510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

CyberKnife Treatment Delivery System

K170788510(k) TraditionalCleared Apr 11, 201726-day reviewClass II21 CFR 892.5050Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Accuray Incorportaed · IYE Accelerator, Linear, Medical

Cleared on Apr 11, 2017 as substantially equivalent to K150873 (Accuray Incorporated, 2015) after a 26-day review. Its family tree reaches back 1 generation to 2015. It has been recalled 4 times.

Decided
Apr 11, 2017
Review
26 days
Class
Class II
Product code
IYE
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K150873 (2015). No later device cites it as a predicate yet.

K150873 · CyberKnife M6 FI Sytem, CyberKnife M6 FM System, CyberKnife M6 FIM System · Accuray Incorporated · 2015 · recalled20162018
  1. Root · 1 generation backK150873 CyberKnife M6 FI Sytem, CyberKnife M6 FM System, CyberKnife M6 FIM…Accuray Incorporated · 2015 · Recalled
  2. This deviceK170788 CyberKnife Treatment Delivery SystemAccuray Incorportaed · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K150873CyberKnife M6 FI Sytem, CyberKnife M6 FM System, CyberKnife M6 FIM SystemAccuray Incorporated20152
0 (this device)K170788CyberKnife Treatment Delivery SystemAccuray Incorportaed2017—
FIG. 1 — Lineage of K170788, 2015–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K150873CyberKnife M6 FI Sytem, CyberKnife M6 FM System, CyberKnife M6 FIM System
Summary page 5
…valence The CyberKnife® Treatment Delivery System is substantially equivalent to the treatment delivery component of the predicate, CyberKnife M6 Systems (cleared under K150873). The intended use for radiologic applications, principles of operation, technological characteristics are the same as the treatm…
Accuray Incorporated20152

Also cleared against K150873: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 16, 2026Class IIZ-2640-2026Software designIn certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the…
Jun 16, 2026Class IIZ-2644-2026Software designIn certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the…
Dec 23, 2024Class IIZ-1023-2025Component design/selectionSnap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.
Feb 26, 2021Class IIZ-1272-2021Equipment maintenanceThe set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 3 other devices also cite its predicate K150873.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The CyberKnife® Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170788: CyberKnife Treatment Delivery System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170788

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026