# K171655: cobas m 511 integrated hematology analyzer

- Pathway: 510(k) (Traditional)
- Applicant: Roche Diagnostics Hematology, Inc.
- Decided: Mar 2, 2018 after 270 days
- Product code: GKZ (Counter, Differential Cell)
- Device class: 2
- Regulation: 21 CFR 864.5220
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K925670 (1993). 7 ancestors in all.
No later device cites it as a predicate yet.
0 of 8 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K112605: Sysmex XN-Series, Sysmex America, Inc. (2012) Summary page 9.
- K092868: Cellavision DM1200 Automated Hematology Analyzer, Model XU-10127, CellaVision AB (2009) Summary page 9.

## Recalls of this device
- May 7, 2018 · Class II · Z-2557-2018: Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell count), HGB (hemoglobi

## Indications for use
The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA-anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.