510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Resection Electrodes with HF cable

K171965510(k) TraditionalCleared Sep 28, 201790-day reviewClass II21 CFR 876.4300Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Winter & Ibe GmbH · FAS Electrode, Electrosurgical, Active, Urological

Cleared on Sep 28, 2017 as substantially equivalent to K152092 (Olympus Winter & Ibe GmbH, 2015) after a 90-day review. Its family tree reaches back 6 generations to 1991. 3 later devices descend from it. It has been recalled once.

Decided
Sep 28, 2017
Review
90 days
Class
Class II
Product code
FAS
Recalls
1
Predicates
1

Its longest line of descent runs back 6 generations to K911904 (1991). 23 ancestors in all, the oldest K897003 (1990). 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 2 other ancestors1994: 3 other ancestors1990: 2 other ancestors2003: 1 other ancestor1999: 4 other ancestors1998: 1 other ancestor2000: 3 other ancestors1996: 1 other ancestor2021: 2 descendants cleared2023: 1 descendant clearedK911904 · Models Ues-10 & PSD-10 · Olympus Corp. · 1991K970184 · Olympus Ues-20 Electrosurgical System and Its Associated Accessories · Olympus America, Inc. · 1997×2: 2 generations cleared close together, K023767 (2002), K030194 (2003) K100275 · Working Element, Resection Button Electrode for Plasma Vaporizaton, HF-Resection Electrode Loops and Bands · Olympus Winter & Ibe GmbH · 2010 · recalledK152092 · Resection Electrodes · Olympus Winter & Ibe GmbH · 2015 · recalled1990200020102020
  1. Root · 6 generations backK911904 Models Ues-10 & PSD-10Olympus Corp. · 1991
  2. 5 generations backK970184 Olympus Ues-20 Electrosurgical System and Its Associated AccessoriesOlympus America, Inc. · 1997
  3. 4 generations backK023767 Ues-30 Electrosurgical Unit and It's Associated AccessoriesThe Olympus Optical Co. · 2002
  4. 3 generations backK030194 Xues-41 Endoscopic Electrosurgical UnitThe Olympus Optical Co. · 2003 · Recalled
  5. 2 generations backK100275 Working Element, Resection Button Electrode for Plasma Vaporizaton…Olympus Winter & Ibe GmbH · 2010 · Recalled
  6. 1 generation backK152092 Resection ElectrodesOlympus Winter & Ibe GmbH · 2015 · Recalled
  7. This deviceK171965 Resection Electrodes with HF cableOlympus Winter & Ibe GmbH · 2017
  8. Descendant · cites it directly · 1 of its ownK210651 Resection Electrodes with HF cableOlympus Winter & Ibe GmbH · 2021 · Recalled
  9. Descendant · 2 generations on · 0 of its ownK221893 KARL STORZ Bipolar Resectoscopes with HF CableKarl Storz Endoscopy America, Inc. · 2023
  10. Descendant · cites it directly · 0 of its ownK210394 Electrosurgical accessoryDornier Medtech America, Inc. · 2021

Most influential descendants

Showing 3 of 3 · See all
  • K210651Resection Electrodes with HF cableOlympus Winter & Ibe GmbH · 2021 · cites it directly1descendant
  • K221893KARL STORZ Bipolar Resectoscopes with HF CableKarl Storz Endoscopy America, Inc. · 2023 · 2 generations on0descendants
  • K210394Electrosurgical accessoryDornier Medtech America, Inc. · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K911904Models Ues-10 & PSD-10Olympus Corp.1991—
−5K970184Olympus Ues-20 Electrosurgical System and Its Associated AccessoriesOlympus America, Inc.1997—
−4K023767Ues-30 Electrosurgical Unit and It's Associated AccessoriesThe Olympus Optical Co.2002—
−3K030194Xues-41 Endoscopic Electrosurgical UnitThe Olympus Optical Co.20033
−2K100275Working Element, Resection Button Electrode for Plasma Vaporizaton…Olympus Winter & Ibe GmbH20101
−1K152092Resection ElectrodesOlympus Winter & Ibe GmbH20151
0 (this device)K171965Resection Electrodes with HF cableOlympus Winter & Ibe GmbH2017—
+1K210651Resection Electrodes with HF cableOlympus Winter & Ibe GmbH20211
+2K221893KARL STORZ Bipolar Resectoscopes with HF CableKarl Storz Endoscopy America, Inc.2023—
+1K210394Electrosurgical accessoryDornier Medtech America, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199020
199430
199610
199720
199810
199940
200030
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202120
202310
FIG. 1 — Lineage of K171965, 1991–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FAS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FAS device

  1. #1 K250522 Multi4 Systemscore 11.4 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152092Resection Electrodes
Summary page 6
…WA22332S, WA22351S, FAS / 876.4300 WA22355S, WA22557S, K152092 Resection Electrodes WA22537S, WA22538S,…
Olympus Winter & Ibe GmbH20151

Also cleared against K152092: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K210651Resection Electrodes with HF cableOlympus Winter & Ibe GmbH20211
K210394Electrosurgical accessoryDornier Medtech America, Inc.20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 4, 2022Class IIZ-0424-2023Packaging process controlAn incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 5.6 years on median.
  • 5 of the 23 ancestors have a recall of any class.
  • Of its descendants, K210651 (Olympus Winter & Ibe GmbH, 2021) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K152092.

How does its category compare?

All FAS devices

FAS Electrode, Electrosurgical, Active, Urological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · FAS

2in 2025

2010: 32011: 12012: 22015: 12016: 12017: 22021: 22022: 12024: 12025: 2
2007high 3 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FAS

150in 2010

2010: 150 days median
2007high 150 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FAS

1in 2023

2016: 22021: 42023: 1
2007high 4 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FAS

395in 2025

2007: 62008: 132009: 92010: 252011: 332012: 302013: 482014: 522015: 822016: 692017: 1212018: 702019: 612020: 1002021: 1022022: 1472023: 3292024: 3102025: 395
2007high 395 (2025) · low 6 (2007)2025

MAUDE reports naming product code FAS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—06
20080—013
20090—09
20103150025
20111—033
20122—030
20130—048
20140—052
20151—082
20161—269
20172—0121
20180—070
20190—061
20200—0100
20212—4102
20221—0147
20230—1329
20241—0310
20252—0395
20260—0—

What is it indicated for?

Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in urological applications: Cutting, ablation, resection, vaporization and coagulation with HF current. Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications. The general indications include transcervical resection, vaporization, ablation, cutting and coagulation of tissue in the…

Read the full statement (100 words)

Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in urological applications: Cutting, ablation, resection, vaporization and coagulation with HF current.

Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications.

The general indications include transcervical resection, vaporization, ablation, cutting and coagulation of tissue in the uterus in conductive irrigation fluid as part of a resectoscope system.

Specific indications:

- transcervical diagnosis and treatment (resection, vaporization, ablation, biopsy, cutting and coagulation) of intrauterine myomas, intrauterine polyps, synechias and endometrium (TCRis)

- lysis of intrauterine septa

- endometrial ablation

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K171965: Resection Electrodes with HF cable.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K171965

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026