# K171953: Bard Mission Disposable Core Biopsy Instrument

- Pathway: 510(k) (Traditional)
- Applicant: Bard Peripheral Vascular, Inc.
- Decided: Sep 14, 2017 after 77 days
- Product code: KNW (Instrument, Biopsy)
- Device class: 2
- Regulation: 21 CFR 876.1075
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 3 generations back to K950732 (1995). 5 ancestors in all.
5 later devices descend from it; 4 cite it directly.
0 of 11 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K161409: M-Biopsy Semi-Automatic Biopsy Instrument M-Biopsy Coaxial Introducer Needle for Sem, Mermaid Medical A/S (2016) Summary page 4.
- K133948: Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R) Max-Core (R) Disposable Core Instrument, Bard Peripheral Vascular, Inc. (2014) Summary page 4.

## Most influential descendants
- K222865: M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic Disposable Biopsy Needle, M Biopsy /SureAim Coaxial Biopsy Needle (2023), 1 descendant of its own
- K250032: Marquee™ Disposable Core Biopsy Instrument and Instrument Kit (2025), 0 descendants of its own
- K241793: Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable Coaxial Biopsy Needle (2025), 0 descendants of its own
- K233031: M·Biopsy /SureCore Automatic Disposable Biopsy Needle, M·Biopsy /SureCore Semi-automatic Disposable Biopsy Needle, M·Biopsy /SureAim Coaxial Biopsy Needle (2024), 0 descendants of its own
- K201166: Temno Elite Biopsy System (2020), 0 descendants of its own

## Recalls of this device
- Nov 8, 2023 · Class II · Z-0643-2024: The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a res
- Apr 11, 2023 · Class II · Z-1564-2023: Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial ca
- Feb 28, 2023 · Class II · Z-1260-2023: The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle 

## Indications for use
The Bard® Mission® Disposable Core Biopsy Instrument is intended for use in obtaining biopsy samples from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors.

Source page: https://510kreport.com/device/K171953