2021Device DesignDue to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury.
Jan 23, 2020Class IIZ-2497-2020Software designA software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user or patient.
Sep 19, 2019Class IIZ-2006-2020Under Investigation by firmThe firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.
Feb 15, 2018Class IIZ-2218-2019OtherLasers/Catheters did not calibrate during set-up prior to use.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 5 companies; a new generation was cleared every 2.7 years on median.
  • 1 of the 26 ancestors has a recall of any class.
  • Of its descendants, K210664 (Ra Medical Systems,Inc, 2021) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K040067.

How does its category compare?

All PDU devices

PDU Catheter For Crossing Total Occlusions: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · PDU

3in 2025

2007: 42008: 22009: 22010: 32011: 32012: 62013: 12014: 32015: 22016: 22017: 12019: 12020: 22021: 32022: 42023: 32024: 12025: 32026 (partial year): 1
2007high 6 (2012) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PDU

59in 2025

2007: 66 days median2010: 49 days median2011: 91 days median2012: 68 days median2014: 97 days median2021: 144 days median2022: 139 days median2023: 221 days median2025: 59 days median
2007high 221 (2023) · low 49 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PDU

1in 2023

2020: 22023: 1
2007high 2 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PDU

179in 2025

2013: 32014: 102015: 122016: 262017: 2112018: 972019: 832020: 1292021: 1892022: 1562023: 712024: 922025: 179
2007high 211 (2017) · low 0 (2007)2025

MAUDE reports naming product code PDU, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200746600
20082—00
20092—00
201034900
201139100
201266800
20131—03
2014397010
20152—012
20162—026
20171—0211
20180—097
20191—083
20202—2129
202131440189
202241390156
20233221171
20241—092
20253590179
20261—0—

What is it indicated for?

The DABRA Laser System is indicated for crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170349: RA-308 Excimer Laser System, Excimer Laser Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170349

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.