# K170592: NuVasive BASE Interfixated Titanium System

- Pathway: 510(k) (Traditional)
- Applicant: Nu Vasive, Incorporated
- Decided: Apr 26, 2017 after 57 days
- Product code: OVD (Intervertebral Fusion Device With Integrated Fixation, Lumbar)
- Device class: 2
- Regulation: 21 CFR 888.3080
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 19 generations back to K841189 (1984). 225 ancestors in all.
186 later devices descend from it; 10 cite it directly.
0 of 412 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K161230: NuVasive Lumbar Interbody Implants, Nu Vasive, Incorporated (2016) Summary page 4.
- K151214: ALIF Interfixated System, Nu Vasive, Incorporated (2015) Summary page 4.
- K152943: NuVasive Foundation-LL System, Nu Vasive, Incorporated (2015) Summary page 4.
- K073109: Stalif TT Intervertebral Body Fusion System, Surgicraft, Ltd. (2008) Summary page 4.

## Most influential descendants
- K193593: NuVasive 3DP Interfixated ALIF System (2020), 157 descendants of its own
- K201820: NuVasive Thoracolumbar Interbody Systems: CoRoent Thoracolumbar System, CoRoent XL Interfixated System, Brigade Standalone System and Brigade Hyperlordotic System, Brigade Lateral System, BASE Interfixated Titanium System, Coalesce Thoracolumbar Interbody Fusion System, Cohere Thoracolumbar Interbody System, Modulus XLIF Interbody System, Modulus TLIF Interbody System, 3DP Interfixated ALIF System (2020), 156 descendants of its own
- K203201: NuVasive Thoracolumbar Interbody Systems: CoRoent Thoracolumbar System, CoRoent XL Interfixated System, Brigade Hyperlordotic System, Brigade Standalone System, Brigade Lateral System, BASE Interfixated Titanium System, Coalesce Thoracolumbar Interbody Fusion System, Cohere Thoracolumbar Interbody System, Modulus XLIF Interbody System, Modulus TLIF Interbody System, Modulus ALIF System (2021), 94 descendants of its own
- K211258: Endoskeleton TC Interbody System, Endoskeleton TCS Interbody System, Endoskeleton TA Interbody System, Endoskeleton TAS & TAS Hyp Interbody System, Endoskeleton TL Interbody System, Endoskeleton TL Hyp. Interbody System, Endoskeleton TO Interbody System, Endoskeleton TT Interbody System (2021), 75 descendants of its own
- K203714: NuVasive Thoracolumbar Interbody Systems: NuVasive CoRoent Thoracolumbar System, NuVasive CoRoent XL Interfixated System, Brigade System, Brigade Lateral System, BASE Interfixated Titanium System, Coalesce Thoracolumbar Interbody Fusion System, NuVasive Cohere Thoracolumbar Interbody System, NuVasive Modulus XLIF Interbody System, NuVasive Modulus TLIF Interbody System, NuVasive Modulus ALIF System, NuVasive Attrax Putty (2021), 66 descendants of its own

## Indications for use
The BASE Interfixated Titanium System is indicated for spinal fusion procedures in skeletally mature patients. The BASE Interfixated Titanium System 10° – 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The BASE Interfixated Titanium System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The BASE Interfixated Titanium System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the BASE Interfixated Titanium System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.