510(k) · Cleared
Recalled · Class II · 2023Olympus Urf-P6/P6R
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Olympus Medical Systems Corp. · FGB Ureteroscope And Accessories, Flexible/Rigid
Cleared on Mar 29, 2018 as substantially equivalent to K912120 (Olympus Corp., 1991) after a 241-day review. Its family tree reaches back 1 generation to 1991. 1 of the 4 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 5 times.
- Decided
- Mar 29, 2018
- Review
- 241 days
- Class
- Class II
- Product code
- FGB
- Recalls
- 5
- Predicates
- 1
Its longest line of descent runs back 1 generation to K912120 (1991). 2 later devices descend from it, 1 citing it directly. 1 of the 4 devices in this family, counting this one, has a Class I recall.
- Root · 1 generation backK912120 Urf-P2 Ureterorenofiberscope/CholedochofiberscopeOlympus Corp. · 1991 · Recalled (Class I)
- This deviceK172298 Olympus Urf-P6/P6ROlympus Medical Systems Corp. · 2018
- Descendant · cites it directly · 1 of its ownK181451 Uretero-Reno Videoscope Olympus Urf-V3/V3R, Uretero-Reno Fiberscope…Olympus Medical Systems Corp. · 2019 · Recalled
- Descendant · 2 generations on · 0 of its ownK191078 BAROnova Access Sheath KitBaronova, Inc. · 2019
Class I recalls in this family
- K912120 Urf-P2 Ureterorenofiberscope/Choledochofiberscopedirect predicate · 1991
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K912120 | Urf-P2 Ureterorenofiberscope/Choledochofiberscope | Olympus Corp. | 1991 | Class I |
| 0 (this device) | K172298 | Olympus Urf-P6/P6R | Olympus Medical Systems Corp. | 2018 | — |
| +1 | K181451 | Uretero-Reno Videoscope Olympus Urf-V3/V3R, Uretero-Reno Fiberscope Olympus… | Olympus Medical Systems Corp. | 2019 | 1 |
| +2 | K191078 | BAROnova Access Sheath Kit | Baronova, Inc. | 2019 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2019 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FGB device
- #1 K251951 Single-Use Video Flexible Ureterorenoscope…score 16.2 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K912120 | Urf-P2 Ureterorenofiberscope/CholedochofiberscopeSummary page 5…Traditional 510(k) URETERO-RENO FIBERSCOPE OLYMPUS URF-P6/P6R 3. Predicate Device K912120 URETERORENOSCOPE OLYMPUS URF TYPE P2 4. Device Description This endoscope is composed from eyepiece section, control section an… | 1991 |
What descends from it? (2)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K181451 | Uretero-Reno Videoscope Olympus Urf-V3/V3R, Uretero-Reno Fiberscope Olympus Urf-P7/P7R | 2019 | 1 |
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 25, 2023 | Class II | Z-0347-2024 | Under Investigation by firm | Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead… |
| Apr 20, 2022 | Class II | Z-1263-2022 | Nonconforming Material/Component | Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body… |
| Apr 20, 2022 | Class II | Z-1264-2022 | Nonconforming Material/Component | Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body… |
| Jan 17, 2018 | Class II | Z-2519-2018 | Device Design | Breakage of the endoscope's insertion tube bending section during surgical procedures. |
| Jan 17, 2018 | Class II | Z-2520-2018 | Device Design | Breakage of the endoscope's insertion tube bending section during surgical procedures. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K912120Urf-P2 Ureterorenofiberscope/CholedochofiberscopeOlympus Corp. · 1991 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
How does its category compare?
All FGB devicesFGB Ureteroscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.
FDA decisions per year · FGB
17in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FGB
134in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FGB
15in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FGB
1,307in 2025
MAUDE reports naming product code FGB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 6 |
| 2008 | 0 | — | 0 | 7 |
| 2009 | 0 | — | 0 | 4 |
| 2010 | 0 | — | 0 | 20 |
| 2011 | 0 | — | 0 | 17 |
| 2012 | 2 | — | 0 | 11 |
| 2013 | 1 | — | 0 | 8 |
| 2014 | 2 | — | 0 | 12 |
| 2015 | 1 | — | 0 | 16 |
| 2016 | 0 | — | 0 | 35 |
| 2017 | 2 | — | 0 | 142 |
| 2018 | 4 | 191 | 0 | 118 |
| 2019 | 2 | — | 0 | 207 |
| 2020 | 2 | — | 0 | 402 |
| 2021 | 3 | 191 | 0 | 325 |
| 2022 | 5 | 205 | 4 | 455 |
| 2023 | 8 | 183 | 1 | 458 |
| 2024 | 16 | 217 | 15 | 572 |
| 2025 | 17 | 134 | 0 | 1,307 |
| 2026 | 9 | 140 | 1 | — |
What is it indicated for?
This instrument has been designed to be used with an Olympus light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct).
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K172298 · printed Nov 5, 2026
Cite this page
510(k) Report. “K172298: Olympus Urf-P6/P6R.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K172298
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026