# K171441: ARIES C. difficile Assay Complete Kit, ARIES C. difficile Assay Protocol File Kit, ARIES C. difficile Assay Kit (24 cassettes), ARIES Stool Resuspension Kit

- Pathway: 510(k) (Traditional)
- Applicant: Luminex Corporation
- Decided: Jul 21, 2017 after 66 days
- Product code: OZN (C. Difficile Toxin Gene Amplification Assay)
- Device class: 2
- Regulation: 21 CFR 866.3130
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K123998 (2013). 2 ancestors in all.
No later device cites it as a predicate yet.
0 of 3 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K123998: Quidel Molecular Direct C. Difficile Assay, Quidel Corp. (2013) Summary page 6.

## Recalls of this device
- Feb 20, 2019 · Class II · Z-2287-2019: Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impa

## Indications for use
The ARIES C. difficile Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of toxigenic Clostridium difficile (C. difficile) nucleic acid in unpreserved, unformed (liquid or soft) stool specimens obtained from patients suspected of having Clostridium difficile infection (CDI). The test targets the C. difficile toxin A gene (tcdA) and toxin B gene (tcdB) and is indicated for use as an aid in the diagnosis of C. difficile infection (CDI). The ARIES C. difficile Assay is indicated for use with ARIES Systems.

Source page: https://510kreport.com/device/K171441