510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Tec 820, Tec 850

K172702510(k) TraditionalCleared Jan 4, 2018119-day reviewClass II21 CFR 868.5880Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda, Inc. · CAD Vaporizer, Anesthesia, Non-Heated

Cleared on Jan 4, 2018 as substantially equivalent to K031027 (Datex-Ohmeda, Inc., 2003) and 1 other predicate after a 119-day review. Its family tree reaches back 3 generations to 1989. It has been recalled 4 times.

Decided
Jan 4, 2018
Review
119 days
Class
Class II
Product code
CAD
Recalls
4
Predicates
2

Its longest line of descent runs back 3 generations to K892057 (1989). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK892057 Ohmeda Tec 5 Continuous Flow VaporizerOhmeda Medical · 1989
  2. 2 generations backK012924 Datex-Ohmeda Tec 7 Anesthesia VaporizerDatex-Ohmeda, Inc. · 2002 · Recalled
  3. 1 generation backK031027 Datex-Ohmeda Tec 7 Anesthesia Vaporizer (Enflurane)Datex-Ohmeda, Inc. · 2003
  4. This deviceK172702 Tec 820, Tec 850Datex-Ohmeda, Inc. · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K892057Ohmeda Tec 5 Continuous Flow VaporizerOhmeda Medical1989—
−2K012924Datex-Ohmeda Tec 7 Anesthesia VaporizerDatex-Ohmeda, Inc.20021
−1K031027Datex-Ohmeda Tec 7 Anesthesia Vaporizer (Enflurane)Datex-Ohmeda, Inc.2003—
0 (this device)K172702Tec 820, Tec 850Datex-Ohmeda, Inc.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
FIG. 1 — Lineage of K172702, 1989–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CAD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CAD device

  1. #1 K240375 V80 Anesthetic Vaporizer (V80)score 8.9 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031027Datex-Ohmeda Tec 7 Anesthesia Vaporizer (Enflurane)
Summary page 4
…de Name: Tec 820, Tec 850 Common/Usual Name: Vaporizer, Anesthesia Classification Names: 21CFR868.5880 Product Code: CAD Predicate Device(s): Tec 7 (K031027 and K012924) Device Description: The Datex-Ohmeda Tec™ 820 and Tec 850 vaporizers are designed…
Datex-Ohmeda, Inc.2003—
*K012924Datex-Ohmeda Tec 7 Anesthesia Vaporizer
Summary page 4
…820, Tec 850 Common/Usual Name: Vaporizer, Anesthesia Classification Names: 21CFR868.5880 Product Code: CAD Predicate Device(s): Tec 7 (K031027 and K012924) Device Description: The Datex-Ohmeda Tec™ 820 and Tec 850 vaporizers are designed for the mete…
Datex-Ohmeda, Inc.20021

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 18, 2024Class IIZ-0815-2025Process controlDue to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Nov 18, 2024Class IIZ-0818-2025Process controlDue to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Nov 18, 2024Class IIZ-0816-2025Process controlDue to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Nov 18, 2024Class IIZ-0817-2025Process controlDue to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 2 companies; a new generation was cleared every 12.8 years on median.
  • 1 of the 4 ancestors has a recall of any class.

How does its category compare?

All CAD devices

CAD Vaporizer, Anesthesia, Non-Heated: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · CAD

1in 2024

2011: 12015: 12018: 12024: 1
2007high 1 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CAD

Too few CAD 510(k)s in any year since 2007 to give a median.

Recalls per year · CAD

2in 2025

2024: 52025: 2, 2 Class I
2007high 5 (2024) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · CAD

2,678in 2025

2007: 302008: 292009: 392010: 322011: 412012: 142013: 122014: 52015: 412016: 92017: 142018: 112019: 122020: 112021: 122022: 132023: 442024: 2552025: 2,678
2007high 2,678 (2025) · low 5 (2014)2025

MAUDE reports naming product code CAD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—030
20080—029
20090—039
20100—032
20111—041
20120—014
20130—012
20140—05
20151—041
20160—09
20170—014
20181—011
20190—012
20200—011
20210—012
20220—013
20230—044
20241—5255
20250—2 (2)2,678
20260—0—

What is it indicated for?

The Tec 820 and Tec 850 vaporizers are designed for use in continuous flow techniques of inhalation anesthesia. They are available in isoflurane and sevoflurane. Each vaporizer is agent specific and is clearly labeled with the anesthetic agent for which it is designed. The vaporizer is temperature, flow and pressure compensated so that its output remains relatively constant despite cooling due to evaporation, variations in inlet flow and fluctuating pressures. The vaporizer is designed for use only with General Electric's Selectatec Series Manifolds.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172702: Tec 820, Tec 850.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K172702

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026