510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR Calibrator Set, BioPlex 2200 Syphilis Total & RPR Control Set

K170413510(k) TraditionalCleared May 11, 201790-day reviewClass II21 CFR 866.3830Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bio-Rad Laboratories · LIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum

Cleared on May 11, 2017 as substantially equivalent to K061247 (DiaSorin, Inc., 2006) after a 90-day review. Its family tree reaches back 2 generations to 2002. It has been recalled 5 times.

Decided
May 11, 2017
Review
90 days
Class
Class II
Product code
LIP
Recalls
5
Predicates
1

Its longest line of descent runs back 2 generations to K014233 (2002). No later device cites it as a predicate yet.

K014233 · Captiam Syphilis-G Assay · Trinity Biotech, Plc · 2002K061247 · Diasorin Liaison Treponema Assay · DiaSorin, Inc. · 20062000200520102015
  1. Root · 2 generations backK014233 Captiam Syphilis-G AssayTrinity Biotech, Plc · 2002
  2. 1 generation backK061247 Diasorin Liaison Treponema AssayDiaSorin, Inc. · 2006
  3. This deviceK170413 BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR…Bio-Rad Laboratories · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K014233Captiam Syphilis-G AssayTrinity Biotech, Plc2002—
−1K061247Diasorin Liaison Treponema AssayDiaSorin, Inc.2006—
0 (this device)K170413BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR…Bio-Rad Laboratories2017—
FIG. 1 — Lineage of K170413, 2002–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LIP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LIP device

  1. #1 K241534 VITROS Immunodiagnostic Products Syphilis…score 9.1 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061247Diasorin Liaison Treponema Assay
Summary page 5
…control Material (Assayed and Chemistry unassayed) (JJX) Unassayed) IV. Predicate Devices LIAISON Treponema Assay, k061247 BD Macro-Vue RPR Card Tests, Pre-amendment prior to May 28, 1976 V. Device Description: BioPlex 2200 Syphilis…
DiaSorin, Inc.2006—
*—BD Macro-Vue RPR Card Test
Named without a submission number
—

Also cleared against K061247: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 8, 2022Class IIZ-0806-2022Under Investigation by firmDue to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet…
Feb 8, 2022Class IIZ-0807-2022Under Investigation by firmDue to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet…
Sep 17, 2021Class IIZ-0235-2022Nonconforming Material/ComponentDue to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID…
Sep 17, 2021Class IIZ-0236-2022Nonconforming Material/ComponentDue to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID…
Apr 21, 2020Class IIZ-2982-2020Unknown/Undetermined by firmThe increased level of RPR reactivity in the BioPlex 2200 Syphilis Total & RPR kit material number 12000650.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 10.8 years on median.
  • None of the 2 ancestors has a recall.
  • 2 other devices also cite its predicate K061247.

How does its category compare?

All LIP devices

LIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LIP

2in 2024

2007: 12009: 12010: 12011: 22012: 22016: 32017: 12021: 12024: 2
2007high 3 (2016) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LIP

170in 2016

2016: 170 days median
2007high 170 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LIP

2in 2022

2007: 22010: 12013: 12014: 12021: 22022: 22026 (partial year): 2
2007high 2 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LIP

112in 2025

2010: 12012: 12013: 42014: 12015: 22016: 12017: 32018: 52019: 92020: 462021: 542022: 442023: 662024: 732025: 112
2007high 112 (2025) · low 0 (2007)2025

MAUDE reports naming product code LIP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—20
20080—00
20091—00
20101—11
20112—00
20122—01
20130—14
20140—11
20150—02
2016317001
20171—03
20180—05
20190—09
20200—046
20211—254
20220—244
20230—066
20242—073
20250—0112
20260—2—

What is it indicated for?

The BioPlex Syphilis Total & RPR kit is a multiplex flow immunoassay intended for the qualitative detection of total (IgG/IgM) antibodies to Treponema pallidum and the qualitative detection and/or titer determination of non-treponemal reagin antibodies in human serum or plasma. The Syphilis Total or RPR assays may be used to supplement a previously determined reactive treponemal or non-treponemal test. The test system should be used in conjunction with…

Read the full statement (206 words)

The BioPlex Syphilis Total & RPR kit is a multiplex flow immunoassay intended for the qualitative detection of total (IgG/IgM) antibodies to Treponema pallidum and the qualitative detection and/or titer determination of non-treponemal reagin antibodies in human serum or plasma. The Syphilis Total or RPR assays may be used to supplement a previously determined reactive treponemal or non-treponemal test. The test system should be used in conjunction with other laboratory tests and clinical findings to aid in the diagnosis of syphilis infection.

The BioPlex 2200 Syphilis Total & RPR kit is not intended for use in screening blood or plasma donors

The BioPlex 2200 Syphilis Total & RPR kit is intended for use with the Bio-Rad BioPlex 2200 System.

The BioPlex 2200 Syphilis Total & RPR Control Set is intended for use as an assayed quality control to monitor the performance of the BioPlex 2200 Instrument and BioPlex 2200 Syphilis Total & RPR assay in the clinical laboratory. The performance of the BioPlex 2200 Syphilis Total & RPR Control Set has not been established with any other Syphilis Total & RPR assays.

The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170413: BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR Calibrator Set, BioPlex 2200 Syphilis Total & RPR Control Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170413

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026