510(k) · Cleared
Recalled · Class II · 2022BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR Calibrator Set, BioPlex 2200 Syphilis Total & RPR Control Set
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Bio-Rad Laboratories · LIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum
Cleared on May 11, 2017 as substantially equivalent to K061247 (DiaSorin, Inc., 2006) after a 90-day review. Its family tree reaches back 2 generations to 2002. It has been recalled 5 times.
- Decided
- May 11, 2017
- Review
- 90 days
- Class
- Class II
- Product code
- LIP
- Recalls
- 5
- Predicates
- 1
Its longest line of descent runs back 2 generations to K014233 (2002). No later device cites it as a predicate yet.
View FIG. 1 as a table
Line of descent and top descendants
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LIP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LIP device
- #1 K241534 VITROS Immunodiagnostic Products Syphilis…score 9.1 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K061247 | Diasorin Liaison Treponema AssaySummary page 5…control Material (Assayed and Chemistry unassayed) (JJX) Unassayed) IV. Predicate Devices LIAISON Treponema Assay, k061247 BD Macro-Vue RPR Card Tests, Pre-amendment prior to May 28, 1976 V. Device Description: BioPlex 2200 Syphilis… | 2006 |
| * | — | BD Macro-Vue RPR Card Test Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 8, 2022 | Class II | Z-0806-2022 | Under Investigation by firm | Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet… |
| Feb 8, 2022 | Class II | Z-0807-2022 | Under Investigation by firm | Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet… |
| Sep 17, 2021 | Class II | Z-0235-2022 | Nonconforming Material/Component | Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID… |
| Sep 17, 2021 | Class II | Z-0236-2022 | Nonconforming Material/Component | Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID… |
| Apr 21, 2020 | Class II | Z-2982-2020 | Unknown/Undetermined by firm | The increased level of RPR reactivity in the BioPlex 2200 Syphilis Total & RPR kit material number 12000650. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 10.8 years on median.
- None of the 2 ancestors has a recall.
- 2 other devices also cite its predicate K061247.
How does its category compare?
All LIP devicesLIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.
FDA decisions per year · LIP
2in 2024
2026 to Oct 25, outlined.
Median 510(k) review, days · LIP
170in 2016
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LIP
2in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LIP
112in 2025
MAUDE reports naming product code LIP, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 2 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 1 | — | 0 | 0 |
| 2010 | 1 | — | 1 | 1 |
| 2011 | 2 | — | 0 | 0 |
| 2012 | 2 | — | 0 | 1 |
| 2013 | 0 | — | 1 | 4 |
| 2014 | 0 | — | 1 | 1 |
| 2015 | 0 | — | 0 | 2 |
| 2016 | 3 | 170 | 0 | 1 |
| 2017 | 1 | — | 0 | 3 |
| 2018 | 0 | — | 0 | 5 |
| 2019 | 0 | — | 0 | 9 |
| 2020 | 0 | — | 0 | 46 |
| 2021 | 1 | — | 2 | 54 |
| 2022 | 0 | — | 2 | 44 |
| 2023 | 0 | — | 0 | 66 |
| 2024 | 2 | — | 0 | 73 |
| 2025 | 0 | — | 0 | 112 |
| 2026 | 0 | — | 2 | — |
What is it indicated for?
The BioPlex Syphilis Total & RPR kit is a multiplex flow immunoassay intended for the qualitative detection of total (IgG/IgM) antibodies to Treponema pallidum and the qualitative detection and/or titer determination of non-treponemal reagin antibodies in human serum or plasma. The Syphilis Total or RPR assays may be used to supplement a previously determined reactive treponemal or non-treponemal test. The test system should be used in conjunction with…
Read the full statement (206 words)
The BioPlex Syphilis Total & RPR kit is a multiplex flow immunoassay intended for the qualitative detection of total (IgG/IgM) antibodies to Treponema pallidum and the qualitative detection and/or titer determination of non-treponemal reagin antibodies in human serum or plasma. The Syphilis Total or RPR assays may be used to supplement a previously determined reactive treponemal or non-treponemal test. The test system should be used in conjunction with other laboratory tests and clinical findings to aid in the diagnosis of syphilis infection.
The BioPlex 2200 Syphilis Total & RPR kit is not intended for use in screening blood or plasma donors
The BioPlex 2200 Syphilis Total & RPR kit is intended for use with the Bio-Rad BioPlex 2200 System.
The BioPlex 2200 Syphilis Total & RPR Control Set is intended for use as an assayed quality control to monitor the performance of the BioPlex 2200 Instrument and BioPlex 2200 Syphilis Total & RPR assay in the clinical laboratory. The performance of the BioPlex 2200 Syphilis Total & RPR Control Set has not been established with any other Syphilis Total & RPR assays.
The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K170413 · printed Nov 6, 2026
Cite this page
510(k) Report. “K170413: BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR Calibrator Set, BioPlex 2200 Syphilis Total & RPR Control Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170413