510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Olympus Urf-V2/V2R

K172246510(k) TraditionalCleared Mar 30, 2018247-day reviewClass II21 CFR 876.1500Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corp. · FGB Ureteroscope And Accessories, Flexible/Rigid

Cleared on Mar 30, 2018 as substantially equivalent to K072957 (Olympus Medical Systems Corporation, 2008) after a 247-day review. Its family tree reaches back 4 generations to 1986. 2 later devices descend from it. It has been recalled 3 times.

Decided
Mar 30, 2018
Review
247 days
Class
Class II
Product code
FGB
Recalls
3
Predicates
1

Its longest line of descent runs back 4 generations to K853585 (1986). 8 ancestors in all. 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2000: 2 other ancestors1993: 1 other ancestor2019: 2 descendants clearedK853585 · Olympus Evs-Endoscopic Video Image & Data Sys · Olympus Corp. · 1986 · recalledK954451 · Evis 140 System · Olympus America, Inc. · 1996 · recalledK051645 · Evis Exera 160A System · Olympus Medical Systems Corporation · 2005 · recalledK072957 · Evis Exera II 180 System · Olympus Medical Systems Corporation · 2008 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 3 other ancestors4 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K072957 · Evis Exera II 180 System · Olympus Medical Systems Corporation · 2008 · recalledK051645 · Evis Exera 160A System · Olympus Medical Systems Corporation · 2005 · recalledK954451 · Evis 140 System · Olympus America, Inc. · 1996 · recalledK853585 · Olympus Evs-Endoscopic Video Image & Data Sys · Olympus Corp. · 1986 · recalled0
  1. Root · 4 generations backK853585 Olympus Evs-Endoscopic Video Image & Data SysOlympus Corp. · 1986 · Recalled
  2. 3 generations backK954451 Evis 140 SystemOlympus America, Inc. · 1996 · Recalled
  3. 2 generations backK051645 Evis Exera 160A SystemOlympus Medical Systems Corporation · 2005 · Recalled
  4. 1 generation backK072957 Evis Exera II 180 SystemOlympus Medical Systems Corporation · 2008 · Recalled
  5. This deviceK172246 Olympus Urf-V2/V2ROlympus Medical Systems Corp. · 2018
  6. Descendant · cites it directly · 1 of its ownK181451 Uretero-Reno Videoscope Olympus Urf-V3/V3R, Uretero-Reno Fiberscope…Olympus Medical Systems Corp. · 2019 · Recalled
  7. Descendant · 2 generations on · 0 of its ownK191078 BAROnova Access Sheath KitBaronova, Inc. · 2019

Most influential descendants

Showing 2 of 2 · See all
  • K181451Uretero-Reno Videoscope Olympus Urf-V3/V3R, Uretero-Reno Fiberscope Olympus Urf-P7/P7ROlympus Medical Systems Corp. · 2019 · cites it directly1descendant
  • K191078BAROnova Access Sheath KitBaronova, Inc. · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K853585Olympus Evs-Endoscopic Video Image & Data SysOlympus Corp.19862
−3K954451Evis 140 SystemOlympus America, Inc.19963
−2K051645Evis Exera 160A SystemOlympus Medical Systems Corporation20053
−1K072957Evis Exera II 180 SystemOlympus Medical Systems Corporation20081
0 (this device)K172246Olympus Urf-V2/V2ROlympus Medical Systems Corp.2018—
+1K181451Uretero-Reno Videoscope Olympus Urf-V3/V3R, Uretero-Reno Fiberscope Olympus…Olympus Medical Systems Corp.20191
+2K191078BAROnova Access Sheath KitBaronova, Inc.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
200020
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201920
FIG. 1 — Lineage of K172246, 1986–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FGB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FGB device

  1. #1 K251951 Single-Use Video Flexible Ureterorenoscope…score 16.2 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072957Evis Exera II 180 System
Summary page 5
…Traditional 510(k) URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2/V2R 3. Predicate Device K072957 VISERA URETERO-RENO VIDEOSCOPE OLYMPUS XURF TYPE V 4. Device Description This endoscope is composed from flexible insertion tube…
Olympus Medical Systems Corporation20081

Also cleared against K072957: 3 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181451Uretero-Reno Videoscope Olympus Urf-V3/V3R, Uretero-Reno Fiberscope Olympus Urf-P7/P7ROlympus Medical Systems Corp.20191

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 25, 2023Class IIZ-0348-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Mar 25, 2022Class IIZ-1116-2022Process change controlThe bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.
Jan 17, 2018Class IIZ-2518-2018Device DesignPotential breakage of the endoscope s insertion tube bending section during surgical procedures

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 4 companies; a new generation was cleared every 10.0 years on median.
  • 6 of the 8 ancestors have a recall of any class.
  • Of its descendants, K181451 (Olympus Medical Systems Corp., 2019) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K072957.

How does its category compare?

All FGB devices

FGB Ureteroscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · FGB

17in 2025

2012: 22013: 12014: 22015: 12017: 22018: 42019: 22020: 22021: 32022: 52023: 82024: 162025: 172026 (partial year): 9
2007high 17 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FGB

134in 2025

2018: 191 days median2021: 191 days median2022: 205 days median2023: 183 days median2024: 217 days median2025: 134 days median2026 (partial year): 140 days median
2007high 217 (2024) · low 134 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FGB

15in 2024

2022: 42023: 12024: 152026 (partial year): 1
2007high 15 (2024) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FGB

1,307in 2025

2007: 62008: 72009: 42010: 202011: 172012: 112013: 82014: 122015: 162016: 352017: 1422018: 1182019: 2072020: 4022021: 3252022: 4552023: 4582024: 5722025: 1,307
2007high 1,307 (2025) · low 4 (2009)2025

MAUDE reports naming product code FGB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—06
20080—07
20090—04
20100—020
20110—017
20122—011
20131—08
20142—012
20151—016
20160—035
20172—0142
201841910118
20192—0207
20202—0402
202131910325
202252054455
202381831458
20241621715572
20251713401,307
202691401—

What is it indicated for?

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the ureter and kidney.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172246: Olympus Urf-V2/V2R.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172246

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026